The FDA asked 26 questions about generative AI medical devices, and 509 filers answered. For device makers and clinicians, each topic runs in four steps: the rules today, what the FDA proposed, how the public responded, and the likely outcome.
This page is not an FDA document, and it does not endorse, recommend or predict any outcome. It is a look at the public comments and at one possible result. Each share is out of the filers who answered that question, and each is a minimum.
The FDA puts each device in one of three classes, from Class I (lowest risk) to Class III (highest). The class depends on what the maker says the device is for, not on its technology, so there is no class for generative AI.
A new risk rating made for generative AI, where today there is only the device’s class.Each function would be rated on two axes: what the device does (informs, directs an action, acts under a clinician’s supervision, or acts alone) and how much harm a wrong output could cause (limited, moderate or severe). The paper also says that a line such as “talk to your doctor” does not necessarily make an output less directive.
A majority accepted the two axes, nearly all of them with changes. The change asked for most was to count whether a mistake can be caught or undone before it causes harm. Nearly every filer who addressed disclaimers said a disclaimer does not lower the risk.
The two-axis rating is likely to be added on top of the three classes, which are set in law. A third factor may be added to the rating: whether a mistake can be caught or undone. A disclaimer is unlikely to be sufficient to lower a device’s risk rating.
The evidence the FDA asks for rises with the class. Most Class II devices are cleared by showing they are equivalent to a device already on the market, and Class III devices need FDA approval, backed by laboratory and clinical studies.
A new two-step evaluation for every generative AI device, modeled on how clinicians are assessed.First, benchmark the finished device, not the model alone, on ten competencies without patients. Then confirm it in real or representative clinical use, which the paper says might not need a prospective clinical study in every case.
Most accepted the two steps, and most said the amount of evidence should rise with a device’s place on the risk axes. Few asked for a new clinical study for every device. Requests for a prospective study centered on the highest-risk functions.
The amount of evidence is likely to follow a device’s place on the risk axes. Lower-risk devices may be confirmed without a new clinical study. The highest-risk functions may still need one.
Each company designs and runs its own tests and sends the results to the FDA. Accredited third parties are already allowed to review some low-to-moderate risk devices.
One list of ten competencies that every device is benchmarked on, where today each company chooses its own tests.The ten run from clinical knowledge to generalizability, and synthetic data could extend testing where real patient inputs are limited. The paper asks who a device should be compared against, and what role independent third parties should have in evaluating devices.
Filers who judged the ten competencies kept them and added more. A majority said synthetic data suits rare cases and stress tests, and a majority said the clinician and the AI should be measured together where a clinician acts on the output. On a company’s own benchmarks, the leading answer was to accept them only with safeguards, and about half asked for rules on third-party evaluators’ qualifications and independence.
A company’s own benchmark may need safeguards, such as test data its developers never see. Synthetic data may be accepted for rare cases and stress tests, and where a clinician acts on the output, performance should be measured for the clinician and the AI together. Third-party evaluators may have to meet rules on qualification and independence.
Device makers must keep complaint files and report deaths, serious injuries and certain malfunctions to the FDA. The FDA’s 2025 draft guidance on AI devices recommends a performance monitoring plan, but the guidance is a draft and its recommendations are not binding.
Ongoing checks of a device’s performance after launch, where today’s rules center on reporting problems once they happen.The benchmark a device passed before launch would become its baseline, and monitoring would watch for drift from it. The paper asks whether closer monitoring can make up for less certainty before launch, whether a second AI can help do the monitoring, and what part clinicians and hospitals should play.
Most said closer monitoring could stand in for some testing before launch, almost always with conditions. About half said the riskiest uses should still be fully tested first, and a third asked for a monitoring plan set before launch. Most said the maker stays accountable, and a majority said that holds even for AI it did not build.
Ongoing checks may come with a monitoring plan set before launch, and closer monitoring may stand in for some testing before launch. The riskiest uses may still need full testing first. Responsibility is likely to stay with the maker, as it is today.
A change that could significantly affect safety or effectiveness needs a new FDA submission before it ships. A maker can avoid that with a predetermined change control plan, authorized with the device, which lists specific planned changes and how each is to be tested.
Rules for a change the device maker does not make itself: an update to the underlying model by its developer.A changed device would be re-tested against its pre-launch benchmark. The paper asks how much re-testing each change needs, and how a change plan can cover changes that cannot be listed in advance.
On a model developer’s update, a majority asked for each of three things: re-test before the new version reaches patients, track which version is running, and get advance notice of changes. Half said the amount of re-testing should follow a change’s clinical effect, not its technical size. About half said a change plan should cover bounded categories of change.
The device maker should re-test before the new version reaches patients, and should track which version is running. Change plans may cover categories of change in place of a list of specific ones. How much re-testing a change needs may depend on its clinical effect.
A supplier can lodge a confidential master file with the FDA, and a device maker can cite it without seeing what is in it. The FDA’s 2025 draft guidance on AI devices does not mention generative AI, foundation models or agents.
A master file made for foundation models, and questions on devices that act on their own.A model developer could give the FDA details of its model in confidence through a voluntary Foundation Model Master File, and device makers could cite it. For a device that plans and acts in several steps, the paper makes no proposal and asks what extra oversight is needed.
Nearly all who took a side backed the master files, most of them with changes, and about half asked the FDA to set standard content for them. For a device that takes actions, the leading requests were a person’s sign-off before risky actions, records of what the device did, tests of the whole chain of actions, and limits on what it may do.
The FDA may set what a master file must contain, and device makers could cite a file in their own submissions. A device that takes actions may need a person’s sign-off before risky actions, and records of what it did.
Filers named six different defaults, and none drew a majority. The most common answer was a tiered set of comparators, followed by whoever does the task in the real workflow. The FDA’s own example, the median clinician, drew almost no support.
Filers set terms for evaluators more often than they took a side. About a quarter asked that third-party evaluation stay optional, industry more often than clinicians. A fifth asked that it depend on the device’s risk.
Filers split. Some would allow a new model version under a change plan agreed in advance. Others said it needs a new FDA review.
Filers split. Some said taking actions makes a device high risk by default, and others said it does not. More than a quarter asked that the rules follow how much the device does on its own and what its actions can affect.
Behind this outline: 87% of the 301 filers who gave a yes or no on the FDA’s core proposals said yes, with changes (Takeaway 5), and 54% of the 479 who took an overall position asked for more oversight than the FDA proposed (Takeaway 3).
Summary report: ten takeawaysResults for all 26 questionsMethodologyThe FDA’s 26 questionsThe FDA’s discussion paper