Q1 · The two-axis risk frameworkQ2 · The spectrum of device activityQ3 · When an output becomes directiveQ4 · Generalist and specialist usersQ5 · Multi-turn conversations that migrateQ7 · The competency-based approachQ9 · The benchmarking structureQ11 · Clinical confirmation without a prospective trialQ13 · Synthetic dataQ14 · Comparators and acceptance criteriaQ16 · Independent third partiesQ18 · Trading premarket certainty for postmarket monitoringQ19 · Postmarket performance evaluationQ20 · Machine-based supervisory agentsQ21 · Clinicians, institutions and societiesQ22 · Re-benchmarking after a modificationQ23 · PCCPs for GenAI devicesQ24 · Third-party foundation model changesQ25 · Foundation Model Master FilesQ26 · Agentic devices
The comment as filed
The attached comments are on behalf of the American Medical Informatics Association (AMIA).
Attachment
Submitted Electronically via Regulations.gov
October 19, 2026
Michelle Tarver, MD, PhD
Director
Center for Devices and Radiological Health
U.S. Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
Re: Docket No. FDA-2026-N-7874—Considerations for the Regulation of Generative AI-Enabled
Medical Devices: Discussion Paper and Request for Feedback
Dear Dr. Tarver:
The American Medical Informatics Association (AMIA) appreciates the opportunity to provide
feedback to the U.S. Food and Drug Administration’s Center for Devices and Radiological Health
(CDRH) on its discussion paper, Considerations for the Regulation of Generative AI-Enabled
Medical Devices. AMIA is the professional home for more than 6,400 clinicians, researchers,
educators, data scientists, and health technology leaders dedicated to advancing health and
healthcare through informatics. Our members develop, evaluate, implement, and oversee health
information systems and artificial intelligence-enabled technologies across clinical care, research,
public health, and education. Drawing on this multidisciplinary expertise, AMIA supports a riskproportionate, evidence-based, and total product lifecycle approach that protects patients,
promotes transparency and accountability, enables interoperability and meaningful real-world
evaluation, and supports responsible innovation.
Scope and Definition of GenAI Enabled Medical Devices
AMIA recommends that CDRH clarify which generative AI-enabled functions qualify as device
functions and therefore fall within FDA's regulatory oversight. Existing guidance, most notably FDA's
"Policy for Device Software Functions and Mobile Medical Applications" 1 and "Clinical Decision
Support Software” 2 guidance, provides numerous examples of software functions that are and are
not the focus of oversight. However, those examples, including the January CDS guidance, do not
U.S. Food and Drug Administration. (2022, September 28). Policy for device software functions and mobile
medical applications: Guidance for industry and Food and Drug Administration staff.
https://www.fda.gov/media/80958/download
U.S. Food and Drug Administration. (2026, January 29). Clinical decision support software: Guidance for
industry and Food and Drug Administration staff. https://www.fda.gov/media/109618/download
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 1 of 22
directly address generative AI-enabled functions, leaving industry stakeholders and developers to
infer how the existing policies apply.
Many issues raised in this discussion paper are also relevant to widely adopted tools that AMIA
believes are non-device software functions. AMIA's responses would likely differ for this broad
group of non-device AI-enabled functions and for the narrower set of AI-enabled device software
functions subject to FDA oversight. Concrete examples, parallel to those in existing guidance,
would help stakeholders understand which technologies FDA intends the proposed frameworks to
cover and provide more focused feedback.
Illustrative examples that we believe fall within scope, depending on their intended use and
functionality, include:
• Generative imaging reconstruction that creates diagnostic-quality images from
undersampled data;
• Multimodal deterioration-detection systems that analyze waveforms, laboratory results,
and clinical notes and generate patient-specific narratives or recommended actions;
• Patient-facing applications that interpret symptoms or physiologic data and recommend a
specific medication dose; and
• AI-enhanced patient monitoring devices that convert measurements into patient-specific
diagnostic, treatment, or escalation recommendations.
Illustrative examples that we believe fall outside of scope include:
• Recording of a clinical encounter, AI-driven transcription, and drafting of a clinical note,
potentially inclusive of drafting orders for physician review
• Summarization of a patient’s chart so long as it is not inclusive of near real-time signal data.
• AI-driven evidence-retrieval tools, including Open Evidence and commercially available
foundation models, whether used by clinicians or patients, so long as the outputs meet the
general examples provided in the CDS guidance.
FDA should clarify when technologies such as these become subject to oversight. CDRH should
map representative GenAI functions to the statutory definition of a device and to each criterion for
non-device clinical decision support, including examples showing when patient-specific outputs,
directives, clinical claims, or influence on diagnosis or treatment bring a function within FDA
oversight. This clarification would ensure that stakeholders evaluate the same categories of
technology and would provide the necessary foundation for the postmarket infrastructure
recommendations that follow.
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 2 of 22
Section IV: Considerations for the Assessment of Risk for GenAI-Enabled Devices
Question 1: Does the two-axis risk framework, organized around device activity and the
consequence of relying on an incorrect output, appropriately capture the dimensions most
relevant to the risk of a GenAI-enabled software function? If there are additional dimensions—
such as the reversibility of a resulting action, the availability of downstream safeguards, the
time pressure of the deployment setting, or the traceability of the output (i.e., to primary
source materials)—that should be represented in a risk framework, please describe and
provide examples of how they should be represented.
AMIA agrees that device activity and the consequences of an incorrect output are relevant, but the
proposed axes should not be treated as a complete risk model. CDRH should anchor classification
in the probability and severity of harm, then apply a limited set of contextual modifiers that can
raise or lower either dimension. This approach keeps the framework usable without reducing a
multidimensional problem to categorical labels such as non-directive, directive, or action-taking.
• The probability that an incorrect or misleading output will be relied upon before correction;
• The reversibility of harm if the output is relied upon;
• Whether the output can be independently reviewed, challenged, and overridden;
• The intended user's relevant expertise, health literacy, and ability to recognize an error;
• Potential influence on the user, highlighted by the strength of the recommendation, degree
of personalization, and cumulative effect of multi-turn interaction;
• The autonomy of the system and its operation;
• The quality and traceability of supporting evidence, including evidence grade where
available, source fidelity, and uncertainty communication; and the
• The scale of deployment and potential population impact.
FDA should use these dimensions to establish transparent risk tiers and illustrative examples. The
framework should not presume that informational functions are inherently safer than actiondirecting functions since contextual information provision itself can be persuasive or that
professional expertise eliminates automation bias. CDRH also should not assume that risk rises
linearly or exponentially with autonomy. Autonomy can create nonlinear risk through scale,
compounding errors, and fewer opportunities to intervene, but some autonomous functions may
reduce human error. Classification should therefore rely on empirical performance in the intended
workflow and under expected fallback conditions.
Human review should count as a risk-reducing safeguard only when the reviewer has the time,
information, clinical and AI expertise, and authority to act, and when human and model failure
modes are not so correlated that review merely confirms the same error. Where possible,
evaluation should compare the device alone, the unaided clinician, and the human-AI team.
Below is an example of a multidimensional risk framework with 4 stages to determine a device’s
risk.
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 3 of 22
1. Assess the probability and severity of harm
First, assess the device using two core dimensions that align with established risk-assessment
practice:
• Probability of harm: How likely is the device to generate an incorrect, misleading, or
incomplete output?
• Severity of harm: If an incorrect output is relied upon, could it cause inconvenience,
unnecessary care, delayed treatment, serious injury, or a life-threatening outcome?
Baseline risk increases as both the likelihood of harm and the severity of that harm increase; the
relationship should not be assumed to be linear. As a contextual modifier, the risk increases if there
is high probability of reliance before correction.
For example:
• A tool that summarizes general clinical information may begin in a lower-risk area only when
errors are unlikely to cause harm, any potential harm is not severe, and users can readily
verify the source information with minimal additional effort.
• A patient-facing tool that recommends whether to seek emergency care would begin in a
higher-risk area because an incorrect output, even if improbable, is likely to be relied upon
without correction. Additionally, it may prompt harmful outcomes: either delayed care or
unnecessary emergency care utilization.
• A system that recommends or autonomously changes an insulin dose would begin in the
highest-risk area because an error could produce immediate and severe harm.
This placement establishes the baseline risk, not the final classification.
2. Apply AMIA’s contextual risk modifiers
Next, assess the circumstances in which the device will operate. The same contextual modifiers
identified above should be applied consistently, with each modifier potentially increasing or
decreasing the estimated probability or severity of harm:
• Reversibility of harm if the output is relied upon: Can an action or resulting harm be
detected and reversed before serious or lasting consequences occur?
• Ability to independently review, challenge, and override: Can a qualified person
meaningfully verify, challenge, stop, or override the device before harm occurs, and does
empirical testing demonstrate that human review corrects errors rather than amplifying
shared reasoning flaws or biases?
• Intended user’s relevant expertise, health literacy, and ability to recognize an error: Is
the output delivered to a patient, caregiver, generalist, specialist, or care team, and has the
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 4 of 22
device been evaluated for the intended users across relevant languages, literacy levels, and
clinical settings?
• Potential influence on the user: How strongly does the output recommend an action, how
personalized is it, and could conversational, anthropomorphic, or multi-turn interactions
increase reliance over time?
• Autonomy of the system and its operation: Does the device provide information,
recommend an action, initiate an action, or autonomously execute a multi-step task, and
where are meaningful opportunities for human intervention?
• Urgency of the decision and availability of downstream safeguards: Is the decision timesensitive, and are independent safeguards available to identify and correct an error before
harm occurs?
• Quality and traceability of supporting evidence: Can users assess the evidence
supporting the output, its source fidelity, applicable evidence grade, limitations, and
validated uncertainty information, and is an appropriate audit trail available?
• Scale of deployment and potential population impact: Could widespread use amplify
errors, inequities, workflow disruption, automation bias, privacy risks, or cybersecurity
vulnerabilities, particularly across historically underserved populations or varied clinical
and technological environments?
These modifiers help prevent two devices with similar intended functions from receiving the same
risk classification when their operating conditions and potential effects differ materially.
As a secondary aid for applying two of these same modifiers, CDRH could use a simple matrix
showing the relationship between an output’s potential influence on the user, particularly the
strength of its recommendation, and the quality and traceability of its supporting evidence. A
strong, personalized recommendation based on weak or unverifiable evidence should increase the
assessed risk even when the output is labeled informational. Conversely, a less directive output
supported by strong, traceable evidence may present lower interaction risk. This matrix should help
apply the contextual modifiers, not replace the underlying probability-and-severity assessment.
3. Assign an overall risk tier
After combining the baseline matrix placement with the contextual modifiers, the reviewer assigns
an overall risk tier. The table below illustrates how the two core axes establish a preliminary risk
level and how contextual modifiers may move a device higher or lower. An upward arrow (↑)
indicates potential upward pressure on risk, while a downward arrow (↓) indicates potential risk
mitigation. No single modifier should automatically determine the final tier.
Risk Tier Probability and severity Contextual modifiers Illustrative example
of harm influencing placement
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 5 of 22
Lower Low probability and ↓ Reversibility: Effects are A clinician-facing tool
limited severity of harm. readily corrected. summarizes general
↓ Independent review: clinical information,
Users can verify the output cites its sources, does
before relying on it. not recommend or
initiate an action, and is
↓ Influence: The output is reviewed before use.
non-directive.
Subgroup testing
↓ Evidence: Sources and identifies no material
uncertainty are clearly performance differences
presented. among intended users.
↓ Scale: Deployment is
limited or controlled.
Moderate Moderate probability or ↑ Intended user: A patient, A patient-facing tool
severity of harm, with caregiver, or generalist may recommends routine
generally manageable or have difficulty recognizing clinical follow-up. A
reversible effects. an error. clinician can review the
↑ Influence: The output is recommendation before
personalized or treatment changes
recommends an action. occur, but testing
identifies reduced
↓ Reversibility and
safeguards: Effects are
comprehension or
reversible, and clinician accuracy for users with
review or another limited English
safeguard is available. proficiency.
↑ Equity: Usability or
performance varies across
language, literacy,
disability, or demographic
groups.
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 6 of 22
High Meaningful probability of ↑ Independent review: A patient-facing
serious harm or a lower- Verification is difficult or emergency-triage tool
probability error with unlikely before reliance. provides strongly
potentially serious ↑ Urgency and worded, personalized
consequences. safeguards: The decision recommendations that
is time-sensitive, with few users are likely to follow
downstream checks. immediately. Testing
identifies higher under-
↑ Influence: The output
strongly directs an action.
escalation rates for a
clinically relevant
↑ Evidence: Supporting subgroup, and
evidence is incomplete or deployment across
difficult to verify. multiple health systems
↑ Scale and equity: Broad could expose many
deployment could amplify patients to the disparity.
subgroup-specific errors or
disparities.
Very High Plausible severe, ↑ Reversibility: Harm may An autonomous agent
irreversible, or life- be difficult or impossible to changes insulin or
threatening harm. reverse. anticoagulant dosing
↑ Autonomy: The system and executes related
initiates or executes orders across a health
clinical actions. system without a
mandatory human
↑ Urgency and
checkpoint. An error
safeguards: There is little
opportunity to intervene
could cause immediate
before harm occurs. serious harm, propagate
through multiple actions,
↑ Compounding actions: and disproportionately
Multi-step outputs can affect a subgroup for
propagate an initial error. which performance is
↑ Scale and population lower.
impact: System-wide
deployment could produce
widespread harm or
magnify inequitable
outcomes.
Equity is a cross-cutting consideration within the intended-user and population-impact modifiers.
Materially worse usability, performance, or outcomes for a clinically relevant subgroup may justify a
higher risk tier even when average performance appears acceptable.
The framework does not necessarily require a numerical score. FDA could instead publish decision
criteria and examples showing when particular modifiers should move a device into a higher or
lower tier.
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 7 of 22
4. Match oversight to the risk tier
The assigned tier should determine the rigor of regulatory evidence, safeguards, and monitoring.
However, usability testing, subgroup analysis, standardized AI exposure and provenance tracking,
basic transparency, complaint and adverse-event monitoring, and periodic reassessment should
apply across all tiers. These measures should help validate the initial risk classification and trigger
reassignment when real-world evidence indicates greater or lesser risk. This is important because
usability problems and unequal performance can increase the risk of otherwise beneficial
technologies and contribute to intervention-generated inequalities. 3
The same interoperable exposure and provenance infrastructure should support every tier,
although the granularity of data collection, frequency of review, and intensity of regulatory oversight
should increase with risk:
• Lower-risk devices: Streamlined validation proportionate to the intended use; baseline
usability testing and subgroup analysis; disclosure of intended use, limitations, and known
risks; and periodic review of exposure, performance, complaint, and safety data.
• Moderate-risk devices: Lower-risk measures conducted with greater rigor, supplemented
by representative benchmarking, scenario-based usability testing, clinically relevant
subgroup analyses, clinical confirmation, defined human-oversight procedures, and more
frequent performance review.
• High-risk devices: Moderate-risk measures supplemented by prospective clinical
evidence, independent evaluation, robust human-factors testing, continuous performance
monitoring, detailed audit logs, strict change controls, and procedures for rapidly
escalating identified safety or equity concerns.
• Very-high-risk devices: High-risk measures supplemented by especially strong premarket
evidence, explicit authorization and operating limits, mandatory human checkpoints where
feasible, rapid shutdown or rollback capabilities, intensive postmarket surveillance, and
frequent reassessment of usability and performance across relevant populations.
Question 2: CDRH seeks input on how to account for the spectrum of GenAI-enabled
informational functions that vary in the degree to which they direct a user to a particular
action (i.e., between “non-directive” and “action-directing”). What characteristics of an
output—such as its wording, specificity, personalization, or context—could be considered as
modifiers of the risk associated with an informational function, after accounting for the
device’s overall functionality and intended use? What additional information would provide
manufacturers with sufficient clarity and predictability around risk assessment for
Veinot TC, Mitchell H, Ancker JS. Good intentions are not enough: how informatics interventions can worsen
inequality. J Am Med Inform Assoc. 2018;25(8):1080–1088. doi:10.1093/jamia/ocy052
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 8 of 22
informational functions while recognizing that directiveness may exist along a continuum
rather than as a binary distinction?
AMIA cautions against treating directiveness as a binary characteristic or a reliable stand-alone
proxy for risk. The distinction between informational and action-directing output cannot be
determined from wording or a sponsor's label alone. An output framed as informational may still
cause action or inaction 4,5 when it is personalized 6, long or repeated 7, presented with unwarranted
certainty 8, or delivered through a conversational interface that develops rapport and trust 9,10. Risk
should reflect the probability and severity of harm in actual use and the output's foreseeable
influence on a clinical decision.
FDA should provide examples addressing specificity, personalization, urgency, imperative
language, recommendation strength, the quality and traceability of supporting evidence, validated
uncertainty, anthropomorphic design, cumulative conversational context, and the likelihood that
an output will alter care. A strong recommendation based on weak or unverifiable evidence should
be treated as a meaningful risk modifier. Directiveness should be evaluated across the full
interaction, and a disclaimer alone should not convert action-directing content into a low-risk
informational function.
CDRH should use usability and human-factors evidence to test how intended users interpret and
act on outputs under realistic conditions. This evidence is more reliable than attempting to classify
a device solely from the formal wording of a single response.
Question 3: CDRH seeks input on whether and when a GenAI-enabled function that results in
delivery of clinical information to patients, as opposed to HCPs, could present different or
Bickmore TW, Trinh H, Olafsson S, O'Leary TK, Asadi R, Rickles NM, Cruz R. Patient and consumer safety
risks when using conversational assistants for medical information: an observational study of Siri, Alexa, and
Google Assistant. Journal of medical Internet research. 2018 Sep 4;20(9):e11510.
https://pubmed.ncbi.nlm.nih.gov/30181110/
Shekar, S., Pataranutaporn, P., Sarabu, C., Cecchi, G. A., & Maes, P. (2025). People Overtrust AI-Generated
Medical Advice despite Low Accuracy. NEJM AI. https://ai.nejm.org/doi/full/10.1056/AIoa2300015
Salvi, F., Horta Ribeiro, M., Gallotti, R. et al. On the conversational persuasiveness of GPT-4. Nat Hum Behav,
1645–1653 (2025). https://doi.org/10.1038/s41562-025-02194-6
Steyvers M, Tejeda H, Kumar A, Belem C, Karny S, Hu X, Mayer LW, Smyth P. What large language models
know and what people think they know. Nature Machine Intelligence. 2025 Feb;7(2):221-31.
https://doi.org/10.1038/s42256-024-00976-7
Zhou K, Hwang J, Ren X, Sap M. Relying on the unreliable: The impact of language models’ reluctance to
express uncertainty. In Proceedings of the 62nd Annual Meeting of the Association for Computational
Linguistics (Volume 1: Long Papers) 2024 Aug (pp. 3623-3643). https://aclanthology.org/2024.acl-long.198/
Li Q, Luximon Y, Zhang J. The Influence of Anthropomorphic Cues on Patients’ Perceived
Anthropomorphism, Social Presence, Trust Building, and Acceptance of Health Care Conversational Agents:
Within-Subject Web-Based Experiment. J Med Internet Res 2023; https://www.jmir.org/2023/1/e44479
Effects of Persuasive Dialogues: Testing Bot Identities and Inquiry Strategies. W. Shi, X. Wang, Y. Oh, J.
Zhang, S. Sahay, and Z. Yu. Proceedings of the 2020 CHI Conference on Human Factors in Computing
Systems. https://doi.org/10.1145/3313831.3376843
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 9 of 22
higher risks, while also recognizing the potential benefits associated with improved patient
empowerment, engagement, and access to clinical information. What device characteristics,
output features, or safeguards might mitigate risks that could arise when a user lacks the
domain knowledge to independently evaluate an output, without unnecessarily
underestimating patient capability?
Patient-facing functions should not automatically be classified as higher risk, but risk increases
when delivered by a computer application imbued with authority based on who and where it is
delivered 11, outputs are personalized, action-directing, difficult to verify, or likely to influence urgent
clinical decisions. Human-like avatars, voice and conversational interfaces may also increase
perceived authority and trust and should be considered in the risk assessment8,12.
Safeguards should include plain-language disclosure that the user is interacting with AI,
communication of uncertainty, citations or source provenance, clear scope limitations, emergencyescalation protocols, access to human assistance, privacy protections, and meaningful opt-in and
opt-out choices. Patient-facing devices should be tested with equal rigor across supported
languages, literacy levels, disabilities, and infrastructure conditions, including low-bandwidth
environments. Patients and caregivers should participate in design and evaluation so that
safeguards reflect real-world health and digital literacy needs.
Question 4: CDRH seeks input on whether and how the distinction between generalist and
specialist physicians might inform the assessment of risk for HCP-facing GenAI-enabled
functions. Under what circumstances, if any, might risk be affected when an HCP who lacks
the relevant clinical specialist knowledge receives information that falls within a specialist
area of practice? What device characteristics or safeguards might mitigate such risks?
The distinction between generalist and specialist users should inform, but not independently
determine, risk. FDA should evaluate whether the intended user can identify an incorrect output,
interpret uncertainty, and recognize when specialist consultation is required. Expertise alone
should not be treated as a sufficient safeguard because specialists may also over-rely on AI outputs
under time and workflow pressures.
Devices should be tested with the clinicians expected to use them, not only subject-matter experts.
Safeguards could include source traceability, empirically calibrated uncertainty information, clear
scope limitations, referral recommendations, and mandatory secondary review for highconsequence outputs or when validated thresholds are not met. For example, an AI-generated
Kopka M, Schmieding ML, Rieger T, Roesler E, Balzer F, Feufel MA. Determinants of Laypersons’ Trust in
Medical Decision Aids: Randomized Controlled Trial. JMIR Hum Factors 2022;9(2):e35219
https://humanfactors.jmir.org/2022/2/e35219
Cohn M, Pushkarna M, Olanubi GO, Moran JM, Padgett D, Mengesha Z, Heldreth C. Believing
anthropomorphism: examining the role of anthropomorphic cues on trust in large language models. In
Extended Abstracts of the CHI Conference on Human Factors in Computing Systems 2024 May 11 (pp. 1-15).
https://dl.acm.org/doi/10.1145/3613905.3650818
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 10 of 22
imaging interpretation should not replace independent review of the underlying image when the
intended workflow requires professional confirmation.
Question 5: For multi-turn conversational GenAI-enabled devices that may migrate from
providing “non-directive” information to “action-directing” information over the course of an
exchange, how should risk be assessed across realistic conversational trajectories? How
could the intended use of such a device be characterized when its behavior is emergent
across a conversation?
FDA should evaluate the complete conversational trajectory, including reasonably foreseeable
situations in which a device moves from general information to personalized or action-directing
advice. The assessment should account for increasing rapport and trust over multiple turns and for
errors that accumulate when later responses build on an earlier incorrect premise.
Testing should include branching scenarios, repeated questioning, incomplete or contradictory
information, attempts to move the device outside its intended scope, and conversations involving
urgent symptoms. Risk should reflect the highest-risk behavior reasonably expected across the
interaction, rather than only the function performed at the first turn.
FDA should require a specific intended-use envelope and test whether the device consistently
maintains those boundaries. Behavior described as emergent should not excuse recurring scope
drift into unvalidated diagnosis, treatment, or other off-label functions; frequent boundary failures
should be treated as a material safety concern.
Section V: A Competency-Based Approach for Premarket Evaluation of GenAI Enabled Devices
Question 7: Is the competency-based approach described above, i.e., device benchmarking
followed by clinical confirmation, a useful and appropriate framework for evaluating GenAIenabled devices?
AMIA supports competency-based evaluation as a useful component of FDA review, but
benchmarking should complement, not replace, clinical validation.
FDA should evaluate the final configured product within its intended workflow, including its
interaction with clinicians, patients, EHRs, external tools, and institutional safeguards. Evaluation
should address clinical performance, health equity, usability, human factors, workflow effects,
privacy, security, failure recovery, and meaningful human oversight.
Question 9: Would a benchmarking structure such as the one described above be likely to
provide adequate evidence of clinical knowledge, analytic capabilities, safety behavior,
communication, and generalizability to support a reasonable assurance of safety and
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 11 of 22
effectiveness? Are there elements that are missing, redundant, or inappropriately
categorized? Are there externally developed standards that could be leveraged?
The proposed categories are a strong starting point, but benchmarking should include more than
clinical knowledge and output accuracy. FDA should also assess:
• Hallucinations and harmful omissions;
• Calibration and communication of uncertainty;
• Source fidelity and traceability;
• Performance across patient subgroups and care settings;
• Robustness to incomplete, unusual, or adversarial inputs;
• Human-AI team performance;
• Usability, automation bias, and clinician workload;
• Safe failure and escalation behavior; and
• Privacy and cybersecurity.
Benchmark reports should quantify rates of hallucinations and harmful omissions rather than
merely record whether they occurred. Results should include denominators, severity, clinical
context, uncertainty intervals, subgroup rates, and performance under both routine and adversarial
conditions. Rare, high-consequence failures should be reported separately so they are not
obscured by aggregate accuracy.
Benchmarks should use independent or sequestered datasets where appropriate, and public
summaries should identify the device version, intended use, test population, performance
thresholds, and known limitations.
For moderate- and high-risk devices, FDA should encourage qualified independent adversarial
testing that probes misleading prompts, unusual inputs, workflow failures, and attempts to bypass
safeguards. Common data models, standardized data elements, and consistent reporting formats
would make results more comparable across manufacturers, populations, and deployment
settings. Benchmarking should be repeatable after material updates and should evaluate the
complete deployed system rather than the foundation model alone.
Question 11: CDRH is considering that clinical confirmation for a GenAI-enabled device might
not require a prospective clinical study in every case, and has described above a range of
approaches of increasing rigor and patient exposure. How might a sponsor select and justify a
confirmation approach tailored to a device’s intended use and proportionate to the device’s
risk profile? Are there device types or risk profiles for which one or more of these approaches
would be insufficient or inappropriate? How might the anticipated distribution of real-world
inputs be taken into consideration? Are there other methods of clinical confirmation that
might help inform the evaluation of GenAI-enabled devices?
FDA should use a staged, risk-proportionate approach:
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 12 of 22
• Lower-risk functions may be supported by retrospective evaluation, usability testing, and
representative workflow simulations.
• Moderate-risk functions may require shadow deployment, multi-site testing, clinician
adjudication, and subgroup analysis.
• High-risk, autonomous, or action-taking functions may require prospective comparative
studies and, when appropriate, randomized trials.
Clinical confirmation should reflect realistic inputs, edge cases, underserved populations, different
care settings, and the performance of the human-AI team when the device is intended to assist
clinicians. Documentation should be standardized and should avoid duplicating information
already maintained through quality-management systems.
Question 13: For which clinical domains, device functions, or subpopulations is synthetic data
particularly well-suited, or particularly inadequate, as a supplement to real world evidence?
What safeguards would mitigate the risk that synthetic data generated by models of the same
class as the device under evaluation reproduces the very performance gaps the evaluation is
intended to detect, particularly for underrepresented subgroups?
Synthetic data can supplement real-world data for rare conditions, safety-critical edge cases, and
populations that are difficult to sample, but it should not replace real-patient evidence for high-risk
uses.
Sponsors should document how synthetic data were generated, demonstrate that clinically
important relationships and subgroup characteristics were preserved, and evaluate whether the
generating model reproduced existing biases. Final performance claims should be confirmed using
independent real-world data, particularly for historically underrepresented populations.
Question 14: For open-ended device outputs, how should performance comparators and
acceptance criteria be selected? When a panel of qualified clinicians serves as the
comparator, how should the applicable standard (for example, the standard of care versus the
performance of a median clinician in practice) be defined and justified? Should generalist or
specialist physicians be used as a performance standard? When should human-AI team
performance, rather than the device operating alone, serve as the basis for evaluation?
Comparators should reflect the device’s intended use and the applicable standard of care. The
performance of a “median clinician” should not be the primary benchmark when prevailing practice
falls below evidence-based standards.
Generalists should serve as comparators for generalist-facing tools, while specialists should be
used for specialty-level tasks. Assistive systems should be evaluated primarily through human-AI
team performance, while autonomous systems should also demonstrate acceptable standalone
performance. Evaluation should include clinical outcomes, harmful omissions, equity, usability,
workload, and resource utilization, not merely agreement with clinicians.
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 13 of 22
CDRH should not assume that adding a human reviewer improves performance. Studies should
compare the unaided clinician, the device alone when relevant, and the human-AI team, and
should examine whether human and model errors are complementary or correlated. Human review
should be credited as a safeguard only when it measurably improves error detection, clinical
decisions, or outcomes under realistic time and workflow constraints.
Question 16: Should independent third parties be involved in some or all aspects of a
competency-based approach, including device benchmarking and clinical confirmation? If so,
in what ways might qualified, independent third-party participation contribute to this
approach, and what qualifications and independence criteria should apply? Are there aspects
of the assessment for which third-party involvement would be impractical or inadvisable?
What safeguards or program design features would be critical to prevent third-party
participation from limiting competition or preventing innovation?
AMIA supports an appropriate role for qualified independent third parties, particularly for
adversarial testing, maintenance of sequestered evaluation datasets, review of sponsor-developed
benchmarks, and clinical confirmation of higher-risk devices. Independent testing can reduce
optimization to known tests and provide greater confidence that performance generalizes beyond
the sponsor's development environment.
FDA should establish transparent qualifications, conflict-of-interest protections, standardized
methods, data-security requirements, and mechanisms for auditing third-party performance.
Participation should be scaled to risk, should not impose disproportionate costs on smaller
developers, and should not transfer the manufacturer's ultimate responsibility for safety and
effectiveness.
Section VI: Postmarket Monitoring for GenAI-Enabled Devices
Overarching Recommendation to CDRH Regarding Questions 18 to 24
After CDRH clarifies which GenAI-enabled functions fall within the medical-device framework,
AMIA recommends that the agency address a related infrastructure need that applies across
Questions 18-24. For functions subject to FDA oversight, CDRH should work with other HHS
agencies and relevant stakeholders to establish a standardized, interoperable framework for
recording AI exposure and provenance.
The framework should allow authorized stakeholders to determine which AI-enabled device or
function was used, which product and model version was deployed, which patient and encounter
were involved, the clinical context in which the technology was used, and what occurred afterward.
These data are necessary to connect real-world use with clinical processes and outcomes as
models, data environments, and workflows change.
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 14 of 22
The framework should be vendor-neutral and should not require clinicians to document each AI
interaction or manufacturers to submit every interaction to FDA. Relevant data should be generated
automatically wherever possible, remain proportionate to the device's risk, and be available
through interoperable formats that existing EHRs, registries, distributed research networks, and
other authorized data infrastructures can use.
The Discussion Paper appropriately recognizes that variable outputs and post-deployment changes
may limit what premarket evaluation can establish. If FDA relies more heavily on postmarket
evidence, standardized exposure and provenance data will be necessary to evaluate safety,
effectiveness, generalizability, subgroup performance, and changes over time.
AMIA therefore encourages CDRH to treat this data infrastructure as a core component of the
postmarket framework for in-scope GenAI-enabled medical devices and to develop it jointly with
ONC, CMS, standards-development organizations, EHR vendors, health systems, manufacturers,
patients, clinicians, and researchers. Existing resources, including multi-institutional EHR networks
such as Epic Cosmos, illustrate how standardized data generated during routine care could support
national-scale evaluation without requiring a new centralized federal database or reliance on a
single proprietary platform.
Question 18: CDRH is considering whether it may be appropriate to accept greater premarket
uncertainty regarding a GenAI-enabled device’s benefit-risk profile through greater reliance
on postmarket monitoring. Under what conditions might such an approach be appropriate,
and what characteristics of a monitoring program would need to be in place to justify reduced
premarket evidence? Are there device types or risk profiles for which this approach would not
be appropriate?
AMIA supports greater reliance on postmarket monitoring only in limited circumstances. Reduced
premarket evidence may be appropriate for lower- or moderate-risk devices when potential harms
are reversible, meaningful human review is available, deployment is controlled, monitoring can
detect safety and subgroup concerns, and the manufacturer can rapidly suspend, correct, roll
back, or withdraw the product. This approach generally should not apply to autonomous,
emergency, diagnostic, treatment-directing, medication-dosing, or other high-consequence
functions. Postmarket monitoring should not replace premarket mitigation of serious foreseeable
risks.
Any monitoring program used to justify reduced premarket evidence should combine active
performance monitoring, independent review, predefined safety thresholds, subgroup analysis,
prompt investigation, and enforceable procedures for corrective action. It should also identify
which patients were exposed to which device and version, the clinical context in which the device
was used, and the subsequent processes and outcomes.
Routine data generally identify medication exposure, and Unique Device Identifiers support
identification of many implanted devices. CDRH should consider a comparable structured AI
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 15 of 22
exposure event for GenAI-enabled medical devices. Depending on the device and its risk, the event
should include:
• The patient and encounter;
• The device or software function, manufacturer, and standardized product identifier;
• The deployed model and version, including the underlying foundation-model version when
relevant;
• The date, time, clinical setting, and user;
• Whether the function was invoked and whether its output was presented to a clinician or
patient;
• The type of output, recommendation, or action;
• The clinician's or patient's response, such as acceptance, modification, or rejection, when
relevant; and
• Sufficient provenance to link the exposure to subsequent care processes and outcomes.
The minimum dataset should be proportionate to risk and generated automatically wherever
possible so the monitoring system does not create new documentation work for clinicians. This
infrastructure would make reliance on postmarket evidence more credible by allowing FDA and
manufacturers to determine where, when, and how a device was used and whether outcomes
differed after exposure.
Question 19: Please comment on the potential approaches to postmarket performance
evaluation, including periodic re-benchmarking, sample-based clinician review, and
performance degradation monitoring. What additional approaches should CDRH consider,
and how should the cadence and triggering events for reassessment be determined?
AMIA supports periodic re-benchmarking, independent clinician review of real-world samples,
performance-degradation monitoring, and adversarial testing as components of a broader
algorithmovigilance program. Monitoring should be continuous or sufficiently frequent to detect
meaningful change, rather than limited to fixed periodic reviews, because performance may shift
with changes in models, data sources, patient populations, connected tools, or clinical workflows.
The program should track clinical outcomes, near misses, harmful outputs, complaints, clinician
overrides, care-escalation patterns, subgroup performance, usability and workflow burden,
cybersecurity events, and integration failures. Where technically valid, calibrated uncertainty
measures may help identify outputs requiring human review, but sponsors should validate those
measures rather than present unverified confidence values as safeguards.
CDRH should also incorporate longitudinal real-world outcome monitoring based on standardized
AI exposure and version data. These data would support evaluation across institutions and clinical
settings, comparisons across patient subgroups, detection of uncommon safety signals, analysis
of clinician interaction with AI outputs, and before-and-after assessment of new device or model
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 16 of 22
versions. Without reliable exposure and version identification, these analyses will remain
incomplete.
The monitoring cadence should be risk-based. Reassessment should also be triggered by model
updates, new data sources, changes in intended use, deployment in new populations or settings,
performance-threshold breaches, adverse events, cybersecurity incidents, and material workflow
changes. Findings should feed promptly into investigation, corrective action, controlled validation,
and change management. Common data models, interoperable reporting formats, and automated
capture should support aggregate learning while minimizing documentation burden.
CDRH should build on existing infrastructure rather than create a disconnected reporting channel.
Relevant pathways include eMDR and MAUDE as medical-device report data become available
through AEMS, MedWatch and MedSun, and NEST or Sentinel for active surveillance, as
appropriate. Because these systems serve different purposes, integration should use common
GenAI data elements for product and model version, exposure, clinical context, outcome, and
event severity. Spontaneous reports alone cannot estimate incidence; reliable rates require
exposure denominators and defined observation windows.
CDRH also should consider an FDA-governed regulatory sandbox or federated registry in which
manufacturers, health systems, clinicians, patients, and researchers can test common data
elements and monitoring methods and share deidentified or appropriately aggregated safety
signals. Governance should protect patient privacy and proprietary information, use role-based
access, define how signals are validated and escalated, and publish aggregate findings when
appropriate. Participation should supplement, not replace, mandatory reporting or manufacturer
accountability.
Question 20: Please comment on whether the proposed approaches to postmarket monitoring
can be facilitated by machine-based supervisory agents. What considerations, including the
evaluation and reliability of the supervisory agent itself, should CDRH take into account for
such an approach?
Machine-based supervisory agents may help CDRH and manufacturers monitor GenAI-enabled
devices by reviewing high volumes of interactions and detecting changes in utilization, output
patterns, clinician acceptance or override rates, subgroup performance, or outcomes after a model
update. Their usefulness will depend on standardized, reliable data that identify AI exposures,
versions, clinical context, and subsequent outcomes.
A supervisory agent should be evaluated as a safety-relevant system in its own right. Validation
should address its intended monitoring task, input-data quality, sensitivity to meaningful safety
signals, false-positive and false-negative rates, subgroup performance, drift, cybersecurity, and
performance under unusual or adversarial conditions. Monitoring plans should include audit logs,
transparent alert thresholds, human review and escalation procedures, periodic revalidation, and
change control for the supervisory agent.
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 17 of 22
Supervisory agents should facilitate, not replace, human monitoring programs, validated source
data, manufacturer investigation, clinical judgment, or FDA oversight. A supervisory agent should
flag and prioritize potential signals for accountable human review rather than make final safety or
regulatory determinations. It also cannot compensate for the inability to identify which device and
version was used or to link that exposure to clinical outcomes.
Question 21: What roles might clinicians, healthcare institutions, professional societies,
standards-setting bodies, and other stakeholders appropriately play in postmarket
monitoring, and how can these roles be structured without diffusing manufacturer
accountability?
Postmarket monitoring should use shared infrastructure and clearly assigned roles, but the
manufacturer must remain accountable for the device's safety and effectiveness. FDA should not
permit contracts to transfer the manufacturer's core monitoring, investigation, reporting, or
corrective-action responsibilities to clinicians or health systems.
An appropriate division of responsibility would include:
• Manufacturers should provide persistent identifiers for products, models, and versions and
remain responsible for monitoring, investigation, regulatory reporting, and corrective
action;
• Health care institutions and EHR vendors should oversee local implementation, integration,
and training; detect site-specific problems; and automatically capture standardized AI
exposure data;
• Clinicians and patients should be able to report safety and usability concerns through
streamlined, accessible mechanisms that do not create excessive administrative burden;
• Professional societies should help develop clinical competencies, benchmarks, reporting
standards, and education;
• Standards-development organizations, research networks, and registries should establish
and use common representations for AI products, versions, exposures, outputs,
provenance, and safety events; and
• HHS agencies should coordinate their authorities: ONC should examine certification
criteria, interoperability standards, and APIs; CMS should consider appropriate quality,
health IT, or payment-program mechanisms; and FDA should define the information
manufacturers must provide and how real-world data may support regulatory evidence.
Standardizing the underlying data, rather than mandating a single repository, would allow existing
EHR networks, registries, distributed research networks, and other real-world data infrastructures
to support multi-institutional evaluation. Epic Cosmos is one illustrative example, but the policy
should remain vendor-neutral. Linking AI exposure with the device, model, version, clinical context,
user interaction, subsequent care, and patient outcome could support subgroup analyses and
detection of uncommon safety signals without requiring a new centralized federal database.
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 18 of 22
This approach recognizes that the data infrastructure is an ecosystem responsibility while
preserving clear manufacturer accountability for each regulated device. Any shared registry or
sandbox should use role-based access, privacy and cybersecurity controls, protections for trade
secrets and confidential commercial information, transparent criteria for escalating safety signals,
and public reporting of appropriately aggregated findings. Participation should not replace
mandatory reporting obligations.
Question 22: CDRH envisions that a manufacturer’s premarket competency-based
assessment might serve as a baseline against which post-deployment modifications could be
re-evaluated. How might the extent of re-benchmarking or other evidence be scaled to the
nature and expected impact of a given modification? For example, are there categories of
change that might not significantly affect safety or effectiveness of a GenAI-enabled device, or
might be appropriately managed within a sponsor’s quality management system versus
requiring FDA premarket review and authorization before implementation? Are there
categories of changes that might be appropriate for inclusion in a PCCP?
CDRH should use a graduated change-control approach based on the likelihood that a
modification could alter clinical function, performance, user interaction, or risk. Nonmaterial
changes that do not affect clinical logic, model behavior, or validated interfaces may be managed
within the manufacturer's quality management system when supported by verification and a
documented rationale. Intermediate changes should receive targeted revalidation and any required
notification, while material changes should receive premarket review unless they fall within an
authorized Predetermined Change Control Plan.
Changes to model weights or architecture, training data, input sources, output generation,
safeguards, intended use, target population, autonomy, or connected tools should trigger targeted
re-benchmarking and, where appropriate, clinical confirmation. A manufacturer's characterization
of a change as an improvement should not determine the review pathway; the pathway should
depend on demonstrated effects on safety, effectiveness, intended use, and foreseeable failure
modes.
The premarket competency assessment should serve as the baseline for each comparison. Every
deployed version should have a persistent identifier, and standardized exposure data should
support before-and-after analysis of real-world performance and outcomes. A PCCP may be
appropriate for reasonably foreseeable categories of change when validation methods, acceptance
thresholds, monitoring requirements, and rollback procedures are prespecified.
Question 23: How might PCCP concepts or other change-control approaches be adapted for
GenAI-enabled devices when the nature or scope of future modifications cannot be fully
prespecified?
When exact modifications cannot be fully prespecified, a PCCP should define bounded categories
of permitted change and prespecify the process for assessing their effects. The plan should identify
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 19 of 22
excluded changes and require additional FDA review for modifications that alter intended use,
materially expand autonomy or risk, exceed defined performance limits, or otherwise fall outside
the authorized scope.
At a minimum, the PCCP should specify data sources, validation methods, clinical and subgroup
acceptance criteria, human-factors and cybersecurity checks, documentation requirements,
deployment controls, post-deployment monitoring, notification and reporting, and stop or rollback
criteria. Within an authorized PCCP, FDA should permit an expedited pathway for an update that
maintains or improves safety and effectiveness when the sponsor meets the prespecified criteria,
demonstrates that the update does not introduce a new material failure mode, assigns a persistent
version identifier, and maintains monitoring and rollback capability. Updates outside those
boundaries should follow the ordinarily applicable premarket review pathway.
Question 24: For devices built on third-party foundation models, changes to the underlying
model may be initiated by the third-party model developer rather than the device
manufacturer. How can a manufacturer detect, evaluate, and respond to such changes in a
timely manner? What mechanisms—for example, contractual, technical, or through a PCCP—
could provide reasonable assurance that third-party developer-initiated changes do not
compromise the safety or effectiveness of the device?
The device manufacturer should remain responsible for detecting, evaluating, and responding to
changes made by a third-party foundation-model developer. Contracts and technical controls
should provide:
• Advance notice of planned and emergency model changes;
• Persistent version identification and, where feasible, version pinning;
• Detailed release notes and agreed criteria for material changes;
• Access to evaluation results, known limitations, incident information, and relevant audit
records;
• Automated regression testing and risk-based validation before a changed model is
deployed in the device;
• Rapid suspension, rollback, and safe fallback capabilities; and
• Immediate notification of safety-related failures or vulnerabilities.
Exposure records should identify both the commercial device and the relevant underlying model or
configuration when technically feasible and relevant to safety or effectiveness. This provenance
would allow manufacturers and regulators to determine whether a change in real-world
performance coincided with a foundation-model update.
A PCCP may address reasonably foreseeable third-party updates, but it should not permit
automatic deployment solely because the third-party developer characterizes an update as an
improvement. The manufacturer should validate the complete device after any material change,
including its safeguards, workflow integration, and connected tools. An update within authorized
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 20 of 22
boundaries may use the expedited pathway described above; a material or out-of-scope change
should receive the ordinarily applicable review. Foundation Model Device Master Files could
support timely information exchange, but they should not replace contractual notice, technical
controls, or the manufacturer's responsibility for continued safety and effectiveness.
Section VII: Other Topics
Question 25: Would voluntary Foundation Model MAFs be practical for foundation model
developers to provide and to keep current, and what content might they include? Given that
participation would be voluntary and that model developers may have limited incentive to
disclose safety-relevant information, what would make such a program sufficiently useful for
premarket review? Are there alternative mechanisms CDRH should consider for obtaining
information about underlying foundation models?
AMIA supports Foundation Model Master Files because they could reduce duplicative submissions
and improve FDA's understanding of models used across multiple devices. However, a purely
voluntary program may provide incomplete information, particularly for safety-critical models.
Files should use standardized, machine-readable templates and include model architecture,
supported uses, training-data provenance, known limitations, failure modes, subgroup evaluations,
guardrails, version history, cybersecurity information, update-notification commitments, and auditlog capabilities. FDA should protect proprietary details while requiring an accessible public
summary of safety-relevant information.
Question 26: Are there additional considerations that inform the premarket and postmarket
evaluation of agentic GenAI-enabled devices, beyond those applicable to non agentic GenAIenabled devices? How should the elevated risk associated with autonomous multi-step
action, tool use, and reduced opportunity for human review be reflected in acceptance
criteria and oversight?
Agentic devices require heightened oversight because errors can compound across multiple steps
before a clinician can intervene. Evaluation should address:
• The device’s permissions and access to external systems;
• The safety of tool and API use;
• End-to-end task completion, rather than isolated outputs;
• Error propagation across sequential actions;
• Human authorization checkpoints;
• Stop, override, and rollback mechanisms;
• Failure containment and safe handoff;
• Immutable audit trails; and
• Performance under adversarial, incomplete, and unexpected conditions.
American Medical Informatics Association
6218 Georgia Avenue NW, Suite #1, PMB 3077, Washington, DC 20011
www.AMIA.org | 301.657.1291
Page 21 of 22
A nominal “human in the loop” should not be treated as an adequate safeguard unless the person
has sufficient time, information, authority, and expertise to meaningfully review and stop the
device’s actions. FDA should increase evidentiary and monitoring requirements as autonomy,
action complexity, and potential harm increase.
AMIA appreciates the opportunity to provide these comments and welcomes continued
engagement with CDRH. We encourage FDA to pursue a clear, risk-proportionate, and lifecyclebased approach that protects patients while supporting responsible innovation. Please contact
Tayler Williams, Senior Manager of Public Policy, at twilliams@amia.org with any questions.
Sincerely,
Eileen Koski
Chair, AMIA Public Policy Committee
American Medical Informatics Association