Stanford Institute for Human-Centered Artificial Intelligence (HAI)

Academia / otherAcademicFiled October 19, 20264,612 words · 1 attachmentFDA-2026-N-7874-0514
LinkedInX

Themes it raises

14 of the 21 themes in the docket, each with the passage we counted, verbatim.
What makes a function high riskFDA Q1, Q2, Q5
“The FDA should clarify how genAI devices with wide-ranging functions are placed on the consequences axis, including whether placement reflects the worst-case use, the typical use, or a distribution across types of users and settings.”
Whether the user can judge the outputFDA Q3, Q4
“For example, a radiologist can readily verify an AI-generated radiology report if they have access to the original images. Yet a patient will struggle to verify an AI-generated summary of their complex medical record.”
Judging devices the way clinicians are credentialedFDA Q7, Q8
“An AI system lacks that experience and the ability to move beyond the domains in which it was tested, so the test is effectively all there is.”
Whether benchmark results prove anythingFDA Q9, Q10, Q16
“A more basic problem is construct validity: benchmarks often do not measure what they claim to measure, and different benchmarks purporting to measure the same property frequently disagree.”
Proving the device works in real careFDA Q11, Q12, Q13, Q14, Q15
“For patient-facing tools, premarket evaluation should also test how patients interpret and are affected by genAI device outputs, as the FDA has done for direct-to-consumer genetic tests.”
Trading premarket certainty for postmarket monitoringFDA Q18
“Moving parts of the evidence burden into the postmarket period is consistent with the FDA’s “least burdensome” approach, but it is only defensible if postmarket commitments are specified up front rather than worked out after the device is in use.”
Watching the device after it shipsFDA Q19, Q20
“When the intended use is to deliver on a specified outcome, postmarket monitoring should not only focus on model output quality but also on downstream patient outcomes that health systems already capture”
Who is accountable when something goes wrongFDA Q21
“However, the responsibility of postmarket monitoring cannot rest on the healthcare institutions adopting the tools alone.”
Whether human oversight is real oversightFDA Q3, Q4, Q14, Q20, Q21, Q26
“Human oversight should be treated as a risk control whose effectiveness must be demonstrated for the task(s) the device performs, rather than always assumed to be helpful.”
Equity, access and under-resourced settingsFDA Q3, Q13, Q21
“But because their monitoring capacity varies widely, there currently exists an uneven self-regulatory landscape and, consequentially, disparate patient protection.”
How this fits rules that already existFDA Q8, Q9, Q16, Q25
“ISO 14971 instructs developers to include reasonably foreseeable misuse in the hazard analysis and validation phases.”
What counts as a reportable eventFDA Q19, Q20
“They should also provide technological mechanisms for frontline users to submit feedback on the AI output and for deploying organizations to receive reports of such feedback both within their organization and across all deployers.”
Escalating too little and too muchFDA Q6
“At Stanford Health Care, clinicians and patients who participate in AI assessments often note patient anxiety due to overescalation as a substantial risk for some types of AI tools.”
What the rules cost sponsors and the marketNot asked by the FDA
“if validation requirements are too demanding, clinical genAI tools could take years to reach authorization, by which point the underlying technology may be obsolete, and the process could become so burdensome that developers choose not to build these tools for clinical use at all”
Machine-assisted draft, pending human review. The source text and highlighted passages appear below. Read the filing on regulations.gov ↗

The comment as filed

Comment submitted on regulations.gov. Passages we counted are highlighted.

See attachment

Attachment

Attachment, text extracted from the filed document. Passages we counted are highlighted.

U.S. Food and Drug Administration
Center for Devices and Radiological Health
Document No. 2026-N-7874
Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper
and Request for Feedback

October 19, 2026

Introduction

We, a group of scholars affiliated with the Stanford Institute for Human-Centered Artificial
Intelligence and Stanford University School of Medicine, submit the following comments in
response to your request for feedback.

As practicing physicians, legal scholars, technologists, ethicists, and policy professionals
working to advance the governance of healthcare AI,1 we are heartened by the discussion paper,
and support the efforts presented to consider how best to regulate genAI-enabled medical
devices. We offer our reflections on the risk assessment, evaluation, and ongoing monitoring of
genAI devices based on our interdisciplinary academic research and our experiences operating a
responsible AI lifecycle governance system at Stanford Health Care.

We share considerations for the operationalization of the two-axis risk assessment framework
and our thoughts on how to ensure that the proposed competency-based approach to evaluation
and monitoring can better align with current developer practices and the science of AI
benchmarking, in an effort to both maximize patient safety and support innovation. While we do
not draw a firm line between general-purpose and purpose-built tools in our submission, the
distinction merits consideration.

We make 7 targeted recommendations:

On the Two-Axis Risk Framework
1.​ Require evidence that human oversight is an effective risk control for the task(s) the
device performs.
2.​ Specify how the consequences axis is determined and consider measures of typical use.
Stanford HAI’s Healthcare AI Policy Steering Committee conducts interdisciplinary, evidence-based research and
convenes multi-stakeholder discussions to develop tangible recommendations for policymakers that help ensure
healthcare AI can benefit patients, doctors, and developers alike. See: Stanford HAI, “Healthcare AI Policy,”
accessed October 14, 2026, https://hai.stanford.edu/policy/policy-efforts/healthcare-ai-policy.
On Premarket Evaluation
3.​ Strengthen benchmarking to address construct validity and prevent metric optimization at
the expense of utility.
4.​ Establish real-world prospective testing as a supplement or alternative to benchmarking.

On Postmarket Evaluation
5.​ Place greater accountability for postmarket monitoring on developers and equip
healthcare institutions with the ability to support continued monitoring.
6.​ Use hybrid monitoring, with machine-based monitoring tools to flag issues and humans
as the final arbiters.
7.​ Ensure that postmarket monitoring occurs by requiring a prespecified monitoring plan.

1. Feedback on the two-axis risk framework for genAI-enabled devices

The proposed two-axis framework for assessing the risks posed by genAI-enabled devices has
strong intuitive appeal. We particularly support the intent to assess a device’s directive functions
(i.e., its categorization as non-directive vs action-directing) based on how the tool is likely to be
used. We agree that developers may otherwise use strategic wording to avoid being placed in a
more stringent risk category.

The discussion paper also rightly recognizes that multi-turn conversations between humans and
genAI devices entail special risks. In addition to becoming more directive over the course of a
multi-turn conversation, evidence suggests that some chatbots engage in psychological
manipulation2 and create significant dependencies3 as they become more intimately familiar with
their user — even if they don’t become more “action-directing” in terms of their functions. This
can undermine the user’s likelihood of seeking to validate its advice through external means.
Manipulative chatbot behavior must therefore be a consideration in the risk framework.

Another strength of the framework is the recognition that over-escalation to care is a harm
worthy of consideration. At Stanford Health Care, clinicians and patients who participate in AI
assessments often note patient anxiety due to overescalation as a substantial risk for some types
of AI tools.
Our governance process for using AI safely, responsibly, and equitably views this as
a legitimate risk that needs to be weighed against the benefits of the tool.4
Nizan Geslevich Packin and Karni Chagal-Feferkorn, “This Is Not a Game: The Addictive Allure of Digital
Companions,” Seattle UL Rev. 48 (2025): 693, https://digitalcommons.law.seattleu.edu/sulr/vol48/iss3/6/.
Myra Cheng et al., “Sycophantic AI decreases prosocial intentions and promotes dependence,” Science 391, no.
6792 (2026): eaec8352, https://www.science.org/doi/10.1126/science.aec8352.
Stanford Health Care, “Our Commitment to Using AI Safely, Responsibly, and Equitably,” accessed October 14,
2026, https://stanfordhealthcare.org/campaigns/ai-education/responsible-use.html.
We identify several key challenges when it comes to operationalizing the two-axis framework:

The risk distinction between advisory and action-taking devices will only hold if human
reviewers are consistently able to meaningfully supervise AI outputs. The activity axis
assumes they can but in practice, this is not always true. The axis spans a broad range of
heterogeneous tasks, and how well a human can check an output depends on the task and the
reviewer. For example, a radiologist can readily verify an AI-generated radiology report if they
have access to the original images. Yet a patient will struggle to verify an AI-generated summary
of their complex medical record.
Individuals also differ in their skepticism: some will triangulate
an AI output with other sources, while others will accept it uncritically.

Oversight also varies with setting and erodes over time. Physicians in one clinical setting may
have the time, resources, expertise, and incentives to scrutinize AI output carefully, while
physicians in another may not. In an emergency care setting, for example, AI output related to a
specialist domain may only receive a short glance from an emergency medicine physician.
Active supervision is also likely to fade over time as users become more familiar and possibly
develop automation complacency. Further, the potential for automation bias is high in many
areas of healthcare, where clinicians work under time pressure and manage workloads that create
chronic exhaustion and burnout.

The framework also doesn’t specify how placement on the consequence axis will be
determined for multi-purpose AI tools. GenAI devices — especially patient-facing tools with
general purpose capabilities — can field thousands of different questions or prompts and thus
have many different potential uses that go beyond the primary intended use. Consequently, the
resulting outcomes of one AI tool may range from trivial to highly consequential. Will the
determination of a tool’s consequence (from “limited” to “severe”) be based on the most severe
possible outcome or some measure of the typical outcome? How will the framework take into
account the reality that the consequences of an AI tool may vary significantly across patient
populations, including clinical subgroups and groups underrepresented in the data used to build
the genAI device? Lastly, how will the framework take into account the fact that the risks
associated with general purpose genAI devices can change over time as use changes?

Recommendation #1. Require evidence that human oversight is an effective risk control for
the task(s) the device performs.

Human oversight should be treated as a risk control whose effectiveness must be demonstrated
for the task(s) the device performs, rather than always assumed to be helpful.
Where a developer
relies on human oversight to reduce the risk of its genAI device, the FDA should require the
developer to provide evidence that oversight will consistently and meaningfully occur in the
intended settings and that reviewers can detect the errors the device is likely to make.
Such evidence could include real-world testing data on how often, how long, and in what ways
AI tool users engage in activities that could be considered supervision or validation of AI output
— and how this varies across different users in different clinical settings. For example, how long
do healthcare professionals (HCPs) spend examining an output before acting on it? How often do
they follow or reject the tool’s recommendation? How often do they click through to source
material, where provided? Do they have access to the necessary information to detect errors?
How do the answers to these questions differ between HCP-facing and patient-facing AI tools?

Recommendation #2. Specify how the consequences axis is determined and consider
measures of typical use.

The FDA should clarify how genAI devices with wide-ranging functions are placed on the
consequences axis, including whether placement reflects the worst-case use, the typical use, or a
distribution across types of users and settings.
We expand on this recommendation in our
feedback on how the FDA should specifically approach premarket and postmarket monitoring
for genAI devices in the following pages.

The framework must also take into account and weigh appropriately harms that are harder to see
or measure when determining the severity of a tool’s consequences. Psychological harms, for
example, have traditionally often been viewed as lower-risk than other, physical harms,5 so a
genAI device that engages with users in distress could be determined to have “limited”
consequences even though the potential for harm is significant. Frameworks such as the FDA's
fair-balance requirements for prescription drug promotion, which state that important risk
information should not be minimized in comparison to efficacy claims, could be applied here.6

2. Feedback on the competency-based approach for premarket evaluation

Given that AI is not indication-based, and given the vast range of potential functions that many
genAI devices may perform, the competency-based approach presents an attractive alternative to
a more burdensome function-by-function review. However, we highlight several concerns about
the effectiveness of a competency-based approach that leans heavily on benchmarks.
Julian Herpertz et al., “FDA-Authorized Software as a Medical Device in Mental Health: A Perspective on
Evidence, Device Lineage, and Regulatory Challenges,” NPJ Mental Health Research 4 (2025): 62,
https://doi.org/10.1038/s44184-025-00174-2.
Hyosun Kim, “Trouble Spots in Online Direct-to-Consumer Prescription Drug Promotion: A Content Analysis of
FDA Warning Letters,” International Journal of Health Policy and Management 4, no. 12 (2015): 813-821,
https://doi.org/10.15171/ijhpm.2015.157.
First, intended use alone may be an incomplete basis for evaluating multi-purpose genAI
devices. The same device can perform functions related to hypertension, diabetes, and heart
failure — and it is challenging to decide what its core competencies should be. When evaluation
is anchored to the product’s stated intended use, a tool may be authorized for a narrow purpose
while being deployed by its users in many other ways, and the evidence generated for the stated
use may say little about the risks of those other uses. This also complicates two parts of the
framework: what developers must evaluate before and after deployment, and where the FDA
places a device on the consequence axis.

One way to address this gap without function-by-function review is through the risk management
process developers already follow. ISO 14971 instructs developers to include reasonably
foreseeable misuse in the hazard analysis and validation phases.
The FDA does not require
developers to follow this standard, but it references it as the gold standard and requires
developers to document a risk management process under the Quality Management System
Regulation (QMSR). The FDA could therefore provide recommendations in guidance documents
on intended use and reasonably foreseeable misuse beyond the stated intended use, and keep that
scoping current as real-world use cases emerge.

Benchmark validity is also a significant source of uncertainty in the proposed
competency-based approach. The discussion paper acknowledges that public benchmarks may
suffer from data contamination, saturation, and limited representativeness of real-world
conditions. A more basic problem is construct validity: benchmarks often do not measure what
they claim to measure, and different benchmarks purporting to measure the same property
frequently disagree.
7,8 Consider a benchmark question created to measure bias that deliberately
gives a model incomplete information, where the correct answer is that there is not enough
information to answer. A model can score well simply by recognizing the trick, regardless of its
underlying biases.

The problem is especially acute in medicine, where many of the skills we want to measure are
inherently hard to measure. A competency-based approach risks resembling medical
examinations with deterministic answers, which may not reflect real medical tasks.9 Performance
may also not carry over when the format changes: models that perform well on multiple-choice
vignettes can fail when the same case is presented as a multi-turn conversation.10
Max Zhang et al., “Why Do Safety Guardrails Degrade Across Languages?,” arXiv preprint arXiv:2605.17173
(2026), https://arxiv.org/abs/2605.17173.
Olawale Salaudeen, “Validating Claims About AI: A Policymaker’s Guide,” Stanford Institute for
Human-Centered Artificial Intelligence, September 24, 2025,
https://hai.stanford.edu/policy/validating-claims-about-ai-a-policymakers-guide.
Inioluwa Deborah Raji et al., “It’s Time to Bench the Medical Exam Benchmark,” NEJM AI 2, no. 2 (2025):
AIe2401235, https://doi.org/10.1056/AIe2401235.
Shrutika Singh et al., “The Pitfalls of Multiple-Choice Questions in Generative AI and Medical Education,”
Scientific Reports 15 (2025): 42096, https://doi.org/10.1038/s41598-025-26036-7.
Weak construct validity may result in measuring convenient outcomes, rather than the
most appropriate ones. As the bias benchmark example above shows, an AI model can pass a
test by recognizing the structure of the test, without delivering what is needed for good
outcomes. The results can still be used by a developer to support the claim that the model works.
There are indications that this is already happening in the mental health context. For example,
the competencies chosen for evaluation within the field (e.g., warmness of response) tend to be
the easiest to measure, but these often miss the mark of what is needed in the real world in order
to produce a good outcome (e.g., appropriate response to suicidality).11,12

Raising the bar for premarket evidence, however, carries its own tension: if validation
requirements are too demanding, clinical genAI tools could take years to reach authorization, by
which point the underlying technology may be obsolete, and the process could become so
burdensome that developers choose not to build these tools for clinical use at all
.

Finally, the clinician credentialing model underlying the competency-based approach has
limits of its own that are likely to be magnified when applied to AI.13 In medicine,
competency assessments are informative because test performance tracks other traits, such as
intelligence, diligence, and professionalism. But the assessment does not capture the years of
training and experience that support a clinician’s practical judgment (phronesis) real-world care
requires. An AI system lacks that experience and the ability to move beyond the domains in
which it was tested, so the test is effectively all there is.
Competency assessment may therefore
offer weaker assurance for AI than it does for physicians.

Recommendation #3: Strengthen benchmarking to address construct validity and prevent
metric optimization at the expense of utility.

Benchmarks are useful evidence only if they measure what matters. Developers should be
required to provide evidence of construct validity, showing that the competency being tested
corresponds to the skill needed for the real task and for producing good outcomes. The
discussion paper already notes that adjudicators should be structurally independent from the
sponsor, and we encourage the FDA to apply the same independence expectation to the process
of selecting benchmark criteria, especially when sponsors propose their own. As mentioned
above, developing benchmarks with high construct validity is expensive. To address this, the
Matthew Flathers et al., “Benchmarking Language Models for Clinical Safety: A Primer for Mental Health
Professionals,” Psychotherapy and Psychosomatics (2026), https://doi.org/10.1159/000553576.
John Torous and Andrea Cipriani, “A Paradigm Shift in Progress: Generative AI’s Evolving Role in Mental
Health Care,” JMIR Mental Health 12 (2025): e82369, https://doi.org/10.2196/82369.
MJMH (Kiki) Lombarts, “Competence-based education misses the essence of the medical profession,”
Perspectives on Medical Education 4, no. 6 (2015): 326-328, https://pmc.ncbi.nlm.nih.gov/articles/PMC4673067/.
FDA could utilize existing processes of guidance documents and comment periods or explore
further mechanisms to support the development of more robust vendor-neutral benchmarks.

To prevent these issues, benchmarks should be pre-registered before testing. This complements
the prespecification the discussion paper already expects by preventing developers from
choosing favorable benchmarks or adjusting criteria after seeing results. However, the tools used
to run these evaluations need the same scrutiny. LLM judges should be validated against a
human expert standard, since research has found that they tend to favor their own outputs, and
those of similar models, over others of comparable quality.14 Patient simulators should count as
evidence only if sponsors can show that simulated users behave like real ones, for example in
how often they accept, question, or push back on outputs, by comparing against a left-out sample
of real interactions. They must also be grounded in real-world data and varied on the traits that
determine whether a user can catch an error, such as health literacy, baseline trust or skepticism
toward AI, and persistence when a device declines to answer. Simulator results lacking this
validation should be labeled exploratory and not pooled with clinical confirmation evidence.

Recommendation #4: Establish real-world prospective testing as a supplement or
alternative to benchmarking.

In contrast to competency exams, real-world usage would give data on performance in the
intended use scenario. Sandboxes, such as Utah’s AI Sandbox,15 may provide the opportunity to
collect real-world data in a monitored setting. This data could be used for submitting premarket
performance to the FDA. It may be beneficial for the FDA to indicate what scope of premarket
data would be acceptable from such a sandbox. Other methods of generating real-world
premarket data in the intended use scenario should also be considered.

For patient-facing tools, premarket evaluation should also test how patients interpret and are
affected by genAI device outputs, as the FDA has done for direct-to-consumer genetic tests.
16

3. Feedback on postmarket monitoring approaches

Postmarket monitoring is one of the most important and challenging aspects of regulating genAI
devices used in healthcare contexts. As already mentioned in Section 1 of our submission, the
range of possible prompts a genAI tool may receive is effectively unbounded, performance can
drift over time, and human oversight is fallible. While performance on predefined tasks can be
Roxana Daneshjou et al., “The Judge is Not Impartial: Self-Preference in Medical LLM Evaluation,” preprint,
Research Square, September 17, 2026, https://doi.org/10.21203/rs.3.rs-11022502/v1.
Utah Office of Artificial Intelligence Policy, “Understanding Utah’s AI Sandbox,” Utah Department of
Commerce, accessed October 19, 2026, https://commerce.utah.gov/ai/regulatory-relief/.
Scott D. Crawford et al., “Decoding FDA DTC Policy: The Complete Series,” SoundRocket (blog), November
2018, https://www.genomes2people.org/wp-content/uploads/2018/11/20181112_sound-rocket.pdf.
assessed during premarket evaluation, real-world safety signals and harms are more difficult to
surface during limited pilot runs, especially when there are harms that may only appear after the
patient leaves their care setting.

However, the responsibility of postmarket monitoring cannot rest on the healthcare institutions
adopting the tools alone.
Technical capacity and staffing across U.S. healthcare organizations
varies widely. Some institutions, including our own, can perform sophisticated in-house
monitoring of how genAI tools are used and what problems HCPs are reporting. Yet even those
institutions are finding that the volume of deployed genAI tools is increasingly outstripping their
ability to monitor them and that assessing the impacts of AI on patient outcomes is quite
challenging.17

Institutions that don’t have the resources or technical capacity to perform their own monitoring
must rely wholly on genAI tool developers to provide them with monitoring information.
However, developers often do not provide sufficient monitoring support to healthcare
institutions. For example, they frequently omit information about the most important failure
modes of their products and suggest only limited monitoring metrics that do not capture the
clinical utility or risks of the tool. Further, shared information may be in the form of dashboards
that do not enable healthcare institutions to conduct their own data analysis.

Recommendation #5. Place greater accountability for postmarket monitoring on developers
and equip healthcare institutions with the ability to support continued monitoring.

Developers should assume a greater role in ensuring the safety and effectiveness of genAI
devices across the total product lifecycle. Existing device requirements (including QMSR,
Medical Device Reporting, and special controls) offer a starting point for conceptualizing
obligations for developers.

Healthcare organizations play a critical role in local evaluation, particularly in surfacing
problems that arise from how AI tools are placed into specific workflows. But because their
monitoring capacity varies widely, there currently exists an uneven self-regulatory landscape
and, consequentially, disparate patient protection.
AI developers should be required to give
deployers usable data access, documentation of known failure modes (with updates as
knowledge evolves), and suggested monitoring metrics that help assess the product’s clinical
utility. They should also provide technological mechanisms for frontline users to submit
feedback on the AI output and for deploying organizations to receive reports of such feedback
both within their organization and across all deployers.
This information would help healthcare
organizations conduct monitoring, identify areas where local workflows could be changed to
Zhongnan Fang et al., “Operationalizing Real-Time Monitoring of Clinical AI,” Stanford Institute for
Human-Centered Artificial Intelligence, May 14, 2026,
https://hai.stanford.edu/policy/operationalizing-real-time-monitoring-of-clinical-ai.
help catch and respond to AI errors before they become consequential, signal when local users
find fault with the AI device so often that discontinuation of the tool should be considered, and
identify specific problems that deployers train users to look out for when using the tool.

Additionally, monitoring should ideally aim to go beyond model outputs to include patient
outcomes and changes in use over time. When the intended use is to deliver on a specified
outcome, postmarket monitoring should not only focus on model output quality but also on
downstream patient outcomes that health systems already capture
, such as unplanned return
visits, escalations in level of care, and re-admissions after discharge. Because harms from
erroneous output often appear after discharge, single-site monitoring will often undercount them:
roughly a third of return visits within three days of an emergency department discharge occur at
a different hospital.18 Thus, these outcomes should be aggregated across sites through a common
data model. The willingness of health systems that deploy the devices and electronic health
record (EHR) vendors to provide developers with access to this real-world deployment and
monitoring data will, in turn, greatly impact developers’ ability to conduct safety monitoring.
Requiring this data to be made available to developers may go beyond the FDA’s authority and
will likely require action from other regulatory bodies, or programs linking patient outcomes
with payment incentives.19

Recommendation #6. Use hybrid monitoring, with machine-based monitoring tools to flag
issues and humans as the final arbiters.

Machine-based screening tools may be needed for comprehensive postmarket monitoring, and,
when appropriate, they should flag issues for human review rather than act on their own.
Because monitoring of genAI is too onerous for many institutions to perform locally, the FDA
should support lightweight tools, qualified through its Medical Device Development Tool
(MDDT) program and hosted by a neutral party, that can run across health systems to detect
outliers. When clusters of mistakes are substantiated on review, the finding should be reported to
the FDA and the developer. Any automated monitoring should itself be validated against human
performance in a way that external auditors can reproduce.

As mentioned on p.2 of this submission, human review is not a safeguard by default. Until fully
automated drift detection matures, reviewers’ agree/disagree adjudications can serve as an
interim proxy for drift, with periodic manual review for high-stakes uses.
Rishi Duseja et al., “Revisit Rates and Associated Costs after an Emergency Department Encounter: A Multistate
Analysis,” Annals of Internal Medicine 162 (2015): 750-756. https://doi.org/10.7326/m14-1616.
Centers for Medicare & Medicaid Services, “ACCESS (Advancing Chronic Care with Effective, Scalable
Solutions) Model,” CMS Innovation Center, accessed October 19, 2026,
https://www.cms.gov/priorities/innovation/innovation-models/access.
Recommendation #7. Ensure that postmarket monitoring occurs by requiring a
prespecified monitoring plan.

We see no alternative to placing heavy emphasis on postmarket monitoring to assure the safety of
genAI tools because of their dynamic, evolving, and multifunctional nature. In doing so, the
FDA should require a postmarket monitoring plan, with defined endpoints and thresholds, to be
in place before authorization. Moving parts of the evidence burden into the postmarket period is
consistent with the FDA’s “least burdensome” approach, but it is only defensible if postmarket
commitments are specified up front rather than worked out after the device is in use.
Many
health systems already work this way: At SHC, each adopted AI tool has a monitoring plan,
designed jointly by the proposing team and the health system’s Data Science Team, that
complements a robust assessment conducted prior to adoption. The FDA can require developers
to do the same.

The FDA must also be able to ensure that such a plan is carried out. In the drug space, the FDA
has struggled to ensure that postmarket surveillance plans are actually followed by
manufacturers, and its legal authorities and levers for enforcing them are limited.20 For genAI
devices, one way to strengthen verification is to require that safety assessment and quality
reporting be designed into the product itself, so that data collection, performance tracking, and
feedback channels are built-in features rather than separate commitments a developer must
remember to honor. This would give the FDA something concrete to verify. We therefore urge
the FDA to consider what mechanisms are available to confirm that postmarket monitoring plans
are actually implemented, and to take action if they are either not fully implemented or reveal
problems that substantially alter the tool’s risk/benefit balance.

***

Overall, the regulatory approach outlined in the discussion paper is on the right track. The
two-axis risk assessment framework and competency-based approach to evaluation are
reasonable starting points for a technology that does not fit conventional device review. Their
value will depend on how they are operationalized, especially with regards to determining the
effectiveness of human oversight and benchmarking as part of monitoring strategies for genAI
devices that have a myriad of potential uses and outcomes.

Postmarket monitoring is where this framework will succeed or fail. Because genAI tools are
dynamic and multifunctional, the framework will lean heavily on what happens after
authorization. Monitoring plans with defined endpoints and thresholds must be in place before
authorization, and the FDA should consider how to verify implementation and act on these
Daniel G. Dauner et al., “FDA’s Unimproved Enforcement of Postmarketing Requirements and Commitments:
Implications for Providers and Patients,” Research in Social and Administrative Pharmacy 16, no. 6 (2020):
844-847, https://doi.org/10.1016/j.sapharm.2019.11.004.
findings. Success also depends on shared responsibility across developers, deployers, and the
FDA, and their willingness to share usable data, documentation of known failure modes, and
feedback mechanisms. Finally, humans, not automated tools, should remain the final arbiters of
what monitoring flags.

We appreciate the opportunity to work collaboratively with the FDA to arrive at frameworks that
support innovation and patient access to useful tools, while ensuring that developers, clinicians,
and healthcare systems face reasonable expectations when it comes to ensuring new genAI
devices are safe and effective.

Sincerely,

Roxana Daneshjou, MD, PhD, FAAD​ David Magnus, PhD
Assistant Professor of Biomedical Data Thomas A. Raffin Professor of Medicine
Science and Dermatology and, by courtesy, and Biomedical Ethics, Pediatrics, and
Radiology, Stanford University School of Medicine, and, by courtesy, of
Medicine Bioengineering, Stanford University

Dev Dash, MD, MPH Josh Makower, MD​
Clinical Assistant Professor of Emergency Co-Founder and Director, Stanford
Medicine, Stanford University School of Mussallem Center for Biodesign; Professor
Medicine of Medicine and Bioengineering, Stanford
University Schools of Medicine and
Diliana Funtarova Engineering
Master’s Candidate, Clinical Informatics
Management, Stanford University Nicole Martinez-Martin, JD, PhD
Assistant Professor of Pediatrics and, by
Curtis Langlotz, MD, PhD courtesy, of Psychiatry and Behavioral
Professor of Radiology, Medicine, and Sciences, Stanford University
Biomedical Data Science, Stanford
University School of Medicine; Senior Caroline Meinhardt
Fellow, Institute for Human-Centered Senior Policy Research Manager, Stanford
Artificial Intelligence (HAI); Senior HAI
Associate Vice Provost for Research;
Stanford University Michelle Mello, JD, PhD
Professor of Health Policy, Stanford
Kara Liu University School of Medicine; Professor of
PhD Candidate, Computer Science, Stanford Law, Stanford Law School
University School of Engineering
Jared Moore
PhD Candidate, Computer Science, Stanford
University School of Engineering

Sandra Waugh Ruggles, PhD
Director of Policy Research, Stanford
Mussallem Center for Biodesign

Kevin Schulman, MD, PhD
Professor of Medicine, Stanford University
School of Medicine; Professor, by courtesy,
of Operations, Information and Technology,
Stanford Graduate School of Business

Nigam Shah, MBBS, PhD
Professor of Medicine, and Biomedical Data
Science, Stanford University School of
Medicine

Artem Trotsyuk, PhD
Postdoctoral Research Fellow, Department
of Genetics, Stanford University School of
Medicine

Caroline Yee
Research Scholar, Stanford HAI