Radiological Society of North America (Carolyn C. Meltzer)

CliniciansClinicianFiled October 19, 20262,859 words · 1 attachmentFDA-2026-N-7874-0432
LinkedInX

Themes it raises

15 of the 21 themes in the docket, each with the passage we counted, verbatim.
What makes a function high riskFDA Q1, Q2, Q5
“The framework should explicitly account for reviewability, action reversibility, time pressure, availability of downstream safeguards, output traceability, intended-user expertise, and the breadth and variability of the device’s functions.”
Whether the user can judge the outputFDA Q3, Q4
“Risk mitigation should instead reflect the user’s ability to understand limitations and independently evaluate the output.”
Judging devices the way clinicians are credentialedFDA Q7, Q8
“This approach is well suited to systems with open-ended inputs and outputs, but it should supplement, not replace, application-specific analytical and clinical validation.”
Whether benchmark results prove anythingFDA Q9, Q10, Q16
“Public benchmarks are useful but may be affected by data contamination, saturation, and limited representativeness.”
Proving the device works in real careFDA Q11, Q12, Q13, Q14, Q15
“Higher-risk, action-directing, autonomous, or poorly reviewable functions should require stronger evidence, including prospective evaluation when appropriate.”
Trading premarket certainty for postmarket monitoringFDA Q18
“RSNA strongly supports robust postmarket monitoring but cautions against using postmarket monitoring as a general substitute for adequate premarket evidence.”
Watching the device after it shipsFDA Q19, Q20
“Periodic re-benchmarking, sample-based clinician review, and performance-degradation monitoring are complementary and should be combined where appropriate.”
Who is accountable when something goes wrongFDA Q21
“Clinicians should not become the default bearers of liability for errors arising from AI systems they cannot reasonably oversee or control.”
Controlling a device that keeps changingFDA Q22, Q23, Q24, Q25
“For devices that rely on third-party foundation models, the device manufacturer must remain responsible for detecting, evaluating, and responding to upstream changes.”
Devices that plan and take actionsFDA Q26
“High-consequence or irreversible actions should include validated human-authorization checkpoints unless the sponsor provides compelling evidence that autonomous operation offers a favorable benefit-risk profile.”
Whether human oversight is real oversightFDA Q3, Q4, Q14, Q20, Q21, Q26
“FDA should consider the qualifications of the intended end user when evaluating whether human oversight provides a meaningful safeguard.”
Records that let investigators reconstruct an eventFDA Q19, Q21, Q24, Q26
“Systems should maintain sufficiently detailed, reviewable audit logs of prompts, retrieved information, actions, tool outputs, failures, overrides, and human interventions to support clinical oversight, incident investigation, and accountability.”
Equity, access and under-resourced settingsFDA Q3, Q13, Q21
“Less resourced and rural institutions may be unable to build independent monitoring systems, creating uneven safety oversight.”
How this fits rules that already existFDA Q8, Q9, Q16, Q25
“These efforts should build upon existing interoperability standards, including DICOM, FHIR, and established common data element frameworks, wherever appropriate.”
What counts as a reportable eventFDA Q19, Q20
“FDA should establish a dual-reporting pathway that allows manufacturers and healthcare organizations or clinicians to report significant performance deviations directly to the agency, especially when vendor response is delayed or inadequate.”
Machine-assisted draft, pending human review. The source text and highlighted passages appear below. Read the filing on regulations.gov ↗

The comment as filed

Comment submitted on regulations.gov. Passages we counted are highlighted.

See attached file(s)

Attachment

Attachment, text extracted from the filed document. Passages we counted are highlighted.

October 19, 2026

Re: Docket Number FDA-2026-N-7874; Considerations for the Regulation of Generative AIEnabled Medical Devices: Discussion Paper and Request for Feedback

The Radiological Society of North America (RSNA) appreciates the opportunity to provide input on the
Food and Drug Administration (FDA) Center for Devices and Radiological Health discussion paper on
generative artificial intelligence (GenAI)-enabled medical devices [Docket Number FDA-2026-N-7874].

RSNA is a leading global organization representing more than 53,000 radiologists and medical imaging
professionals across 150 countries. Radiology is among the most data-intensive fields in medicine and
has extensive experience evaluating, implementing, and monitoring AI-enabled technologies in clinical
practice.

RSNA supports FDA’s effort to develop a nimble, risk-based, total product lifecycle approach for GenAIenabled devices. These technologies may improve the quality, timeliness, and accessibility of care, but
their stochastic outputs, broad task capabilities, dependence on foundation models, and potential to
evolve after deployment create risks that cannot be addressed through one-time premarket review
alone. Patient safety should remain the central organizing principle, and regulatory flexibility should be
paired with clear, enforceable safeguards that promote transparency, accountability, clinical validation,
and continuous real-world monitoring.

Our comments focus on six priorities: (1) broadening the proposed risk framework; (2) pairing
competency-based benchmarking with clinical confirmation; (3) establishing minimum, risk-based
continuous monitoring expectations; (4) assigning complementary responsibilities without diffusing
manufacturer accountability; (5) strengthening change control, version transparency, and foundationmodel governance; and (6) ensuring scalable implementation across healthcare settings.

I. Summary of Recommendations
• FDA should retain the proposed activity-by-consequence framework as a useful starting point, but
it should not be the sole basis for risk classification. Risk also depends on the user’s ability to
independently review an output, the reversibility and time sensitivity of resulting actions, the
availability of downstream safeguards, the breadth of the intended use, and the traceability of
outputs to source information.
• A competency-based approach can complement existing device evaluation, provided it assesses the
final deployed configuration and is paired with clinical confirmation that reflects the intended
population, workflow, users, modalities, and clinical setting. Benchmark performance alone is
insufficient to establish safe and effective real-world use.
• FDA should not accept materially greater premarket uncertainty for high-consequence,
autonomous, action-taking, or poorly reviewable functions. Where a limited shift toward
postmarket evidence is permitted, it should depend on an enforceable monitoring plan, validated
detection thresholds, timely escalation and corrective-action requirements, transparent versioning,
and FDA access to meaningful real-world evidence.
• Continuous monitoring should evaluate the clinical AI system, not the model alone. This includes
model performance, subgroup performance, human-AI interaction, workflow effects, appropriate
use, model and system changes, and cumulative effects when multiple AI tools operate in the same
environment.
• Manufacturers must remain accountable for device safety and effectiveness. Healthcare
organizations, clinicians, professional societies, and standards bodies have complementary roles
based on what each can observe and control, but those roles should not substitute for
manufacturer obligations.
• FDA should strengthen Predetermined Change Control Plans (PCCPs) for GenAI-enabled devices,
require timely and detailed communication of changes to models, prompts, and system
configurations, and establish mechanisms for end users to report performance deviations and
request review of change-control plans based on real-world experience. Manufacturers should
provide sufficient transparency to support safe, validated customization for local patient
populations and clinical workflows while protecting proprietary intellectual property.

II. Risk Assessment Should Reflect Clinical Context and Reviewability
RSNA agrees that the degree of device activity and the consequence of relying on an incorrect output
are central dimensions of risk. However, a two-axis framework may understate risk where outputs
appear informational but are difficult for users to independently evaluate, are presented with high
fluency or authority, or influence downstream decisions indirectly. FDA should therefore apply the
framework as an organizing heuristic within a broader, multidimensional assessment.

The framework should explicitly account for reviewability, action reversibility, time pressure,
availability of downstream safeguards, output traceability, intended-user expertise, and the breadth
and variability of the device’s functions.
A radiology output that cannot be independently verified
without specialist expertise may warrant greater oversight even if it is labeled non-directive. Similarly,
an output used in an urgent stroke, trauma, or oncology workflow may pose greater risk because the
opportunity to detect and correct an error is limited.

When clinical oversight is relied upon to mitigate risk, the intended end user should have the
qualifications and training necessary to independently perform the equivalent clinical task and assess
the validity of the AI output. FDA should consider the qualifications of the intended end user when
evaluating whether human oversight provides a meaningful safeguard.

FDA should assess directiveness based on the substance, specificity, personalization, presentation, and
foreseeable effect of an output, rather than terminology alone. Repeated suggestions, ranked options,
confident recommendations, workflow prioritization, or default selections may influence action even
when the device avoids words such as “recommend.” For multi-turn systems, intended use and risk
should be evaluated across realistic conversational trajectories, including foreseeable migration from
information retrieval to recommendation or action.

FDA should require manufacturers to establish safeguards that identify when user requests or system
behavior move beyond a device's validated indications for use. These safeguards should provide clear
warnings about the limitations of the device's validated performance without unnecessarily restricting
appropriate clinical judgment. For conversational systems, safeguards should operate across multi-turn
interactions rather than evaluating individual prompts in isolation. Manufacturers should validate
these safeguards and provide auditable records of detected deviations, warnings, and subsequent
system actions.

Patient-facing and generalist-facing functions should not automatically be treated as unsafe, because
they may improve access and engagement. Risk mitigation should instead reflect the user’s ability to
understand limitations and independently evaluate the output.
Safeguards may include plain-language
explanations, traceable source information, calibrated uncertainty, explicit limits on scope, clear
escalation pathways, specialist review for higher-risk use, and testing across health literacy levels and
communication styles.

III. Competency-Based Evaluation Must Be Coupled with Clinical Confirmation
RSNA supports further development of a competency-based approach that combines structured device
benchmarking with clinical confirmation. This approach is well suited to systems with open-ended
inputs and outputs, but it should supplement, not replace, application-specific analytical and clinical
validation.
Evaluation should address the final user-facing device in its deployed or representative
configuration, including prompts, retrieval components, guardrails, orchestration logic, user interface,
and relevant third-party dependencies.

Benchmarking should be risk-based and multidimensional. In addition to clinical knowledge and task
fidelity, FDA should require evaluation of reproducibility, stochastic variation, hallucination and
confabulation risk, boundary adherence, calibration and uncertainty communication, subgroup
performance, robustness to distribution shift, workflow integration, automation bias, usability, and
foreseeable failure modes. For radiology applications, testing should reflect variation in modalities,
manufacturers, scanners, imaging protocols, acquisition conditions, patient populations, disease
prevalence, and clinical practice settings.

Public benchmarks are useful but may be affected by data contamination, saturation, and limited
representativeness.
Sponsors should demonstrate construct validity and explain how benchmark
performance predicts safe and effective behavior for the intended use. FDA should encourage
sequestered datasets, independent expert adjudication, multi-institutional testing, predefined
acceptance criteria, and transparent reporting of dataset composition and limitations. Sponsordeveloped benchmarks may be appropriate for specialized functions but should be subject to
independent review and safeguards against optimization to the test.

Evidence requirements for GenAI-enabled devices should reflect the differences between AI and
traditional therapeutic interventions. For many diagnostic and workflow applications, clinical benefit
depends not only on device performance but also on how clinicians interpret and act on AI outputs.
Moreover, performance may vary across patient populations, imaging equipment, and clinical settings,
limiting the generalizability of a single clinical trial. FDA should therefore support flexible, riskcalibrated evidentiary approaches that incorporate real-world evidence, multi-institutional validation,
and prospective clinical studies as appropriate to the intended use and potential consequences of
error.

Clinical confirmation should be proportionate to risk. Lower-risk assistive functions may be supported
by retrospective multi-site studies or silent prospective evaluation. Higher-risk, action-directing,
autonomous, or poorly reviewable functions should require stronger evidence, including prospective
evaluation when appropriate.
FDA should assess both device-alone performance and human-AI team
performance when the intended use depends on clinician interaction. A technically accurate system
may still create harm through poor workflow integration, alert fatigue, automation bias, or unclear
communication.
IV. Continuous Monitoring Should Be Standardized, Actionable, and System-Focused
RSNA strongly supports robust postmarket monitoring but cautions against using postmarket
monitoring as a general substitute for adequate premarket evidence.
Greater premarket uncertainty
may be acceptable only for lower-risk, reviewable functions when the expected benefits are
meaningful and a mature, enforceable monitoring infrastructure is in place. It should not be accepted
for high-consequence, autonomous, action-taking, or difficult-to-review functions where an error may
cause serious or irreversible harm before it can be detected.

Continuous monitoring should be an ongoing lifecycle process that can detect signals, review their
clinical significance, intervene when appropriate, and reassess after intervention. It should evaluate
the full clinical AI system, including local model performance, human-AI interaction, patterns of
appropriate and inappropriate reliance, workflow effects, intended use, off-label use, subgroup
performance, model and system versions, and changes over time.

FDA should establish minimum manufacturer expectations that are adaptable to clinical risk,
technological characteristics, and degree of reviewability. Monitoring plans should identify measures,
data sources, cadence, thresholds, triggering events, responsible parties, escalation pathways, and
actions that may follow a significant signal. Triggering events should include model or software
updates, changes to an underlying foundation model, changes in imaging equipment or firmware,
significant shifts in patient or data distributions, material workflow changes, cybersecurity events, and
observed changes in error patterns.

Periodic re-benchmarking, sample-based clinician review, and performance-degradation monitoring
are complementary and should be combined where appropriate.
FDA should also consider
concordance and discordance analysis, review of near-misses, structured end-user feedback, subgroup
performance surveillance, audit-log review, and monitoring of processing failures or incomplete
analyses. Machine-based supervisory agents may assist with monitoring, but they should themselves
be validated, versioned, auditable, and monitored. Automated supervision should not become an
opaque second layer of risk.

V. Responsibilities Should Follow Observability, Capability, and Control
No single stakeholder can fully observe or manage all risks in a deployed GenAI-enabled device. FDA
should establish complementary responsibilities while preserving clear manufacturer accountability:

• Manufacturers should monitor technical performance, model integrity, processing failures,
updates, known limitations, and safety signals; investigate product-related degradation; provide
monitoring-enabling information and safeguards to identify and communicate use beyond
validated indications for use; and support timely corrective action.
• Healthcare organizations should maintain governance, data capture, model and version tracking,
audit trails, escalation processes, and the ability to restrict, pause, or discontinue use when
warranted.
• Clinical AI and specialty teams should define clinically meaningful measures and thresholds,
interpret signals, evaluate human-AI performance, and recommend revalidation or intervention.
• End users should understand the device's intended use and limitations, exercise appropriate clinical
oversight, and have easily accessible mechanisms to report suspected errors, unexpected behavior,
and workflow or safety concerns.
• Professional societies and standards organizations can develop consensus measures, independent
benchmarks, implementation guidance, education, and multi-institutional monitoring
infrastructure.
These complementary roles should not diffuse accountability. Manufacturer responsibility should
remain primary for the safety and effectiveness of the marketed device, including performance
changes caused by components or third-party models incorporated into the device. FDA should
establish a dual-reporting pathway that allows manufacturers and healthcare organizations or
clinicians to report significant performance deviations directly to the agency, especially when vendor
response is delayed or inadequate.

FDA should provide greater clarity regarding accountability when AI-enabled devices contribute to
patient harm. As generative AI systems assume increasingly autonomous and near-autonomous
functions, responsibility should be proportional to each stakeholder's ability to observe, evaluate, and
control system behavior. Clinicians should not become the default bearers of liability for errors arising
from AI systems they cannot reasonably oversee or control.

VI. Transparency and Monitoring-Enabling Information Are Essential
Healthcare organizations cannot meet monitoring expectations without timely access to information
needed to identify, investigate, and act on safety signals. FDA should require manufacturers to provide
model and system identifiers; complete version histories; descriptions of changes and expected
performance effects; known limitations and failure modes; relevant input and output data
specifications; processing status and error information; subgroup performance information; and audit
information sufficient for root-cause analysis.

For generative AI-enabled devices, manufacturers should provide healthcare organizations with
sufficient transparency regarding the prompts, system instructions, and configurations that govern
device behavior. This transparency is necessary to support clinical evaluation, ongoing monitoring, and
appropriate adaptation to local patient populations and workflows. Manufacturers should support
controlled customization and validation of these components when appropriate, while maintaining
protections for proprietary intellectual property. Changes that could materially affect clinical
performance should be documented and subject to appropriate revalidation.

When clinically meaningful, point-of-care transparency should indicate whether eligible inputs were
successfully analyzed, which images, series, text, or other inputs were used, whether exclusions or
processing errors occurred, and, where validated and appropriate, provide output-specific uncertainty
or confidence information for the individual patient rather than relying solely on general statements in
model documentation. Because GenAI systems can produce convincing but incorrect outputs, this
information should help users recognize when additional review or verification is warranted. Static
documentation is not sufficient when safe interpretation depends on information available within the
workflow.

FDA should require standardized model cards and dataset reporting elements that are updated with
each material version. For foundation models supporting multiple applications, documentation should
include a core model description and application-specific sections that distinguish validated uses from
exploratory or off-label uses.

VII. PCCPs and Foundation Model Dependencies Require Stronger Change Controls
PCCPs can support iterative improvement, but future changes to GenAI-enabled devices may be
difficult to fully pre-specify. FDA should permit flexibility only within defined safety boundaries and
require sponsors to identify the types of changes, methods of evaluation, acceptance criteria,
communication procedures, and circumstances that require a new submission. Changes affecting
intended use, autonomy, user population, clinical domain, safety-critical behavior, or validated
performance boundaries should generally require heightened review.
FDA should require timely notice to deploying organizations before material updates, except when
urgent security or safety action is necessary. Notices should explain what changed, why it changed,
which functions or populations may be affected, how performance compares with the prior version,
and whether local revalidation is recommended. Healthcare organizations should retain access to prior
version information for diagnostic consistency, troubleshooting, and root-cause analysis.

For devices that rely on third-party foundation models, the device manufacturer must remain
responsible for detecting, evaluating, and responding to upstream changes.
Contracts, technical
controls, version pinning, change notifications, audit logs, fallback procedures, and re-benchmarking
requirements should support this responsibility. A voluntary “Foundation Model Device Master File”
could improve regulatory efficiency, but participation should not reduce the device sponsor’s
obligation to demonstrate safety and effectiveness for each intended use. FDA should consider
incentives for participation and minimum update-notification commitments so that the files remain
current and useful.
VIII. Agentic Systems Warrant Heightened Oversight
Agentic systems may plan and execute multi-step actions, interact with external tools, and propagate
errors across interconnected workflows. FDA should evaluate not only individual outputs but also
decision pathways, tool selection, sequencing, autonomy boundaries, handoff quality, error recovery,
and cumulative downstream effects. Acceptance criteria should be more stringent as autonomy and
consequences increase.

High-consequence or irreversible actions should include validated human-authorization checkpoints
unless the sponsor provides compelling evidence that autonomous operation offers a favorable
benefit-risk profile.
Systems should maintain sufficiently detailed, reviewable audit logs of prompts,
retrieved information, actions, tool outputs, failures, overrides, and human interventions to support
clinical oversight, incident investigation, and accountability.

IX. FDA Should Support Scalable Implementation and Harmonization
Meaningful monitoring should not depend on the resources of an individual healthcare organization.
Less resourced and rural institutions may be unable to build independent monitoring systems, creating
uneven safety oversight.
FDA should coordinate with other Department of Health and Human Services
agencies to support interoperable data and reporting standards, shared monitoring tools, multiinstitutional validation networks, quality datasets, professional education, and research on human-AI
interaction.

FDA should work with clinical experts, professional societies, standards organizations, patients,
healthcare organizations, and industry to define common data elements for AI identification and
versioning, provenance, processing status, outputs, audit information, and safety reporting.
Harmonized standards should reduce duplicative local burden and support cross-institutional learning
while preserving privacy and proprietary protections. These efforts should build upon existing
interoperability standards, including DICOM, FHIR, and established common data element frameworks,
wherever appropriate.

Conclusion
RSNA supports FDA’s effort to modernize oversight for GenAI-enabled medical devices. A successful
framework should be risk-based and adaptable, but also sufficiently rigorous to address variable
outputs, broad capabilities, foundation-model dependencies, and increasing autonomy. Premarket
evaluation and postmarket monitoring should function as complementary safeguards, not substitutes.

RSNA encourages FDA to center patient safety; evaluate the final clinical AI system rather than the
model alone; require clinically meaningful transparency and traceability; align responsibilities with
stakeholder observability, capability, and control; preserve manufacturer accountability; and support
scalable monitoring infrastructure. These principles can enable innovation while maintaining the public
and professional trust necessary for GenAI-enabled devices to improve patient care.

RSNA appreciates the opportunity to provide input and looks forward to continued collaboration with
FDA. For additional information, please contact Libby O’Hare, Director of Government Relations, at
eohare@rsna.org.

Sincerely,

Carolyn C. Meltzer, MD
Chair of the Board
Radiological Society of North America