Q1 · The two-axis risk frameworkQ2 · The spectrum of device activityQ3 · When an output becomes directiveQ6 · Care escalation functionsQ7 · The competency-based approachQ9 · The benchmarking structureQ10 · Benchmark contamination and saturationQ11 · Clinical confirmation without a prospective trialQ14 · Comparators and acceptance criteriaQ15 · Performance against usual careQ16 · Independent third partiesQ18 · Trading premarket certainty for postmarket monitoringQ19 · Postmarket performance evaluationQ20 · Machine-based supervisory agentsQ21 · Clinicians, institutions and societiesQ22 · Re-benchmarking after a modificationQ23 · PCCPs for GenAI devicesQ25 · Foundation Model Master FilesQ26 · Agentic devices
The comment as filed
Please see attached comments from the Health Innovation Alliance (HIA).
Attachment
October 19, 2026
Food and Drug Administration
Center for Devices and Radiological Health
Digital Health Center of Excellence
10903 New Hampshire Avenue
Silver Spring, MD 20993
Re: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper
and Request for Feedback (Docket No. FDA-2026-N-7874)
Submitted electronically via regulations.gov
Dear FDA Digital Health Center of Excellence Team,
HIA is a coalition of patient advocates, specialty care providers, and technology companies that has
advanced a risk-based, innovation-forward approach to health technology policy for nearly two decades.
Our response below builds on HIA's February 2026 comments on HHS's Health Sector AI RFI i, HIA's
earlier comments on FDA's AI-enabled device performance RFI ii, HIA's Clinical AI Confidence Compact
concept, which was shared directly with FDA staff before this discussion paper's publication, and
conversations with patient advocacy and provider organizations over the summer of 2026, none of which
called for a pause or new regulatory oversight of clinical AI.
Autonomous clinical AI is not a future scenario for FDA to regulate cautiously into existence. It is here,
and it is needed. What's missing is clarity and certainty on whether and how oversight makes sense. Our
core message is simple: accelerate real-world evidence generation.
Independent studies in imaging, primary care, and cardiology show AI, alone or alongside clinicians,
meeting or exceeding clinician-only benchmarks on specific, well-defined tasks.iii Independent evaluators
also caution that measurable, real-world outcome evidence remains thin relative to model capability. iv
HIA reads that gap as the case for phased, monitored, time-limited deployment that generates the missing
evidence.
The real risk to patients is unsubstantiated risk itself. Treating every use case as equally uncertain and
equally dangerous, without weighing alternatives such as no care at all, does not protect patients. It
compounds an access problem this technology is positioned to help solve. The United States faces a
looming physician shortage. v Ninety-two percent of rural counties are federally designated primary care
shortage areas.vi For a patient in one of those counties, the realistic comparator for an AI-delivered
diagnosis is not “a clinician, or an AI” — it is “an AI, or nothing.”
A second, more immediate risk is a state-by-state patchwork of AI-in-health-care legislation, much of it
moving faster than this federal process, that threatens to create fifty different, cumbersome pathways for
bringing the same innovation to market. Federal action can lessen fragmentation by giving states a
credible, evidence-generating federal alternative to enacting rules alone.
Industry's own posture confirms the cost of continued ambiguity. Most generative AI tools on the market
today are, in one attorney's description, “dancing around the FDA-regulated space” — deliberately
avoiding device-triggering functions until FDA clarifies what will actually be required, because, in her
words, “[n]o one really wants to be the guinea pig.” vii Companies that have already accepted FDA
regulation, the same attorney adds, are looking for more clarity, and at least one company CEO describes
regulatory leadership from FDA as the most important safeguard against either a patchwork of different
regulations or bad actors and reactionary responses.viii Ambiguity limits innovative and safe products
from reaching patients.
HIA welcomes this discussion paper and supports its general framing, in particular:
• Its two-axis risk framework, with two refinements: the activity axis should measure reliance
(whose judgment an outcome depends on), and the consequence axis should reflect the chance,
frequency, and severity of an inaccurate output.
• Its openness to accepting greater premarket uncertainty in exchange for real-world postmarket
monitoring.
• Its competency-based approach, benchmarking paired with clinical confirmation, modeled on how
physicians are trained and credentialed, and its recognition that clinical confirmation need not
always require a prospective study.
We do caution that the framing of this paper could be read to reach software functions outside the device
definition. In particular, FDA's framework should:
• Continue to distinguish AI-enabled tools that support health care professionals' independent
decision-making from software that performs regulated medical device functions.
• Reference the non-device CDS criteria that the 21st Century Cures Act added to the FD&C Act
at section 520(o)(1)(E).
A discussion paper is not a policy. HIA is asking FDA to act now: build on TEMPO to create a Clinical
AI Confidence Compact (“Compact”) with interested U.S. states and companies to generate the evidence
this paper says it needs, while patients are already being served.
HIA’s proposed Compact is a time-limited agreement under which a defined AI-delivered care model
operates with phased autonomy (full clinician review, then sampling review, then monitoring) and
continuous reporting, to generate real-world evidence on where risk actually lies. Such Compacts would
let real-world evidence, generated under a phased, monitored, and temporary deployment, count as valid
clinical confirmation evidence under the framework this paper proposes.
Detailed Response
The responses below follow the discussion paper's own section structure. HIA's principal
recommendations are in Questions 1, 2, 11, 15, 18, and 22 and in Part E.
A. Section IV — Risk Assessment Framework
Question 1 — Does the two-axis framework capture the right dimensions?
HIA agrees the two-axis framework — activity and consequence — captures the dimensions that matter
most. We would caution against adding dimensions such as reversibility or traceability as independent
axes. They should be treated as modifiers within the existing framework.
HIA would clarify what the activity axis is actually measuring. Read narrowly, "activity" can seem to
describe what the AI does. What matters clinically is different: as a given diagnosis, treatment course, or
clinical finding is reached, how much does it depend on an informed person's judgment, versus the AI
operating on its own. That reliance is what should place a function on this axis (see more, Question 2).
HIA also recommends clarifying the consequence axis. HIA has previously defined health IT risk as a
function of three factors: the chance that a system's output will be inaccurate, the frequency of potential
inaccuracy, and the severity of the outcome if an inaccurate output is relied upon, including whether the
harm is reversible.ix HIA urges FDA to treat consequence as these three components.
Severity sets a reference point between clinical domains: an inaccurate insulin dose carries more severe
potential consequences than an inaccurate poison-ivy recommendation, as a category. That comparison
does not mean every instance of autonomous insulin dosing carries the same risk. Within a single clinical
domain, actual risk still varies with the chance and frequency of inaccuracy. Benchmarking, real-world
evidence, and monitoring affect those risk factors. A function's consequence tier should therefore be
assessed using real-world data on its actual chance and frequency of error, including missed escalation
triggers that were not anticipated. All three components should also be judged against the realistic
alternative available to the patient (see more, Question 15).
Question 2 — How should the non-directive to action-directing spectrum be assessed?
Consistent with Question 1 above, directiveness exists on a continuum. Wording alone ("recommend,"
"should") is a poor proxy for where a function sits on it. Time pressure is also an uncertain proxy; the
automation-bias literature reports mixed effects. The variable that matters is reliance: as a given diagnosis,
treatment course, or clinical finding was reached, how much did it depend on a person's judgment, versus
the AI operating on its own.
Congress already used reliance as the organizing idea in the CDS exclusion that the 21st Century Cures
Act added to the FD&C Act (section 520(o)(1)(E)). HIA's point is that reliance is better measured directly
by asking whose judgment produced the outcome.
Reliance is a property of a function's design and intended use. Escalation triggers or pathways for
clinician review and the system's containment capabilities, including the ability to withdraw autonomy
rapidly, shift reliance from the AI function to an HCP. Consider an AI function for uncomplicated urinary
tract infection, one of the low-complexity conditions HIA's Clinical AI Confidence Compact identifies for
Track 1. A version of this function that routes red-flag presentations to a clinician sits closer to the
clinician-reliant end of the axis than an otherwise identical function with no such trigger.
In short, a function's position on the axis is set by reliance, as assessed from its design (including its
escalation pathways) and the evidence on how it performs in use. That evidence can support a defined
change to the function's design and scope, such as narrowing or widening the conditions under which
autonomy is withdrawn, through the PCCP mechanism discussed under Question 22.
Question 3 — Do patient-facing functions present different or higher risk than HCP-facing functions?
Yes, and consistent with Questions 1 and 2: a patient-facing function could sit higher on the activity axis,
depending on the user's reliance on the AI system versus a clinician. Yet, the relevant comparison for a
patient with no HCP in the loop is AI versus whatever the patient would rely on otherwise: unmoderated
information, delayed care, or no care at all (Question 15).
Also, its consequence assessment should still be evaluated on the three dimensions above: chance,
frequency, and severity of an inaccurate output, including reversibility.
Question 6 — Weighing under- vs. over-escalation error
Both failure modes should be weighed together. Over-escalation carries its own cost: unnecessary
emergency-department utilization, patient anxiety, and, over time, erosion of trust that pushes patients
back toward no care at all. In an access-constrained system, directing a patient toward care they cannot
reach or afford has much the same effect as under-escalation. Benchmarking for escalation functions
should measure both directions against defined thresholds.
B. Section V — Competency-Based Premarket Evaluation
FDA grounds this competency-based approach in how physicians themselves are evaluated and
credentialed, a construct similar to HIA's Clinical AI Confidence Compact.
Device benchmarking is necessary, but insufficient alone to demonstrate real-world competence. Clinical
confirmation is where independence expands as competence is demonstrated. The Compact's phases track
this directly with supervision narrowing as performance holds up. At each stage, the question is the one
raised in Question 2 above: on whose judgment the clinical outcome actually depends.
Question 7 — Is the competency-based approach (benchmarking plus clinical confirmation) sound?
Yes. The two-part structure fits GenAI's open-ended behavior better than exhaustive input/output testing,
for the reasons above. FDA's Section V.C methods — shadow deployment, clinician adjudication,
prospective study — already describe most of the phased structure in HIA's Compact concept. See
Question 11.
Question 9 — Are there missing or redundant benchmarking elements?
HIA recommends one addition: an explicit standard-of-care/guideline-concordance element, alongside the
ground-truth-based elements already listed, measuring not only whether an output is technically correct
but whether it matches or exceeds what a qualified clinician panel, or a recognized clinical-guideline
source, would recommend for the same patient in the same circumstances. This is the comparator question
raised in Section V.D.1 and deserves its own benchmarking element.
Questions 10 and 16 — Benchmark validity and independent third parties
Benchmarks created by device developers themselves should remain an acceptable component of
competency-based evaluation, provided their methodology is transparent and their clinical relevance is
validated against established clinical guidelines and representative clinical scenarios. Depending on the
function's risk profile, validation may include independent qualified clinician review, established clinicalguideline and evidence platforms, or other appropriate independent corroboration. Limiting evaluation to
third-party or FDA-developed benchmarks alone could concentrate evaluation capacity in a small number
of institutions and slow the field where CDRH states it wants a scalable, high-throughput approach. For
the same reason, third-party evaluation should be an option where warranted.
Question 11 — Should prospective study be required in every case?
It should not. FDA's named methods — shadow deployment, clinician adjudication, prospective study —
describe increasing levels of rigor that a real-world deployment structure like HIA's Compact is designed
to move through as evidence accrues. HIA urges FDA to state plainly that evidence generated in this way
satisfies Section V.C clinical confirmation.
Question 14 — How should comparators and acceptance criteria be selected for open-ended outputs?
Comparing performance to a panel of qualified clinicians reflecting the standard of care is sound. FDA
should not treat performance equal to a median clinician as the ceiling of an appropriate comparator
standard. For narrowly defined, high-volume, well-evidenced tasks, the right comparator may be the
performance of expert, seasoned specialists, because for those specific tasks the evidence shows some
systems already clear it. HIA also cautions against defaulting to specialist-panel comparators as uniformly
more rigorous than generalist comparators; the right standard depends on the population the device is
intended to serve and the alternative realistically available to it, addressed further under Question 15.
On human-AI team versus fully autonomous comparison: HIA urges FDA not to treat human-AI team
performance as inherently the safer or preferred standard by default. A pooled analysis of 52 studies
found human-AI teams beat the clinician alone in 95 percent of conditions but beat the AI alone in only
44 percent.x Others reach similar conclusions. xi
Autonomous operation is not uniformly safer regardless of task: a 2026 safety benchmark found top
models produced potentially severely harmful recommendations in roughly one in fourteen consultations,
though they outperformed generalist physicians on the same cases.xii This is why competency-based
benchmarking, paired with phased, monitored deployment, is the right tool for managing this risk.
Requiring human review as a categorical safety measure, without evidence that the specific human-AI
configuration improves outcomes, risks codifying a comparator that sounds safer than it measurably is.
Question 15 — Should performance be assessed relative to the care likely to occur in the device's
absence?
HIA urges FDA to formally recognize no clinical encounter at all as a common and important comparator
for real-world deployment, particularly for populations with limited access to alternatives. xiii For a
substantial share of the population these tools are intended to serve, the realistic choice is AI or nothing. A
framework that measures AI function only against clinician performance, without also measuring what
happens when a comparably situated patient has no functioning access to a clinician at all, will understate
this technology's benefit where that benefit could be largest.
C. Section VI — Postmarket Monitoring
Question 18 — Should FDA accept greater premarket uncertainty through greater reliance on
postmarket monitoring?
Yes, greater reliance on postmarket monitoring makes sense, in proportion to risk. CDRH's own examples
already draw this line, contrasting autonomous antibiotic prescribing for confirmed strep throat with
autonomous thrombolytic initiation in a stroke workflow. Strep throat sits squarely in the same category
as HIA's Clinical AI Confidence Compact Track 1 conditions: well-established, evidence-based protocols;
low clinical complexity; episodic conditions such as uncomplicated UTI, acute sinusitis, or bacterial
conjunctivitis. Where an established protocol lowers the chance of inaccuracy, the condition's severity is
low even if mismanaged, and escalation triggers bound the frequency of unmitigated error, FDA should
accept meaningfully reduced premarket evidence in exchange for robust, real-time postmarket
monitoring. The appropriate balance should reflect the clinical risk of the intended function, the strength
of existing evidence, and the ability of monitoring and escalation safeguards to identify and mitigate
potential harm.
Question 19 — Periodic re-benchmarking, sample-based review, and drift monitoring
HIA supports these as the right building blocks. They are how the chance-and-frequency evidence
described under Question 1 gets generated. HIA's Clinical AI Confidence Compact concept explores a
similar structure on a temporary basis, regular reporting on volume, escalation rates, safety events, and
complaints, with a defined transition from full clinician review to sampling to steady-state monitoring, as
a way to generate evidence on whether and where risk actually lies.
Question 20 — Machine-based supervisory agents
This is a promising area for continued exploration. HIA would support a future, specifically scoped FDA
effort here once the Section V framework matures.
Question 21 — Roles for clinicians and institutions without diffusing manufacturer accountability
HIA recommends a sliding-scale approach. The manufacturer's postmarket monitoring and real-time
escalation obligations should scale with the function's activity and consequence tier, so accountability
remains with the manufacturer by default and is only shared where a treating clinician or institution is
actually present. FDA's own interoperable automated insulin delivery framework already reflects this
logic, holding manufacturers accountable for real-time, autonomous dosing decisions made with no
clinician in the loop at the time of use, allocated by each component's function. xiv This connects to the
liability question raised in E.1 below, which this discussion paper does not address.
Question 22 — Premarket assessment as a baseline for postmarket modifications
HIA has separately and repeatedly urged FDA to unlock the PCCP pathway's full potential. Using the
premarket competency-based assessment as a baseline against which modifications are re-evaluated is the
right model. HIA urges FDA to pair this discussion paper's framework with a commitment to approve
broader, pre-specified evolutionary pathways for GenAI-enabled devices specifically, given how
frequently these systems are expected to change relative to traditional software.
Question 23 — PCCP when future modifications cannot be fully prespecified
HIA would support a bounded evolutionary envelope approach: prespecifying categories and limits of
change, such as retraining cadence or permitted guardrail modifications. Requiring every specific future
change to be named in advance is unworkable.
D. Section VII — Other Topics
Question 25 — Foundation Model Device Master Files
HIA supports the voluntary Foundation Model MAF concept in principle, as a way to give FDA reference
information without a new mandatory disclosure burden on device sponsors who rely on third-party
models. Its value will depend on whether model developers see sufficient benefit, such as faster review
for devices built on their models, to participate voluntarily.
Question 26 — Additional considerations for agentic AI systems
The agentic AI competencies FDA proposes under element A.1 are a reasonable start, consistent with how
the field is now defining this category. xv HIA notes that Track 2 of its Compact concept, continuous
monitoring with defined escalation triggers for complex, ongoing conditions, is directly relevant to
agentic systems that plan and execute multi-step actions, and welcomes the opportunity to discuss how
Compact-style monitoring could complement agentic-specific premarket benchmarking.
E. Issues This Discussion Paper Does Not Ask About, But Should
1. Liability and safe harbor
This discussion paper is silent on liability. HIA has proposed a safe harbor model tied to substantial
compliance with a monitoring structure like the Compact: a participant in substantial compliance with
phased-autonomy monitoring and reporting obligations should have protection against liability claims
except in cases of fraud, gross negligence, or willful misconduct, while patients who decline AI-delivered
care retain full recourse under ordinary standards through the non-AI alternative that must remain
available to them. HIA urges FDA to engage on how its evidentiary framework could inform, or be
informed by, parallel HHS and state work on liability standards for compliant deployments.
2. State practice-of-medicine authority
FDA authorization of a device software function does not resolve whether that function may lawfully
deliver care without a licensed clinician under a state's practice-of-medicine laws. The resulting state-bystate patchwork is a significant barrier to clinical AI deployment. HIA urges FDA to support a
mechanism, such as the tri-party structure in HIA's Compact, that pairs FDA's enforcement discretion
with state practice-of-medicine relief under a single time-bound agreement, and to use what it learns to
inform any longer-term policy fix.
3. Scope of the framework
FDA's paper addresses software functions that meet the device definition. Section 520(o)(1)(E) of the
FD&C Act, added by the 21st Century Cures Act, excludes from that definition clinical decision support
that supports a health care professional's judgment and enables the professional to independently review
the basis for its recommendations, so that the professional does not rely primarily on them. HIA asks FDA
to state in any follow-on guidance that the risk framework, including its treatment of action-directing
functions, applies only to functions that meet the device definition and does not alter the 520(o)(1)(E)
criteria.
Conclusion
HIA appreciates FDA's engagement on this topic and shares its view that a risk-based, proportionate
approach is right. We ask FDA to commit to recognizing real-world evidence generated under phased,
monitored, time-limited Compact deployments as valid clinical confirmation evidence under Section V.C,
and, through TEMPO or a similar structure, to identify a cohort of low-complexity, well-establishedprotocol functions consistent with Track 1 for a near-term pilot. HIA and its members stand ready to help.
Sincerely,
Roslyn Docktor
Executive Director
Health Innovation Alliance
i HIA_AI_RFI_Response_2.23.26.pdf
ii HIA-Comments-FDA-AI-Enabled-Device-Performance-RFI-vFIN-1.pdf
iii Eisemann N, Bunk S, Mukama T, et al. Nationwide real-world implementation of AI for cancer detection in populationbased mammography screening. Nat Med. 2025;31:917–924. https://www.nature.com/articles/s41591-024-03408-6; Abràmoff
MD, Lavin PT, Birch M, Shah N, Folk JC. Pivotal trial of an autonomous AI-based diagnostic system for detection of diabetic
retinopathy in primary care offices. npj Digit Med. 2018;1:39. https://www.nature.com/articles/s41746-018-0040-6; see also
Huang JJ, et al. Autonomous artificial intelligence for diabetic eye disease increases access and health equity in underserved
populations. npj Digit Med. 2024;7:196. https://www.nature.com/articles/s41746-024-01197-3; Korom R, et al. Real-world
deployment of a generative AI clinical decision support tool across primary care clinics in Kenya (39,849 patient visits), as
reported in ARISE Network, The 2026 Healthcare AI Industry Report (July 2026), at 34, https://www.ariseai.org/reports/2026-healthcare-ai-industry-report.pdf; Zhang, et al. Clinician-artificial intelligence collaboration: a win-win
solution for efficiency and reliability in atrial fibrillation diagnosis (2025), as reported in id. at 39.
iv ARISE Network (Perez A, Tusty M, Liu C, et al.). The 2026 Healthcare AI Industry Report. Stanford, CA; July 2026.
https://www.arise-ai.org/reports/2026-healthcare-ai-industry-report.pdf
v American Hospital Association. Fact Sheet: Telehealth. Feb. 2025 (reporting the Association of American Medical Colleges’
projected shortage of up to 86,000 physicians by 2036). https://www.aha.org/fact-sheets/2025-02-07-fact-sheet-telehealth
vi Horstman C, Shah A. The State of Rural Primary Care in the United States. The Commonwealth Fund; Nov. 17, 2025
(reporting that 92 percent of rural counties were designated primary care health professional shortage areas in 2023).
https://www.commonwealthfund.org/publications/issue-briefs/2025/nov/state-rural-primary-care-united-states
vii Reuter E. 4 questions about the FDA’s approach to generative AI. MedTech Dive; Sept. 9, 2026 (quoting Suzanne Levy
Friedman, Partner, Honigman). https://www.medtechdive.com/news/4-questions-about-the-fdas-approach-to-generativeai/829871/
viii Reuter, supra note vii (paraphrasing Sharif Vakili, CEO, UpDoc).
ix Health Innovation Alliance. Use Cases for AI Tools to Improve Health Care. March 2025, at 7 (defining health IT risk); see
also id. at 10 (listing level of reliance on AI output for decision or action among the factors influencing risk). https://healthinnovation.org/wp-content/uploads/2025/03/HIA-AI-Use-Cases-Report-2025.pdf
x Liu Y, et al. Pooled reliability analysis of human-AI team performance across 52 studies and 87 clinical conditions (2025), as
reported in ARISE Network, supra note iv, at 41.
xi Emanuel EJ, Baker-Butler A, Khosla N, Khosla V. Will autonomous AI exceed AI-aided physicians as the best medical care?
JAMA. 2026;336(11):915–918. https://doi.org/10.1001/jama.2026.15380; Warren LM, et al. Retrospective evaluation of AI as
a second reader in NHS breast cancer screening, including arbitration outcomes (2026), as reported in ARISE Network, supra
note iv, at 42.
xii Wu D, et al. NOHARM: a benchmark for clinically harmful model outputs (2026), as reported in ARISE Network, supra
note iv, at 26–27.
xiii See supra notes v and vi; see also Samuels J. America’s least-discussed physician shortage is getting worse. Forbes; Aug. 4,
2026 (reporting access gaps in geriatrics, sleep medicine, and family medicine).
https://www.forbes.com/sites/johnsamuels/2026/08/04/americas-least-discussed-physician-shortage-is-getting-worse/
xiv See 21 C.F.R. § 862.1356 (interoperable automated glycemic controller; Class II, special controls),
https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-862/subpart-B/section-862.1356; see also 21 C.F.R. §
862.1355 (integrated continuous glucose monitoring system); 21 C.F.R. § 880.5730 (alternate controller enabled infusion
pump).
xv Angus DC, Khera R, Lieu T, et al. AI, Health, and Health Care Today and Tomorrow: The JAMA Summit Report on
Artificial Intelligence. JAMA. 2025;334(18):1650–1664. https://doi.org/10.1001/jama.2025.18490