American College of Radiology (Dana H. Smetherman)

CliniciansClinicianFiled October 19, 20268,809 words · 1 attachmentFDA-2026-N-7874-0399
LinkedInX

Themes it raises

16 of the 21 themes in the docket, each with the passage we counted, verbatim.
What makes a function high riskFDA Q1, Q2, Q5
“We recommend adding contextual modifiers that refine risk assessments within the framework rather than adding dedicated axes that may complicate the framework without adding clarity.”
Whether the user can judge the outputFDA Q3, Q4
“Patient users have widely varying health literacy levels, and this changes the risk assessment paradigm of many functions due to different error detection capabilities, output reliance, downstream action, and time-to-treatment.”
Escalating too little and too muchFDA Q6
“FDA should therefore require sponsors to report under-escalation and over-escalation separately, with predefined acceptance criteria for each.”
Judging devices the way clinicians are credentialedFDA Q7, Q8
“Finally, FDA should exercise caution when drawing analogies between GenAI evaluation and the education, training, or credentialing of healthcare professionals.”
Whether benchmark results prove anythingFDA Q9, Q10, Q16
“While the extent of benchmark contamination in medical AI evaluation may not always be known, sponsors should disclose known overlaps, characterize remaining uncertainty, and describe measures taken to reduce bias in performance estimates.”
Proving the device works in real careFDA Q11, Q12, Q13, Q14, Q15
“Finally, FDA should require clinical confirmation using real patient data whenever the intended user cannot independently assess the accuracy of the output.”
Trading premarket certainty for postmarket monitoringFDA Q18
“This tradeoff is more defensible when a qualified end user reviews each output before it influences patient care and is less appropriate for patient-facing or autonomous functions.”
Watching the device after it shipsFDA Q19, Q20
“Monitoring programs should capture all eligible cases, including processing failures, missing outputs, abstentions, raw outputs before clinician editing, and clinically relevant downstream outcomes.”
Who is accountable when something goes wrongFDA Q21
“For each deployment, stakeholders must assign named responsibility for investigating alerts, stopping use, notifying affected sites, reviewing cases already affected, and maintaining care during outages.”
Controlling a device that keeps changingFDA Q22, Q23, Q24, Q25
“Agreements with suppliers should address behavior-affecting changes and model retirement and provide sufficient notice to complete verification.”
Devices that plan and take actionsFDA Q26
“Drafting an order, submitting it, and submitting it irreversibly present different risks and may warrant different levels of validation and oversight.”
Whether human oversight is real oversightFDA Q3, Q4, Q14, Q20, Q21, Q26
“This oversight should earn credit only when sponsors demonstrate its effectiveness in the intended workflow (e.g., error detection and correction under realistic workload, time pressure, and interface conditions, with residual errors reported after review).”
Records that let investigators reconstruct an eventFDA Q19, Q21, Q24, Q26
“Traceability/source attribution: The ability to identify information underlying an output and reconstruct decisions with confidence that citations are accurately and correctly attributed.”
Security, dependencies and what happens when they failFDA Q1, Q9, Q24
“Agentic systems may introduce cybersecurity and prompt-injection risks through user inputs, images, retrieved content, tool outputs, and other external data sources.”
How this fits rules that already existFDA Q8, Q9, Q16, Q25
“In any future regulatory framework specific to GenAI, FDA should clarify applicability to currently authorized medical device functions, such as those enabling generative image reconstruction, denoising, and synthesis.”
Equity, access and under-resourced settingsFDA Q3, Q13, Q21
“Finally, clinically consequential subgroup claims should be confirmed with independent real-world data.”

FDA questions it names

Questions this filing names by number.

Q1 · The two-axis risk frameworkQ2 · The spectrum of device activityQ3 · When an output becomes directiveQ4 · Generalist and specialist usersQ5 · Multi-turn conversations that migrateQ6 · Care escalation functionsQ7 · The competency-based approachQ8 · Mapping the risk grid to evidenceQ9 · The benchmarking structureQ10 · Benchmark contamination and saturationQ11 · Clinical confirmation without a prospective trialQ12 · Statistically meaningful performanceQ13 · Synthetic dataQ14 · Comparators and acceptance criteriaQ15 · Performance against usual careQ16 · Independent third partiesQ17 · Devices with many functionsQ18 · Trading premarket certainty for postmarket monitoringQ19 · Postmarket performance evaluationQ20 · Machine-based supervisory agentsQ21 · Clinicians, institutions and societiesQ22 · Re-benchmarking after a modificationQ23 · PCCPs for GenAI devicesQ24 · Third-party foundation model changesQ25 · Foundation Model Master FilesQ26 · Agentic devices

Machine-assisted draft, pending human review. The source text and highlighted passages appear below. Read the filing on regulations.gov ↗

The comment as filed

Comment submitted on regulations.gov. Passages we counted are highlighted.

Please see the attached American College of Radiology (ACR) comment letter to FDA regarding "Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for Feedback" (FDA-2026-N-7874). Thank you.

Attachment

Attachment, text extracted from the filed document. Passages we counted are highlighted.

October 19, 2026

Dockets Management
Staff (HFA-305), Food and Drug Administration
5630 Fishers Lane, Rm. 1061
Rockville, MD 20852

Re: (Docket ID: FDA-2026-N-7874) Considerations for the Regulation of Generative AIEnabled Medical Devices: Discussion Paper and Request for Feedback; Comments of
the American College of Radiology

The American College of Radiology (ACR) – a professional association representing more than
40,000 physicians practicing diagnostic radiology, interventional radiology, radiation oncology, and
nuclear medicine, as well as medical physicists – appreciates the opportunity to comment on the
discussion paper published August 18, 2026, by the Food and Drug Administration (FDA) titled
Considerations for the Regulation of Generative AI (GenAI)-Enabled Medical Devices. ACR
previously presented on this subtopic in-person to the FDA Digital Health Advisory Committee
during its public meeting on Nov. 20-21, 2024, and filed written comments on Jan. 21, 2025, during
the subsequent comment period corresponding with the same meeting. ACR supports FDA’s
continued efforts to gather stakeholder perspectives on this topic before proposing any new policy.

General Comments

Radiology is a leading specialty domain for GenAI-enabled medical device innovation, and imageinformed draft reporting is among the most active areas of development and early clinical
evaluation. A central challenge for FDA oversight of GenAI-enabled devices in radiology and other
specialties is that device performance often depends not only on the software itself, but also on
the clinicians, workflows, data sources, governance structures, and monitoring systems with
which it operates.

The ACR’s complementary approach to post-deployment AI safety—including continuous postmarket performance monitoring, future practice accreditation for healthcare facilities, and the
responsible use of radiology AI in practice—has the ability to pick up on this composite system
performance, particularly as these safeguards become increasingly connected to longitudinal
outcomes data. We welcome continued partnership between FDA and the medical community to
take on these challenges.

Consistent with this view, we urge FDA to incorporate the following principles into any future
oversight framework for GenAI-enabled medical devices:

● Regulate what the system does and the risk it creates. FDA should regulate the complete
clinical function and its associated risk. The same underlying foundation model may
support low-risk documentation or high-risk autonomous diagnosis. As a result, regulatory
requirements should reflect intended use, potential patient harm, workflow, and degree of
autonomy.
● Explicitly credit qualified end-user oversight, but only if demonstrated. A GenAI output
reviewed and affirmatively accepted by a qualified end-user before the output influences
patient care has a fundamentally different risk profile from an output that independently
makes, communicates, or acts upon a clinical determination. FDA should therefore
recognize qualified-user review as an important risk control, but only while accounting for
error detectability, automation bias, and the user’s ability to independently assess the
output. This oversight should earn credit only when sponsors demonstrate its effectiveness
in the intended workflow (e.g., error detection and correction under realistic workload, time
pressure, and interface conditions, with residual errors reported after review).

● Allow technology to evolve. FDA should move toward validated performance envelopes,
predetermined change control plans, and lifecycle oversight where these mechanisms are
appropriate. Changes to models, prompts, retrieval, orchestration, or guardrails that
remain within validated performance parameters should not automatically require a new
submission provided safety and effectiveness remain the same.
● Counterbalance premarket flexibility with real-world performance monitoring. Greater
flexibility for continuous improvement should be counterbalanced by robust postmarket
monitoring for clinically meaningful failures, performance drift, subgroup effects, and
emerging risks. For GenAI, postmarket monitoring should be viewed not simply as an
additional requirement but as a core component of the regulatory framework that enables
safe continuous improvement.
● Measure the failures that matter clinically. Evaluation should focus on clinically
consequential performance rather than generic hallucination rates, nondeterminism, or
textual similarity. ACR and other medical societies can help define meaningful categories
such as significant false assertions, omissions, laterality/localization errors, incorrect
comparisons, unsupported recommendations, contradictions, and communication
failures. GenAI requirements should also scale with autonomy.
● Clarify relevance to already-authorized medical devices. In any future regulatory framework
specific to GenAI, FDA should clarify applicability to currently authorized medical device
functions, such as those enabling generative image reconstruction, denoising, and
synthesis.

● Add worked imaging examples. Given the level of innovation in GenAI-enabled medical
devices within the radiology domain, any future regulatory framework, guidance, and other
FDA resources should include imaging examples.
● Seek imaging community feedback via dedicated activities. The discussion paper is broadly
applicable across all GenAI-enabled medical device issues. To better address imaging, FDA
should convene a radiology-focused public workshop, Digital Health Advisory Committee
meeting, and/or publish a targeted solicitation for written comments on imaging-specific
considerations prior to establishing a future regulatory framework, developing draft
guidance, or other policy developments.

Specific Questions

Q1. Does the two-axis risk framework, organized around device activity and the consequence
of relying on an incorrect output, appropriately capture the dimensions most relevant to the
risk of a GenAI-enabled software function? If there are additional dimensions—such as the
reversibility of a resulting action, the availability of downstream safeguards, the time
pressure of the deployment setting, or the traceability of the output (i.e., to primary source
materials)—that should be represented in a risk framework, please describe and provide
examples of how they should be represented.

A1. The described two-axis risk framework, which evaluates device activity and the consequences
of relying on an incorrect output, may oversimplify or overlook important contextual considerations
that influence risk. Risk in most clinical applications is not a function of the AI system alone, but of
the broader human-AI/socio-technical system in which it is deployed and used. Exceptions may
include highly autonomous AI functions or narrowly defined functions such as workflow triage,
image transformation, or text transformation that serve as individual components within a larger
clinical system.

We recommend adding contextual modifiers that refine risk assessments within the framework
rather than adding dedicated axes that may complicate the framework without adding clarity.

These considerations should include:

● Severity of harm from wrong outputs: Delineation of different levels of potential harm. FDA
describes severity with examples, but does not define the Limited, Moderate, and Severe
levels of the consequences axis.
● Expected frequency of wrong outputs: The frequency of erroneous outputs produced by the
device across the expected input mix and conditions of use.
● Detectability: Whether, and at what point, the intended user detects an erroneous output
within the real-world workflow. Different users may act differently when they detect an
error, which may impact subsequent actions and harm. Poor detectability increases the
likelihood that an erroneous output will result in harm and may increase the number of
affected patients before the user or system identifies the issue.
● Safeguards/deployment exposure: Separate consideration of device safeguards and what
deploying organizations must do to ensure safety. Rather than merely requiring description
of safeguards, require evidence that safeguards are effective under expected conditions of
use.
● Reversibility: Whether there is a practical and meaningful opportunity for human
intervention to undo the action between detection of an erroneous output and its effect on
patient care.
● Traceability/source attribution: The ability to identify information underlying an output and
reconstruct decisions with confidence that citations are accurately and correctly
attributed.
An erroneously generated or misattributed source is itself a failure mode that
can exacerbate safety issues via persuasiveness.

Finally, any future FDA guidance or other resource that establishes this framework as agency policy
should include axis-placement criteria and examples, including examples of low-autonomy
functions with severe downstream consequences. Risk assessment of GenAI will require greater
nuance than the discussion paper can illustrate via an abstract, generalized framework, and
therefore providing more examples would enhance clarity.

Q2. CDRH seeks input on how to account for the spectrum of GenAI-enabled informational
functions that vary in the degree to which they direct a user to a particular action (i.e.,
between “non-directive” and “action-directing”). What characteristics of an output—such as
its wording, specificity, personalization, or context—could be considered as modifiers of the
risk associated with an informational function, after accounting for the device’s overall
functionality and intended use? What additional information would provide manufacturers
with sufficient clarity and predictability around risk assessment for informational functions
while recognizing that directiveness may exist along a continuum rather than as a binary
distinction?

A2. The characteristics of an output described in question 2 are appropriate, but operationally
challenging. ACR supports FDA’s notion in the discussion paper that disclaimer content in a
patient-targeted output does not neutralize the core directive in that output.

Regarding context as an output characteristic, the key question in imaging is whether the reader
can independently assess the output and identify consequential errors. That assessment typically
lowers risk without establishing “low risk.” Harm severity if an erroneous output goes undetected,
reader time pressures, directly or indirectly triggered or suggested action(s), and presentation still
matter, however. The FDA should require the sponsor to demonstrate the detectability of errors,
including whether readers catch errors under real-world workloads and within vendor-provided
interfaces. The sponsor should also produce those measurements.

Personalization of the output can increase risk. For patient-users, personalization can make the
output seem more directive and detailed, which raises risk level. For clinician-users, a
personalized output, such as matching a radiologist's phrasing, can lower editing burden, but it
makes the AI output read like the radiologists’ own, which can lower scrutiny by the reporting
physician. FDA should evaluate detection performance in the personalized setting, not exclusively
in the default setting.

Finally, FDA should clarify whether directiveness is a characteristic of the device function or of a
particular output. This distinction is especially important for conversational systems, whose
outputs may range from purely informational to highly directive depending on user prompts.

Q3. CDRH seeks input on whether and when a GenAI-enabled function that results in delivery
of clinical information to patients, as opposed to HCPs, could present different or higher
risks, while also recognizing the potential benefits associated with improved patient
empowerment, engagement, and access to clinical information. What device characteristics,
output features, or safeguards might mitigate risks that could arise when a user lacks the
domain knowledge to independently evaluate an output, without unnecessarily
underestimating patient capability?

A3. FDA should consider the functions and safeguards in place for patient-delivered outputs,
rather than defaulting to higher risk determinations merely because the patient is an intended user.
Patient users have widely varying health literacy levels, and this changes the risk assessment
paradigm of many functions due to different error detection capabilities, output reliance,
downstream action, and time-to-treatment.
In addition, the potential risk is impacted by the
intended use of an AI product. For example, a CT prep instruction function has different risk
considerations than a symptom-triage function in practical terms. As both factors impact the
probability of harm and its severity in different ways, FDA should separate the two rather than
putting both on the consequences axis.
Safeguards should improve accuracy, uncertainty management, and user comprehension, rather
than rely primarily on users observing and adhering to disclaimers. Examples include grounding
responses in current and authoritative information sources, transparency regarding retrieved
content and its limitations, validated approaches for communicating uncertainty with appropriate
escalation thresholds, and demonstrated comprehension across the intended user population,
including patients with varying levels of health literacy. For higher-risk clinical decisions, systems
should explicitly avoid directing patients to initiate, discontinue, or modify treatment without
appropriate healthcare professional oversight.

Q4. CDRH seeks input on whether and how the distinction between generalist and specialist
physicians might inform the assessment of risk for HCP-facing GenAI-enabled functions.
Under what circumstances, if any, might risk be affected when an HCP who lacks the relevant
clinical specialist knowledge receives information that falls within a specialist area of
practice? What device characteristics or safeguards might mitigate such risks?

A4. The answer depends primarily on the specific function's intended use, potential consequences
of error, available safeguards, and whether the intended user can independently recognize and
appropriately respond to consequential errors.

The 2026 ACR–SIIM Practice Parameter for Imaging Artificial Intelligence defines a “qualified enduser” in this context as “a licensed physician with the necessary qualifications and training to
independently provide the equivalent medical service without the aid of the AI solution, and who
possesses qualifications and training in the use of the AI solution, including the ability to assess
the validity of its output.” FDA should recognize qualified end-user review as an important risk
control when the user evaluates and affirms the output before it influences patient care decisions
or actions. However, the effectiveness of this safeguard depends on the user's expertise, the
detectability of errors, the potential for automation bias, and whether specialist consultation or
referral is readily available in a timely fashion and effective within the intended workflow.

Additional safeguards might include output-uncertainty communication, recommendations for
specialist review or escalation in uncertain or high-risk situations, and system-level performance
monitoring.

Q5. For multi-turn conversational GenAI-enabled devices that may migrate from providing
“non-directive” information to “action-directing” information over the course of an exchange,
how should risk be assessed across realistic conversational trajectories? How could the
intended use of such a device be characterized when its behavior is emergent across a
conversation?

A5. The paper discusses drift symptoms across three benchmark items but leaves these
unconnected. For example, S.1 discusses “resistance to over-reassurance when users minimize
symptoms or resist escalation.” S.2 asks about "multi-turn conversations in which individual turns
appear in-scope, but the cumulative interaction drifts out of scope." R.1 lists degradation across
long conversations.

Sponsors should define intended use across the conversational trajectories the device supports,
including the transition from informative outputs to directive or suggestive outputs.
Informational/non-directive claims need evidence that the device holds that boundary. Where
safety depends on specialist escalation or user scope limitations, multi-turn adversarial testing
under sustained user pushback and the presence of sponsors to measure whether the device
withdraws or softens a warranted escalation with prespecified acceptance criteria should be
required.

Q6. For GenAI-enabled care escalation functions, CDRH is considering how the evaluation
may account for both under-escalation and over-escalation. How could manufacturers
characterize and weigh these two directions of error, given that they may not be
commensurable and that acceptable trade-offs may vary by clinical context?

A6. ACR recommends evaluating under-escalation and over-escalation as distinct performance
dimensions, with separate validation criteria, risk assessments, and mitigation strategies for each.
For patient-facing or patient-directing AI, sponsors should clearly communicate risks associated
with both under-escalation and over-escalation.

The acceptable balance between under-escalation and over-escalation is highly context
dependent. In medical imaging, for example, over-calling an actionable incidental finding may lead
to additional imaging, radiation exposure, healthcare costs, and patient anxiety. Once embedded
in a clinical report, an unnecessary but documented recommendation may also be difficult for
clinicians to reverse. Conversely, under-calling a finding may delay diagnosis or treatment. As
demonstrated by the ACR Incidental Findings Committee's work, determining appropriate tradeoffs requires consideration of the specific clinical context and evidence base.

FDA should therefore require sponsors to report under-escalation and over-escalation separately,
with predefined acceptance criteria for each.
Combining the two into a single aggregate escalation
metric could create incentives to offset under-escalation with over-escalation conservatism.

Q7. Is the competency-based approach described above, i.e., device benchmarking followed
by clinical confirmation, a useful and appropriate framework for evaluating GenAI-enabled
devices?

A7. ACR supports a framework in which the sponsor follows benchmark testing with clinically
grounded confirmation of performance for the device's intended use. Benchmark evaluation can
establish a minimum performance threshold, but benchmark results alone do not demonstrate
that a device will perform safely and effectively within diverse real-world clinical settings,
populations, or workflows. The sponsor should tie clinical confirmation to the device's intended
use, target population, and conditions of use.

FDA should avoid relying on aggregate benchmark scores alone. Benchmark performance may be
sensitive to prompt wording, question ordering, repeated runs, and other testing conditions, and
high benchmark performance does not necessarily predict safe and effective performance for the
intended use, population, or workflow. Sponsors should therefore provide evidence that
benchmark results are predictive of real-world performance for the intended use and supplement
summary metrics with sensitivity testing, failure analysis, and evaluation of clinically
consequential errors.
Additionally, if FDA intends to use ongoing benchmarking as part of a lifecycle oversight approach,
the agency should provide clear expectations regarding performance monitoring, re-evaluation
triggers and thresholds, remediation requirements, and appropriate regulatory actions when the
regulator, sponsor, or other stakeholder identifies performance degradation.

Finally, FDA should exercise caution when drawing analogies between GenAI evaluation and the
education, training, or credentialing of healthcare professionals.
Competency is a characteristic of
individuals, whereas software medical devices are evaluated based on demonstrated performance
for defined tasks and intended uses. While such analogies may be useful for illustration, regulatory
requirements should remain focused on evidence supporting the device's safety and effectiveness
for its intended use.

Q8. How might the two-axis risk framework described in Section IV be considered within the
competency-based approach described above to help determine the level of evidence
needed for a premarket submission?

A8. ACR agrees that evidence requirements should vary based on the intended use, level of device
activity, and potential consequences of reliance on incorrect outputs. If FDA adopts the proposed
framework, the agency should first establish clear placement criteria and provide illustrative
examples before proposing corresponding evidence requirements. Evidence expectations should
then use the contextual risk modifiers described in our response to question 1.

For imaging applications, ACR believes FDA should require clinical confirmation whenever the
intended user cannot independently assess the accuracy of the output, regardless of where the
function falls on the activity axis of the framework. This distinction helps ensure that the agency
does not treat functions, such as generative image reconstruction, that indirectly inform diagnoses
as low-risk informational tools solely because they do not directly offer diagnostic
recommendations. Instead, the level of evidence should reflect the user's ability to detect
consequential errors and the potential downstream impact on clinical decision-making.

Q9. Would a benchmarking structure such as the one described above be likely to provide
adequate evidence of clinical knowledge, analytic capabilities, safety behavior,
communication, and generalizability to support a reasonable assurance of safety and
effectiveness? Are there elements that are missing, redundant, or inappropriately
categorized? Are there externally developed standards that could be leveraged?

A9. FDA should evaluate failures that matter clinically rather than generic hallucination rates,
nondeterminism, or textual similarity. Medical societies like ACR can help define meaningful
categories, and any such requirements should also scale with autonomy. To that end, there are two
gaps that are missing or inadequately addressed with respect to imaging GenAI specifically—an
imaging failure taxonomy and a scoring architecture.

With respect to compiling a failure taxonomy, this must have testability, operational definitions,
severity classifications, reference evidence per category, and thresholds. FDA should partner with
medical societies, such as ACR, to build and maintain that taxonomy for relevant subdomains. For
imaging, clinically meaningful failures can include:
● laterality errors,
● localization errors: correct finding, wrong site, or wrong level,
● fabricated or incorrect comparison to a prior, including comparison to a study that does not
exist,
● measurement hallucination, or a correct measurement attached to the wrong lesion,
● omission of a clinically significant incidental finding,
● incorrect study, protocol, or technique attribution,
● report-image discordance,
● failure to communicate critical or actionable results,
● erroneous or absent follow-up recommendations, including recommendations
inconsistent with published guidelines,
● anatomic mislabeling, and,
● for reconstruction and synthesis: structures introduced that are not in the acquired data,
and true findings lost or attenuated.

Regarding scoring architecture, the discussion paper does not clearly indicate whether FDA might
intend for performance assessment to be compensatory or non-compensatory. As written, strong
performance in one domain could offset poor performance on clinically important safety measures
within a composite score. ACR recommends that clinically important safety measures carry
prespecified minimum performance thresholds that cannot be offset by strong device performance
in unrelated evaluation domains. This does not mean that every observed safety error should result
in automatic failure; rather, FDA should define appropriate endpoints, performance thresholds,
and uncertainty bounds. The goal should be to avoid aggregate scores that mask unacceptable
safety-related performance.

FDA should also clarify the regulatory implications of failing to meet a safety threshold, including
whether the agency would require additional evidence, risk controls, labeling restrictions, narrower
intended uses, or other actions. The relevant regulatory question is whether the device's
demonstrated performance supports a favorable benefit-risk profile for the proposed intended use.

With respect to external standards and clinical guidelines, FDA should encourage the use of
independently developed, specialty-specific standards where they exist. For medical imaging, the
ACR Appropriateness Criteria® are an example of the guideline-based frameworks described in the
discussion paper. Developed through evidence review, multidisciplinary expert input, and
structured consensus processes, such resources can provide clinically grounded reference
standards for evaluation rubrics, performance measures, and assessment of GenAI outputs.
Existing experience using the ACR Appropriateness Criteria as a reference standard for evaluating
LLM-generated imaging recommendations also demonstrates how established clinical guidelines
can support more clinically meaningful assessment than generic benchmark metrics alone.

Q10. CDRH recognizes that publicly available benchmarking assets may be subject to data
contamination, saturation, and limited real-world representativeness. How should a sponsor
establish that performance on a given benchmark predicts safe and effective real-world
behavior for the device’s intended use? What evidence should support the construct validity
of a benchmark used to gate device evaluation, and what role should sponsor-developed
benchmarks play given potential concerns around independence and optimization to the
test?
A10. ACR agrees that benchmark datasets should be representative of the intended use, patient
population, input data, and real-world deployment environment. Clinical confirmation
demonstrating predictive value for real-world performance should corroborate benchmark
performance.

For imaging applications, FDA should encourage the use of both clinically representative patient
data and appropriately validated data phantoms. Patient data are necessary to evaluate
performance under real-world conditions, while phantoms can provide controlled reference
standards for specific technical characteristics for which assessment using clinical cases alone is
inadequate.

Assessments may use sponsor-developed benchmarks for evaluating specialized use cases or
addressing gaps not covered by publicly available resources. Nonetheless, these benchmarks
should not serve as the sole basis for demonstrating safety or effectiveness, particularly for higherrisk functions. FDA should consider requiring independent evaluation datasets, independent
clinical review, or other external validation measures when sponsors use benchmark results to
support higher-risk claims. At a minimum, sponsors should provide transparent documentation
regarding benchmark construction, evaluation methodologies, adjudication processes,
acceptance criteria, known limitations, and measures taken to maintain independence between
development and evaluation activities.

FDA should also address potential overlap between training and evaluation data. While the extent
of benchmark contamination in medical AI evaluation may not always be known, sponsors should
disclose known overlaps, characterize remaining uncertainty, and describe measures taken to
reduce bias in performance estimates.
Additional independent, temporally separated, or realworld clinical datasets may be needed if meaningful overlap cannot be excluded. Assessors should
also periodically refresh or update benchmark datasets to preserve their long-term validity as
evaluation tools.

Q11. CDRH is considering that clinical confirmation for a GenAI-enabled device might not
require a prospective clinical study in every case, and has described above a range of
approaches of increasing rigor and patient exposure. How might a sponsor select and justify a
confirmation approach tailored to a device’s intended use and proportionate to the device’s
risk profile? Are there device types or risk profiles for which one or more of these approaches
would be insufficient or inappropriate? How might the anticipated distribution of real-world
inputs be taken into consideration? Are there other methods of clinical confirmation that
might help inform the evaluation of GenAI-enabled devices?

A11. The rigor of clinical confirmation should increase with the degree of autonomy, reliance on the
output, and potential patient impact. Functions that are autonomous, patient-facing, or otherwise
intended for use without effective qualified end-user oversight warrant more rigorous clinical
confirmation than functions whose outputs are qualified end-user-vetted before influencing care.

FDA should also clarify that clinical realism, evaluation rigor, and exposure to device-influenced
care are distinct concepts that do not necessarily increase together. For example, a shadow
deployment may involve no exposure to device-influenced care while still evaluating performance
on real patients in a real clinical workflow. Separating these dimensions would improve
consistency in selecting appropriate clinical confirmation approaches.

Sponsors should define the anticipated deployment environment and demonstrate adequate
coverage of the expected input distribution, including patient demographics, care settings,
equipment manufacturers and models, acquisition protocols, and other variables material to the
intended use.

Finally, FDA should require clinical confirmation using real patient data whenever the intended
user cannot independently assess the accuracy of the output.
This consideration may be
particularly important for functions involving generative reconstruction, synthesis, or other outputs
where clinically consequential errors may not be readily detectable.

Q12. How should sponsors achieve statistically meaningful performance measurement for
GenAI-enabled devices? Where synthetically generated inputs supplement real patient data,
how should sponsors account for differences between the synthetic and real-world
distributions when estimating performance, and under what conditions, if any, is it
appropriate to combine benchmarking evidence and clinical confirmation evidence to
support a single performance estimate?

A12. Sponsors should anchor clinical performance claims in representative real-world patient
data. Synthetic data, digital phantoms, and other simulated inputs may support development,
testing, and certain validation activities, but their evidentiary value depends on the claim being
supported, the availability of known reference truth, and their representativeness of the intended
use environment. Accordingly, FDA should require that the sponsor separately evaluate
performance on real-world and synthetic data.

Regardless of data source, FDA should require sponsors to: (1) prespecify clinically relevant
endpoints, acceptance criteria, sample-size justification, confidence bounds, and statistical
analysis methods; (2) clearly define the unit of analysis and account for dependence across
patients, sites, readers, conversations, or repeated model runs; (3) prespecify relevant subgroups
and evaluate clinically consequential subgroup claims on representative data; and (4) document
the deployed system configuration, including model version, prompts, retrieval mechanisms,
guardrails, and other components that may influence performance.

FDA should also address performance variability. For GenAI systems, identical inputs may produce
different outputs across repeated runs. Sponsors should therefore assess variability in clinically
meaningful conclusions and safety-related outcomes using prespecified acceptance criteria and
uncertainty bounds. Variability reporting should inform the safety and effectiveness assessment
but should not substitute for meeting performance requirements.

Finally, benchmarking and clinical confirmation serve different evidentiary purposes, and sponsors
report these separately. Combining results may be appropriate only when sponsors justify
comparability, weighting, and statistical assumptions, and demonstrate that the resulting estimate
remains representative of the intended use population and deployment environment.
Q13. For which clinical domains, device functions, or subpopulations is synthetic data
particularly well-suited, or particularly inadequate, as a supplement to real-world evidence?
What safeguards would mitigate the risk that synthetic data generated by models of the same
class as the device under evaluation reproduces the very performance gaps the evaluation is
intended to detect, particularly for underrepresented subgroups?

A13. Representative real-world patient data should remain the primary basis for demonstrating
clinical performance. Synthetic data appropriateness depends on the claim the sponsor intends
for that data to support. FDA should avoid treating all synthetic data as a single category.

Different uses of synthetic data warrant different evidentiary standards:

● Physics-based simulations and digital phantoms may provide controlled reference truth
but require validation of their assumptions and realism.
● Rare-event, adversarial, and stress-testing applications may reveal important failure
modes but do not establish real-world performance or prevalence.
● Training augmentation should be evaluated based on performance improvements
demonstrated on independent real-world datasets. Such evidence supports the value of
the augmentation but does not establish the absence of unintended effects.
● Clinical performance assessment using synthetic data should require a justified
relationship to the intended-use population, with synthetic and real-world results reported
separately.

FDA's concern should extend beyond reproducing known performance gaps. Synthetic data
generation may introduce, amplify, suppress, or otherwise alter clinically relevant attributes that
were not the intended target of manipulation. Accordingly, sponsors should evaluate potential
unintended changes relevant to the proposed use, characterize the limits of those analyses, and
disclose remaining uncertainty.

Sponsors should also disclose the synthetic data generation process, including the generator,
available information regarding training-data provenance, the relationship between the generator
and the evaluated device, and any material uncertainties they cannot resolve. If sponsors generate
synthetic data recursively or to augment future training datasets, they should assess and
document potential impacts on diversity, representativeness, and performance.

Finally, clinically consequential subgroup claims should be confirmed with independent real-world
data.
Sponsors should also demonstrate that their evaluation approach can detect relevant device
failures despite potential dependencies arising from shared models, datasets, assumptions, or
evaluation tools.

Q14. For open-ended device outputs, how should performance comparators and acceptance
criteria be selected? When a panel of qualified clinicians serves as the comparator, how
should the applicable standard (for example, the standard of care versus the performance of
a median clinician in practice) be defined and justified? Should generalist or specialist
physicians be used as a performance standard? When should human-AI team performance,
rather than the device operating alone, serve as the basis for evaluation?
A14. FDA should distinguish among three separate questions that may require different evidence:
(1) whether an output is clinically acceptable, (2) how performance compares to existing clinical
practice, and (3) whether the device has met minimum safety and effectiveness requirements.
These considerations can coexist, but one should not substitute for another.

Clinical acceptability should be evaluated against either an objective reference standard where
one exists (e.g., pathology, adjudicated outcomes, or follow-up imaging) or through independent
expert adjudication using predefined criteria. For open-ended outputs, multiple responses may be
clinically acceptable, and qualified expert panels may be better suited to defining acceptable
ranges of performance than a single "correct" answer.

Comparator selection should reflect the intended use, deployment setting, and clinician who
would otherwise perform the task without the device. For imaging interpretation, the appropriate
comparator would typically be the radiologist who would otherwise perform that function. FDA
should clarify the distinction between "standard of care" and "median clinician performance," as
these concepts are not equivalent and may lead to different conclusions regarding benefit and risk.
Sponsors should justify the selected comparator based on the clinicians and workflows the device
is intended to support.

Finally, device-only performance and human-AI team performance are separate considerations
whenever performance depends materially on user interaction, oversight, or workflow integration.
Standalone device accuracy does not necessarily predict performance within a clinical workflow,
particularly for GenAI functions where automation bias, plausible omissions, or editing burden may
affect real-world use. The extent of workflow evaluation and human oversight should be
proportionate to the device's intended use, autonomy, and risk profile.

Q15. Are there ways in which performance might be assessed relative to the care, technology,
or course of action likely to occur in the absence of the device, rather than to the comparators
described in this section? What approaches might be used to identify and justify a
comparator such as unaided clinical judgment, delayed specialist review, or no intervention?

A15. The value of a GenAI-enabled device should be assessed relative to the care that would
realistically occur in the absence of the device. The appropriate comparator may vary across
intended uses and deployment settings. In some environments, the relevant comparator may be
prompt specialist interpretation; in others, it may be a delay of specialist review, preliminary review
by another clinician, or no timely interpretation. As a result, the same device performance may
produce different benefit-risk profiles across different practice settings.

FDA should therefore require sponsors to justify the selected counterfactual using evidence from
the intended deployment environment, including existing workflows, practice patterns, resource
availability, and relevant clinical data. The sponsor should include in intended use/labeling all
claims based on delayed, limited, or absent specialist access.

At the same time, a favorable counterfactual should not substitute for independent
demonstrations of accuracy, safety, and effectiveness. Assessments of clinical acceptability,
comparisons to existing care, and minimum safety assurances address different regulatory
questions, and each should be evaluated separately.
Finally, because the validity of a counterfactual depends on the deployment environment, FDA
should consider postmarket monitoring when authorization relies on assumptions regarding site
capabilities, specialist availability, or clinical workflow. Continued use outside the settings
supporting the original benefit-risk assessment may warrant additional evaluation.

Q16. Should independent third parties be involved in some or all aspects of a competencybased approach, including device benchmarking and clinical confirmation? If so, in what
ways might qualified, independent third-party participation contribute to this approach, and
what qualifications and independence criteria should apply? Are there aspects of the
assessment for which third-party involvement would be impractical or inadvisable? What
safeguards or program design features would be critical to prevent third-party participation
from limiting competition or preventing innovation?

A16. ACR supports the use of qualified, independent third parties for benchmarking, clinical
confirmation, and postmarket assessment. For purposes of this comment, we define a qualified
third-party as “a trusted, unconflicted entity that is independent of both the manufacturer and the
regulator.” At a minimum, FDA should require third-party assessors to have the following
characteristics of independence:

● disclosure of relevant financial relationships at both the organizational and individual
reviewer levels.
● individual recusal requirements with an FDA-defined look-back period.
● transparent funding arrangements insulated from assessment outcomes.
● organizational separation between assessment activities and related commercial services
provided to the same sponsors.
● prespecified evaluation protocols, reporting of results regardless of outcome, and no
sponsor right to suppress findings.
● prohibition on manufacturers evaluating their own devices or those of competitors; and
● multiple qualified evaluators rather than a single gatekeeper.

FDA should require manufacturers to make device outputs and relevant performance information
(e.g., AI outputs and related data) available in standardized, interoperable formats (e.g., DICOM
Structured Reports or HL7 FHIR) that support timely independent evaluation with near-real-time
release rather than batch releases. Without such access, meaningful and timely third-party
assessment is not feasible.

ACR's Assess-AI initiative, the first nationwide imaging AI data service including a registry function,
represents existing postmarket monitoring infrastructure that may help inform future independent
assessment approaches. ACR would also support development of independently governed
evaluation datasets and adjudication services, with validation, governance, funding, and
operational capacity established before such resources were used by sponsors/FDA to support
regulatory decisions. ACR's experience with Assess-AI, AI Central, Transparent-AI, and ARCH-AI
recognition may provide useful perspectives in this area. However, any organization serving a thirdparty assessment role should be subject to the same independence, governance, transparency,
and conflict-of-interest requirements.
Finally, FDA should distinguish between assessment, certification, accreditation, and FDA
authorization of devices, as these activities serve different purposes. Certification-style
examinations or sequestered testing assets may help address benchmark contamination
concerns, but they do not by themselves establish safe and effective performance of a device in
clinical practice. Performance on clinically relevant real-world tasks should remain the primary
basis for regulatory decision-making.

Q17. CDRH is exploring whether a competency-based approach could be applied to devices
that incorporate a variety of underlying model architectures, such as multimodal visionlanguage models and generative or predictive world models. Are there aspects of the
described approach that might be ineffective or inapplicable to these devices? What
additional or distinct considerations might need to be addressed?

A17. FDA should clarify how the framework applies to generative reconstruction, enhancement,
and image synthesis functions. While these functions may not directly provide diagnostic
conclusions, they generate or modify the images upon which subsequent clinical decisions
depend. As a result, a function's apparent level of activity may not accurately reflect its risk.

For reconstruction and synthesis functions, FDA should require evaluation using datasets with
known reference truth, including appropriately validated phantoms, inserted lesions, or similar
methods. Assessment should specifically evaluate whether the function introduced, altered,
obscured, or lost clinically relevant findings.

Additionally, FDA should not assume the reading radiologist will always catch a reconstruction
error. Instead, device-specific evidence should be required before crediting visual review as a
safeguard.

Q18. CDRH is considering whether it may be appropriate to accept greater premarket
uncertainty regarding a GenAI-enabled device’s benefit-risk profile through greater reliance
on postmarket monitoring. Under what conditions might such an approach be appropriate,
and what characteristics of a monitoring program would need to be in place to justify reduced
premarket evidence? Are there device types or risk profiles for which this approach would not
be appropriate?

A18. In certain circumstances, greater premarket uncertainty may be acceptable when
counterbalanced by robust postmarket monitoring. However, the appropriateness of this approach
depends on the device's intended use, degree of autonomy, availability of qualified end-user
oversight, and the potential consequences of error. This tradeoff is more defensible when a
qualified end user reviews each output before it influences patient care and is less appropriate for
patient-facing or autonomous functions.

Deferral is appropriate only when postmarket monitoring can reliably detect clinically meaningful
failures, performance drift, subgroup effects, and emerging risks in time to prevent or limit patient
harm. Sponsors should identify the signals they are monitoring, the underlying data sources, their
expected detection timelines, and corrective action plans.
FDA should require prespecified performance thresholds, remediation timelines, notification
procedures, and conditions for deployment restrictions or withdrawal. Sponsors should also
establish workable mechanisms for obtaining the data necessary to support monitoring and
demonstrate that adequate resources exist to sustain those activities.

Finally, greater reliance on postmarket evidence may be inappropriate where unresolved
uncertainty involves the potential for serious, irreversible, or catastrophic harm. Even when such
harms are theoretically detectable through monitoring, the inability to intervene before meaningful
patient exposure occurs may warrant additional premarket evidence.

Q19. Please comment on the potential approaches to postmarket performance evaluation,
including periodic re-benchmarking, sample-based clinician review, and performance
degradation monitoring. What additional approaches should CDRH consider, and how should
the cadence and triggering events for reassessment be determined?

A19. ACR supports the use of re-benchmarking, sample review, and drift monitoring as
complementary components of a lifecycle oversight framework. These approaches serve different
purposes, and FDA should not consider these interchangeable. Re-benchmarking can identify
performance regression against established evaluation criteria, sample review can identify
unanticipated failure modes, and drift monitoring can identify changes in the population, clinical
environment, or device behavior over time.

Effective monitoring requires access to real-world clinical data and workflows that are often not
available to manufacturers alone. Therefore, partnerships with organizations that hold those data
are necessary. ACR's Assess-AI provides existing infrastructure for this purpose. Participating sites
submit de-identified AI outputs, report text, and study metadata through locally installed software.
Concordance against report-derived surrogate labels is then computed centrally, and sites receive
longitudinal performance trends and benchmark comparisons and can utilize local reidentification capability for case review. Assess-AI currently supports classification use cases such
as intracranial hemorrhage, pulmonary embolism, and large-vessel occlusion. ACR intends to
extend this architecture to GenAI outputs in the future.

Monitoring programs should capture all eligible cases, including processing failures, missing
outputs, abstentions, raw outputs before clinician editing, and clinically relevant downstream
outcomes.
At the same time, data completeness and other limitations must be transparently
reported. Sponsors should also conduct impact assessments for material changes to models,
prompts, retrieval systems, orchestration components, guardrails, or other elements that may
affect performance.

Finally, reassessment should occur on a risk-informed basis, triggered by predefined events,
including safety signals, performance drift, significant model or workflow changes, or established
time intervals. Assessors should periodically refresh or rotate evaluation datasets to reduce
overfitting from repeated testing. Disclosures of important methodological limitations, including
imperfect reference standards, incomplete data capture, and validation efforts that remain
ongoing, should accompany the monitoring results.
Q20. Please comment on whether the proposed approaches to postmarket monitoring can be
facilitated by machine-based supervisory agents. What considerations, including the
evaluation and reliability of the supervisory agent itself, should CDRH take into account for
such an approach?

A20. Machine-based supervisory agents will be necessary to evaluate and monitor GenAI-enabled
devices at the scale of modern healthcare workflows, including medical imaging. However, FDA
should treat supervisory systems as safety-critical components subject to independent validation
and ongoing oversight. The acceptable degree of machine supervision should vary based on the
device's intended use, degree of autonomy, and whether the device is patient-facing or providerfacing.

We caution that LLM-based evaluators may exhibit preferential scoring, including favoring outputs
produced by similar models or by the evaluator itself. Agreement between machine supervisors
and human reviewers does not necessarily establish safety and reliability. Moreover, the supervisor
is an attack surface, and short adversarial phrases appended to a text can push judge LLMs to
maximum scores regardless of content. Therefore, FDA should require the following:

● Disclose the supervisor's base model, lineage, and available training provenance, including
what is unknown. Shared lineage warrants scrutiny, and different lineage does not establish
independence. Sponsors should have to measure correlated failure of the supervisor.
● Evaluate for rare consequential failures and abstention behavior rather than overall
agreement. A supervisor that never identifies uncertainty may not be functioning as an
effective reviewer.
● Measure how often the supervisory system fails to identify consequential device errors and
whether the addition of supervision meaningfully improves safety relative to the
unsupervised device.
● Validate the adjudication method against explicit clinical acceptability criteria, including
cases where every candidate output is unsafe. Test adversarial and out-of-distribution
inputs before the sponsor uses the supervisory system to support safety or effectiveness
claims.
● Apply the same drift monitoring as the device, and have humans audit a defined sample of
the device’s judgments on an ongoing basis.
● For panels, measure joint failures and the panel's added value over single evaluators.
Neither the number of agents nor lineage diversity establishes an adequate control.

Q21. What roles might clinicians, healthcare institutions, professional societies, standardssetting bodies, and other stakeholders appropriately play in postmarket monitoring, and how
can these roles be structured without diffusing manufacturer accountability?

A21. Professional societies should define what gets measured and maintain the measurement
tools. For example, in radiology, ACR can contribute resources with distinct roles, such as AssessAI for post-deployment surveillance and future practice accreditation for deploying healthcare
facilities. This strengthens rather than diffuses manufacturer accountability by ensuring someone
other than the manufacturer is also monitoring performance. Section VI.B recognizes these
complementary roles, but does not specify how FDA would assign measurement, reporting, and
corrective-action responsibilities without shifting the manufacturer's obligations onto clinicians
and institutions. FDA should clarify the above consideration in any future relevant policy.

Additionally, the stakeholder ecosystem is more complicated than suggested in the discussion
paper and question 21. In imaging, one organization may acquire the images, a second may employ
the interpreting radiologist, and a third may operate the AI. For each deployment, stakeholders
must assign named responsibility for investigating alerts, stopping use, notifying affected sites,
reviewing cases already affected, and maintaining care during outages.
Such stakeholders should
settle data-sharing and escalation considerations before going live. Manufacturer accountability
stays intact when an institution or registry does this work.

Finally, FDA should recognize that all stakeholders engaged in these activities need to fund the
monitoring efforts performed, whereas Section VI.B does not address this key component of
establishing an effective and trustworthy monitoring paradigm. The agency should work closely
with identified stakeholders to explore typical costs that enable their surveillance role and consider
where responsibility lies for recovering those costs.

Q22. CDRH envisions that a manufacturer’s premarket competency-based assessment might
serve as a baseline against which post-deployment modifications could be re-evaluated. How
might the extent of re-benchmarking or other evidence be scaled to the nature and expected
impact of a given modification? For example, are there categories of change that might not
significantly affect safety or effectiveness of a GenAI-enabled device, or might be
appropriately managed within a sponsor’s quality management system versus requiring FDA
premarket review and authorization before implementation? Are there categories of changes
that might be appropriate for inclusion in a PCCP?

A22. FDA should move from regulation that effectively freezes specific model versions, prompts, or
configurations toward validated performance envelopes, predetermined change control plans
(PCCPs), and lifecycle oversight, all tiered by what the change does. Changes to models, prompts,
retrieval, orchestration, or guardrails that remain within validated performance parameters should
not automatically require a new submission. Regulation should not create an incentive to retain an
older, inferior model simply because upgrading creates regulatory burden. However, it is important
to note that "newer" and "scores higher overall" are different from “better.” A model can improve its
composite while also decreasing its precision in important clinical tasks, such as escalating a
critical finding or performance for a subgroup. Evidence, rather than recency, should inform
justification for reduced barriers.

There are three categories of modification: (1) changes that need no new submission under the
applicable modification rules are documented and validated within the quality system; (2) changes
covered by an authorized PCCP that may be implemented when they meet that plan's specified
modifications, validation protocol, and acceptance criteria; and (3) changes that require additional
authorization and must undergo premarket review before implementation. Passing a sponsordefined performance threshold alone should not decide the change category. Safety-critical
behaviors and relevant subgroups need their own acceptance criteria, so that aggregate
improvement does not conceal unacceptable regression. The PCCP guidance already calls for
sequestered test data and prespecified criteria, but we recommend independent custody of
evaluation data at higher risk tiers as an added safeguard.
Additionally, the current version of the PCCP guidance does not appear to address GenAI-enabled
devices explicitly. FDA should clarify the conditions under which a PCCP applies, as behavior can
shift through prompts, retrieval, orchestration, guardrails, or an upstream model the sponsor does
not control.

Q23. How might PCCP concepts or other change-control approaches be adapted for GenAIenabled devices when the nature or scope of future modifications cannot be fully
prespecified?

A23. Fundamentally, if a sponsor cannot predetermine the change, including specifying in advance
the modification, validation protocols, and acceptance criteria, a PCCP is not the appropriate
mechanism.

The current FDA guidance on PCCPs requires test data "never shown to the algorithm during
training," "pre-defined acceptance criteria, and statistical tests for each planned modification,"
and a commitment that a modification failing evaluation "will not be implemented." It also requires
modifications that are "specific, and that can be verified and validated” with an impact assessment
component that rests on that specificity. To address the GenAI gap with an extension of the current
architecture, FDA should consider bounded categories rather than open-ended envelopes and
permit a change type with limits such as model version updates within a named family, retrievalcorpus updates from a specified source, and prompt changes within a defined function. Each
change category should have corresponding acceptance criteria and an impact assessment. Then,
anything outside the authorized plan must go through the applicable modification rules again to
determine whether it requires a new submission.

The alternative of prespecifying the test - instead of the change - would represent a policy change
that removes the specificity on which the impact assessment depends. FDA would need to clarify
how an impact assessment works when no modification has been prespecified. Passing the same
benchmark does not demonstrate that new failure modes have been assessed, the intended use
remains unchanged, subgroup performance has been maintained, or changes to the supplier or
system architecture have introduced no new hazards.

In either of the above approaches, notification would be necessary. The sponsor must notify FDA
and all deployment sites whenever they modify the device to enable regulatory oversight and local
governance.

Q24. For devices built on third-party foundation models, changes to the underlying model may
be initiated by the third-party model developer rather than the device manufacturer. How can
a manufacturer detect, evaluate, and respond to such changes in a timely manner? What
mechanisms—for example, contractual, technical, or through a PCCP—could provide
reasonable assurance that third-party developer-initiated changes do not compromise the
safety or effectiveness of the device?

A24. Continuous monitoring is important for identifying downstream impacts and mitigating harm,
but it should be paired with controls that prevent unverified changes from entering clinical use.
Foundation models and related services can change over time, even when presented as the same
product. Manufacturers therefore need clear visibility into which elements of the deployed
configuration they control, and which may change upstream.

FDA should require appropriate supplier controls, configuration management, change detection,
pre-deployment verification, and a tested fallback process. Agreements with suppliers should
address behavior-affecting changes and model retirement and provide sufficient notice to
complete verification.
When a device can no longer maintain a validated configuration, it should
transition to a validated alternative or safely suspend the affected function.

FDA can regulate the device's integrated clinical function while requiring version identification,
supplier transparency, change notification, and configuration control. As the discussion paper
notes, a foundation-model master file would not indicate authorization of the underlying model for
a device intended use. These measures reflect established supplier-control principles rather than
direct regulation of foundation models as a class.

Finally, effective local performance monitoring requires deploying organizations to have access to
device outputs and relevant metadata in a structured, machine-readable format. FDA should
require manufacturers to provide such access when authorization depends on local or postmarket
performance monitoring.

Q25. Would voluntary Foundation Model MAFs be practical for foundation model developers
to provide and to keep current, and what content might they include? Given that participation
would be voluntary and that model developers may have limited incentive to disclose safetyrelevant information, what would make such a program sufficiently useful for premarket
review? Are there alternative mechanisms CDRH should consider for obtaining information
about underlying foundation models?

A25. A foundation model master file (MAF) could provide FDA access to confidential component
information without requiring public disclosure. However, because participation would likely be
voluntary, FDA should not assume that a MAF alone will consistently provide the information
needed for device review. A MAF should identify the covered model version and provide
information, as appropriate, on training-data provenance, healthcare-relevant evaluation results,
known limitations and failure modes, safety controls, update practices, and audit-log availability,
with the level of detail proportionate to its relevance to the device's safety and effectiveness. If FDA
determines that foundation model information is necessary for device review, sponsors should be
required to provide that information directly or through an authorized reference to a current MAF.

Q26. Are there additional considerations that inform the premarket and postmarket
evaluation of agentic GenAI-enabled devices, beyond those applicable to non-agentic GenAIenabled devices? How should the elevated risk associated with autonomous multi-step
action, tool use, and reduced opportunity for human review be reflected in acceptance
criteria and oversight?

A26. FDA should refine its definition of agentic AI and clarify whether the agency would intend to
evaluate the characteristics identified in the discussion paper individually or in combination. FDA
should assess risk based on a system's capabilities and degree of autonomy, including its ability to
take actions, access external information or tools, operate across multiple steps, and perform
actions that are difficult to reverse.

FDA should also distinguish between autonomy and verifiability. If a qualified end-user cannot
independently verify a system’s outputs, the system should not receive credit for nominal human
oversight.

As risks may arise during task execution even when the final result is correct, evaluation of agentic
systems should assess both outcomes and agent actions leading to those outcomes. Testing
should examine failure modes such as incorrect patient or study selection, duplicate actions, stale
or incomplete data, failed handoffs, partial task execution, and variability across repeated runs of
the same task. Sponsors should evaluate repeated runs of the same task and report meaningful
variability in outcomes and intermediate actions rather than solely final outputs.

Agentic systems may introduce cybersecurity and prompt-injection risks through user inputs,
images, retrieved content, tool outputs, and other external data sources.
FDA should align agenticAI expectations with existing cybersecurity requirements, including testing of these attack
pathways and application of least-privilege principles for tool access and external actions.

Controls should be proportionate to the consequences and reversibility of the actions a system
can perform. Drafting an order, submitting it, and submitting it irreversibly present different risks
and may warrant different levels of validation and oversight.
Higher-risk functions may warrant
clinically representative validation, phased deployment, postmarket monitoring, and reversibility
analysis. Where human supervision is a claimed safeguard, FDA should evaluate that interaction
directly and define the supervisor's role.

As always, the ACR welcomes continued dialogue with the FDA on these and other radiologyrelevant topics. For more information, please contact Michael Peters, Senior Director, ACR
Government Affairs, at mpeters@acr.org.

Sincerely,

Dana H. Smetherman, MD, MPH, MBA, FACR
Chief Executive Officer