American Hospital Association (Ashley Thompson)

IndustryTrade associationFiled October 19, 20262,589 words · 1 attachmentFDA-2026-N-7874-0380
LinkedInX

Themes it raises

12 of the 21 themes in the docket, each with the passage we counted, verbatim.
Whether the user can judge the outputFDA Q3, Q4
“Additionally, this should specify the level to which healthcare professionals are in the decision loop and the type of clinician (general versus specialist) involved.”
Judging devices the way clinicians are credentialedFDA Q7, Q8
“The AHA agrees in concept with a competency-based approach for premarket evaluation, but encourages clarity of sponsor and developer requirements.”
Trading premarket certainty for postmarket monitoringFDA Q18
“opportunities exist to better synchronize post-deployment measurement and evaluation activities in ways that could reduce unnecessary premarket barriers.”
Watching the device after it shipsFDA Q19, Q20
“Standards should include metrics for performance, thresholds for further evaluation, and communication requirements to end users related to ongoing performance.”
Who is accountable when something goes wrongFDA Q21
“vendors and developers must be responsible for the ongoing performance, maintenance and integrity of the tools they market.”
How this fits rules that already existFDA Q8, Q9, Q16, Q25
“The FDA also should clarify that measurement and evaluation standards would not apply to clinical decision support tools and administrative AI tools.”
What the rules cost sponsors and the marketNot asked by the FDA
“This approach may limit the ability to expand otherwise safe and effective uses of AI and result in added costs to vendors and providers alike, as application fees have often been passed down to end users through increased prices.”
What makes a function high riskFDA Q1, Q2, Q5
“The AHA also recommends that the FDA add variables for likelihood or probability of consequences, which are currently missing from the risk matrix and discussion paper.”
Escalating too little and too muchFDA Q6
“A recent Nature article found that certain commercial AI tools over-triaged non-emergent cases and under-triaged emergency conditions.”
Whether benchmark results prove anythingFDA Q9, Q10, Q16
“If this were pursued, we encourage the agency to explore models where multiple third parties could support instead of a single third party.”
Proving the device works in real careFDA Q11, Q12, Q13, Q14, Q15
“Instead of relying on a panel, we encourage the agency to leverage existing consensus-based clinical practice guidelines. Standards should be transparent and should be denoted in labelling.”
Equity, access and under-resourced settingsFDA Q3, Q13, Q21
“Additionally, some hospitals (particularly rural, critical access and other safety net hospitals) may not have the staff or resources to support governance structures and ongoing measurement activities.”
Machine-assisted draft, pending human review. The source text and highlighted passages appear below. Read the filing on regulations.gov ↗

The comment as filed

Comment submitted on regulations.gov. Passages we counted are highlighted.

Please see the AHA comments attached: FDA-2026-N-7874 Considerations for the Regulation of Generative AI-Enabled
Medical Devices: Discussion Paper and Request for Feedback

Attachment

Attachment, text extracted from the filed document. Passages we counted are highlighted.

October 19, 2026

Michelle Tarver
Director
Center for Devices and Radiological Health
Food and Drug Administration
10903 New Hampshire Ave.
Silver Spring, MD 20993

Submitted Electronically

RE: FDA-2026-N-7874 Considerations for the Regulation of Generative AI-Enabled
Medical Devices: Discussion Paper and Request for Feedback

Dear Director Tarver,

On behalf of our nearly 5,000 member hospitals, health systems and other healthcare
organizations, our clinician partners — including more than 270,000 affiliated
physicians, 2 million nurses and other caregivers — and the 43,000 healthcare leaders
who belong to our professional membership groups, the American Hospital Association
(AHA) appreciates the opportunity to provide feedback to the Food and Drug
Administration (FDA) on regulatory considerations for generative artificial intelligence
(GenAI)-enabled medical devices.

AI-enabled healthcare applications are transforming care delivery in innumerable ways,
holding tremendous promise for improving healthcare outcomes and reducing costs. As
GenAI technologies advance, policymakers will grapple with novel challenges related to
model bias, hallucinations and model drift, which are not yet fully accounted for in
existing medical device frameworks. AI tools are inherently designed to be agile and
adaptive, taking in new data points, discerning patterns and continually updating to
improve model accuracy. This is especially true for GenAI. The FDA will face questions
about which GenAI applications constitute medical device versus non-device functions,
and how the agency’s regulatory frameworks should account for these tools.

In general, the AHA supports a flexible, risk-based regulatory framework for GenAIenabled medical devices that protects patients without impeding responsible innovation.
To achieve this balance, the AHA recommends that the FDA:
Director Michelle Tarver
October 19, 2026
Page 2 of 8

• Refine the proposed risk heuristic for evaluating the risk of GenAI applications.
• Provide labelling guidance on “informational action directing” devices versus
“action taking” devices.
• Align premarket clearance and post-deployment evaluation and monitoring
activities under a total product lifecycle approach.
• Clarify that GenAI-enabled clinical decision support and administrative tools are
not subject to FDA oversight.
• Assign primary responsibility for premarket and post-deployment monitoring and
evaluation activities to developers.

Below are our detailed comments.

CONSIDERATIONS FOR RISK ASSESSMENT

FDA’s discussion paper proposes a risk-based approach to generative AI-enabled
medical device evaluation. The request for information (RFI) seeks feedback on a
conceptual framework for assessing risk, which the FDA has structured as a heuristic
comprised of two variables. The first variable reflects the “activity level” of the GenAIenabled device, ranging from providing informational but non-directive outputs to a
device user to fully autonomous action. The second variable reflects the
“consequences” of the GenAI-enabled medical device’s actions, ranging from limited to
severe. Figure 1, drawn from the RFI, includes a diagram of this risk-based approach.

Figure 1. FDA’s proposed 2-variable risk heuristic for generative AI-enabled
medical devices

Below is our detailed feedback on the proposed risk heuristic.

Expand Risk Heuristic to Account for Variables Like Probability of Consequence

The AHA generally supports the pursuit of risk-based approaches for the evaluation of
generative AI-enabled medical devices. However, we recommend the FDA add
Director Michelle Tarver
October 19, 2026
Page 3 of 8

variables to its proposed heuristic, such as reversibility, downstream safeguards and
output traceability in its risk-based evaluation. The AHA also recommends that the FDA
add variables for likelihood or probability of consequences, which are currently missing
from the risk matrix and discussion paper.
The likelihood of a negative outcome, error
(type I or II) or other consequence is a vital component of risk evaluation, especially in
the context of statistical tools like GenAI.1

Update or Remove Proposed “Activity Level” Variable

The FDA’s proposed framework correctly recognizes that the activity level of GenAI
devices exists on a spectrum. However, activity level alone may not sufficiently predict
the risk of particular devices, and the relationship to risk and activity level is not as linear
as the FDA’s heuristic suggests. As such, the AHA encourages the FDA to reconsider
the structure of this variable.

For example, the discussion paper suggests that “informational action directing”
activities may include a “strongly worded patient-facing recommendation about whether
to seek emergency care.” Recent studies of certain commercial products on triaging
efficacy demonstrate a range of risks or consequences for these tools. A recent Nature
article found that certain commercial AI tools over-triaged non-emergent cases and
under-triaged emergency conditions.
2 Specifically, the tools had a 52% mis-triage rate
for certain patients that required escalation to an emergency department (e.g., diabetic
ketoacidosis patients or patients with impending respiratory failure). At the same time,
the tools over-triaged 64% of non-emergent cases. As the article states, “The 51.6%
undertriage rate for true emergencies represents the most concerning finding, as
missed emergencies can result in patient harm, while the 64.8% overtriage rate for
nonurgent cases, although less dangerous, risks unnecessary healthcare utilization at
scale.”3

In the context of the FDA’s heuristic, this study demonstrates that “informational action
directing” applications, in and of themselves, do not necessarily present lower
consequences or risk than “action taking HCP-supervised applications.” As the article
states, “The undertriaged cases—rising pCO2 signaling respiratory failure, metabolic
acidosis in DKA—are presentations that no experienced clinician would delay.”4 For
these reasons, the FDA’s heuristic should consider variables, such as likelihood of
error, negative outcome or consequence, as more heavily weighted than whether an
activity is informational versus action-directing.
In statistical models, a “type I” error denoted by α occurs when there is a false positive result (i.e.,
incorrectly rejecting a null hypothesis). A “type II” error denoted by β occurs when there is a false negative
result (i.e., incorrectly failing to reject a null hypothesis).
https://www.nature.com/articles/s41591-026-04297-7
Ibid.
Ibid.
Director Michelle Tarver
October 19, 2026
Page 4 of 8

Clarify Labeling Guidance to Distinguish Medical Devices and Non-medical
Devices, and Informational and Action-taking Devices

The example above related to triaging efficacy also underscores the importance of clear
product labelling. As we have previously commented, for consumers, it can be
confusing to distinguish between medical and non-medical device applications, or
devices intended to be informational versus action-directing.5 This is particularly true for
models that rely on similar types of physiological data and generate similar
recommendations or outputs, yet differ in their intended use, validation processes and
regulatory status.

We encourage the FDA to update guidance on labelling in collaboration with the Federal
Trade Commission (FTC). Such guidance should describe how the FDA and FTC would
distinguish between informational and action-taking devices, and suggest disclaimer
language for informational products. Additionally, this should specify the level to which
healthcare professionals are in the decision loop and the type of clinician (general
versus specialist) involved.
Transparency in labelling can better inform consumers of
the extent to which a tool is autonomous and the type of clinician in the loop. The FTC
has a broad mandate to prevent unfair or deceptive acts or practices, and the FTC and
FDA share jurisdiction over marketing of certain health-related products (including
devices).

TOTAL PRODUCT LIFECYCLE APPROACH TO REGULATION OF GENAIENABLED MEDICAL DEVICES

The discussion paper requests feedback on the applicability of total product lifecycle
(TPLC) approaches to the oversight of GenAI-enabled medical devices, an approach
the FDA has long promoted for medical device regulation. In general, the AHA agrees
that risk-based evaluation should be integrated into a TPLC approach for GenAIenabled medical device oversight. A risk-based approach to measurement and
evaluation aligned with TPLC could focus scarce resources, such as time, personnel
and cost, on the highest-risk applications and also coincide with the approach of
domestically accredited and international standards development groups.

GenAI is perhaps one of the most prominent examples where TPLC approaches, and
alignment of premarket clearance requirements with post-deployment monitoring
activities based on risk, would be most beneficial because GenAI models are constantly
evolving. The potential for model drift and hallucinations underscores the importance of
post-deployment monitoring activities in ways that perhaps previous static devices did
not necessitate.
https://www.aha.org/letterscomments/2026-08-13-aha-responds-fda-request-input-development-21stcentury-cures-act-section-3060-required-report
Director Michelle Tarver
October 19, 2026
Page 5 of 8

COMPETENCY-BASED APPROACH FOR PREMARKET EVALUATION

The discussion paper also requests feedback on a two-part competency-based
approach for premarket evaluation comprised of non-clinical device benchmarking and
clinical confirmation. Device benchmarking elements would determine whether a GenAIenabled device demonstrates the clinical knowledge, analytic capabilities, safety
behavior, communication and generalizability to support reasonable assurance of
device safety and effectiveness for its intended use. The agency acknowledges that
benchmarking alone may not fully account for real-world conditions, and therefore
proposes integration of clinical confirmation activities in premarket evaluation to provide
review of clinically representative conditions.

The AHA agrees in concept with a competency-based approach for premarket
evaluation, but encourages clarity of sponsor and developer requirements.
Specifically,
when it comes to clinical confirmation, providers often lack visibility into an AI model’s
underlying weights, making it difficult to determine alignment with real-world patient
populations. Therefore, the agency should clarify device benchmarking and clinical
confirmation activities that are the responsibility of the sponsor/device developer.

The AHA also encourages the FDA not to unintentionally restrict clinicians in utilizing
clinical judgment for tools. The discussion paper references standards of performance
and proposes comparing performance to a “panel of qualified clinicians whose
consensus reflects the applicable standard.” Instead of relying on a panel, we
encourage the agency to leverage existing consensus-based clinical practice
guidelines. Standards should be transparent and should be denoted in labelling.

The FDA also explores the role of third parties in evaluation and benchmarking. If this
were pursued, we encourage the agency to explore models where multiple third parties
could support instead of a single third party.
Having a single third-party evaluator could
limit competition and artificially inflate pricing.

POST-MARKET MONITORING AND EVALUATION

The discussion paper also requests input on post-market monitoring for GenAI-enabled
devices. Our detailed feedback is below.

Synchronize Post-deployment Monitoring Activities With TPLC to Reduce Preclearance Barriers

As previously noted, opportunities exist to better synchronize post-deployment
measurement and evaluation activities in ways that could reduce unnecessary
premarket barriers.
6 For example, currently over 96% of AI-enabled medical devices are
https://www.aha.org/lettercomment/2025-12-01-aha-letter-fda-ai-enabled-medical-devices
Director Michelle Tarver
October 19, 2026
Page 6 of 8

approved under the 510k clearance process, but the 510k process caps the number of
indications for which applicants can seek approval at a given time. By its very nature,
GenAI is adaptable, resulting in new capabilities that could augment the number of
clinical indications for which a particular AI application is suited. Yet, the current 510k
framework generally requires vendors and other developers to submit multiple
clearance applications for substantially similar technologies. This approach may limit the
ability to expand otherwise safe and effective uses of AI and result in added costs to
vendors and providers alike, as application fees have often been passed down to end
users through increased prices.

The AHA encourages the FDA to explore clearance pathways whereby vendors and
other developers can provide detailed post-market evaluation and monitoring plans,
including additional special controls, to streamline 510k clearances under fewer
applications. Such an approach could shorten the time to get evidence-based, safe
uses of AI approved for use, benefiting patients and providers. It could also help reduce
regulatory burden and cost.

Clarify Boundaries of Device and Non-device Definitions

The FDA also should clarify that measurement and evaluation standards would not
apply to clinical decision support tools and administrative AI tools.
The 21st Century
Cures Act excluded certain AI tools from the definition of “medical devices,” so long as
they support clinical decision-making, with the clinician independently reviewing
recommendations. Clinical decision support tools are often developed to address
specific clinical challenges over limited periods. Additional requirements can add
barriers to delivering timely solutions and may be impractical if the tool is only designed
to support clinicians for a short timeframe (like six months). To maximize scarce
resources, the FDA should focus monitoring and evaluation activities on higher-risk
areas related to the diagnosis of disease or other conditions, or in the cure, mitigation,
treatment or prevention of disease.

Place Primary Post-market Monitoring Responsibility on Developers

While hospitals and health systems support evaluation and monitoring activities,
vendors and developers must be responsible for the ongoing performance,
maintenance and integrity of the tools they market.
Hospitals and health systems have
broadly developed multi-disciplinary governance teams to review AI use cases,
technologies and value over time. According to the AHA’s Health IT supplemental
survey, 74% of hospitals have multiple teams responsible for evaluating predictive AI,
including (but not limited to) senior leaders, department leaders and IT staff, with many
having specific committees for machine learning and/or clinical decision support.7
https://healthit.gov/data/data-briefs/hospital-trends-use-evaluation-and-governance-predictive-ai-20232024/
Director Michelle Tarver
October 19, 2026
Page 7 of 8

Hospitals and health systems continually assess the strengths and limitations of AI
models they use, but the “black box” nature of many AI systems can make it more
challenging for hospitals and health systems to identify flaws in models that may affect
the accuracy and validity of an AI tool’s analyses and recommendations. As such, postmarket measurement and evaluation standards should be developed for vendors.
Standards should include metrics for performance, thresholds for further evaluation, and
communication requirements to end users related to ongoing performance.

Additionally, some hospitals (particularly rural, critical access and other safety net
hospitals) may not have the staff or resources to support governance structures and
ongoing measurement activities.
As such, many of these healthcare providers have not
deployed AI tools. Resource and infrastructure barriers can exacerbate the “digital
divide” in certain geographies, where rural and other underserved areas have less
access to digital services, including AI tools for clinicians and patients. While addressing
incentives and broader infrastructure barriers to AI is not the sole responsibility of the
FDA, we encourage cross-agency collaboration to develop training, technical assistance
and potential grant funding opportunities to support educational efforts and governance
activities for AI health tools. This includes education and resources to support
evaluation and monitoring activities. Cross-agency coordination could include
collaboration among Health and Human Services, the FDA, the Federal
Communications Commission, the Department of Commerce, the Department of
Agriculture and the Department of Education.

OTHER TOPICS

The agency also requests feedback on a variety of other related topics like Model
Device Master Files and voluntary submission of AI model cards to the FDA. We note
that transparency of AI model cards has been a part of the Office of the National
Coordinator for Health IT’s (ONC) certified health IT program. We have heard from
hospitals that this criterion is one of the few examples of transparency standards
providing information on AI to help inform tool procurement and implementation. One of
the primary barriers to AI adoption has been the “black box” nature of algorithms. This
criterion provides information on how a predictive or GenAI application was designed,
developed, tested, evaluated and should be used. The AHA opposed the proposed
removal of these criteria in ONC’s HTI-5 proposed rule, and continues to advocate that
transparency of model card data is imperative to foster trust in AI tools and ensure
patient safety.8

The AHA looks forward to working with the FDA on a risk-based approach to GenAIenabled device regulation that protects patients, provides clear expectations for
developers, aligns with operational realities of care delivery and supports responsible
https://www.aha.org/lettercomment/2026-02-27-aha-comments-astponc-health-care-technologyinteroperability-proposed-rule
Director Michelle Tarver
October 19, 2026
Page 8 of 8

innovation. Please contact me if you have questions, or feel free to have a member of
your team contact Jennifer Holloman, AHA director of health IT policy, at
jholloman@aha.org.

Sincerely,

/s/

Ashley Thompson
Senior Vice President
Public Policy Analysis and Development