Q1 · The two-axis risk frameworkQ2 · The spectrum of device activityQ3 · When an output becomes directiveQ4 · Generalist and specialist usersQ5 · Multi-turn conversations that migrateQ6 · Care escalation functionsQ7 · The competency-based approachQ8 · Mapping the risk grid to evidenceQ9 · The benchmarking structureQ10 · Benchmark contamination and saturationQ11 · Clinical confirmation without a prospective trialQ12 · Statistically meaningful performanceQ13 · Synthetic dataQ14 · Comparators and acceptance criteriaQ15 · Performance against usual careQ16 · Independent third partiesQ17 · Devices with many functionsQ18 · Trading premarket certainty for postmarket monitoringQ19 · Postmarket performance evaluationQ20 · Machine-based supervisory agentsQ21 · Clinicians, institutions and societiesQ22 · Re-benchmarking after a modificationQ23 · PCCPs for GenAI devicesQ24 · Third-party foundation model changes
The comment as filed
See attached file(s)
Attachment
October 19, 2026
Center for Devices and Radiological Health
Food and Drug Administration
Department of Health and Human Services
Subject: Response to FDA Considerations for the Regulation of Generative AI-Enabled
Medical Devices: Discussion Paper and Request for Feedback
Docket No. FDA-2026-N-7874
To Whom It May Concern:
Thank you for the opportunity to comment on FDA’s discussion paper, Considerations for the
Regulation of Generative AI-Enabled Medical Devices (U.S. Food and Drug Administration,
2026).
Harrison.ai is a global healthcare technology company that expands clinician capacity and
improves patient care through AI automation. Clinician-led and patient-first, our comprehensive
solutions support earlier, more accurate radiology diagnoses and integrate seamlessly into
clinical workflows. Harrison.ai solutions are cleared in more than 40 countries, available to half
of all Australian radiologists, and clinically deployed at more than 1,000 customer sites globally,
including more than 55 NHS Trusts. Our tools have impacted the care of more than 7 million
patients to date. Harrison.ai has a growing footprint in the United States, including 9 FDA
clearances covering 13 chest x-ray and CT brain indications, 3 FDA Breakthrough Device
designations, and New Technology Add-on Payment (NTAP) reimbursement. We are committed
to supporting innovation, quality, and safety in American healthcare.
Building on our experience developing and deploying multitask AI models, we have invested
heavily in generative AI (GenAI) and foundation model technology. Our mission is to use AI to
help build healthcare capacity worldwide. For example, draft reporting can help address
physician shortages: the technology detects radiological abnormalities and organizes them into a
draft report for the clinician to review. Our investment extends beyond technology development
to methods for evaluating these technologies before and after they reach the market. To support
transparency, we have published details of this work in progress in our Harrison.Rad 1.5
technical report (Mall, et al., 2026), which is included in full as an appendix to our comments.
Our research foundation model has also been evaluated independently, including in a peerreviewed multi-reader study in the American Journal of Roentgenology (Hong, Song, &
O’Riordan, 2026).
We greatly appreciate FDA’s efforts to share its perspectives and encourage broader public
discussion. We recognize that the paper is intended to solicit discussion, does not establish
regulatory expectations, and does not determine whether additional legal authority would be
needed to implement the approaches under consideration. Our comments should be read in that
context.
Page 1 of 12
We generally support AdvaMed’s comments on the discussion paper (AdvaMed, 2026) and
highlight key points from that organization under the relevant questions below. We also offer
additional responses where Harrison.ai can share specific insights from our experience
developing and evaluating GenAI-enabled radiology technologies.
Section IV: Considerations for the Assessment of Risk for GenAI-Enabled
Devices (Q1-6)
Q1: Does the two-axis risk framework, organized around device activity and the
consequence of relying on an incorrect output, appropriately capture the dimensions most
relevant to the risk of a GenAI-enabled software function? If there are additional
dimensions—such as the reversibility of a resulting action, the availability of downstream
safeguards, the time pressure of the deployment setting, or the traceability of the output
(i.e., to primary source materials)—that should be represented in a risk framework, please
describe and provide examples of how they should be represented.
A1: We support AdvaMed’s recommendation that approaches to GenAI should “build
upon the existing foundation of risk assessment and device classification,” such as ISO
14971 (International Organization for Standardization, 2019) and IMDRF N81
(International Medical Device Regulators Forum, 2025), and that “the presence of GenAI
or the degree of autonomy should not be the sole drivers of assessing device risk. Rather,
consistent with existing practice, a device’s risks should be evaluated holistically with
consideration to its intended use and device-specific technological considerations to more
fully contextualize clinical use-related risks” (AdvaMed, 2026).
Q2: CDRH seeks input on how to account for the spectrum of GenAI-enabled
informational functions that vary in the degree to which they direct a user to a particular
action (i.e., between “non-directive” and “action-directing”). What characteristics of an
output—such as its wording, specificity, personalization, or context—could be considered
as modifiers of the risk associated with an informational function, after accounting for the
device’s overall functionality and intended use? What additional information would
provide manufacturers with sufficient clarity and predictability around risk assessment for
informational functions while recognizing that directiveness may exist along a continuum
rather than as a binary distinction?
A2: This question is difficult to answer without specific device examples. Without that
context, some of these outputs may not meet the statutory definition of device software,
as described in FDA’s Clinical Decision Support Software guidance (U.S. Food and Drug
Administration, 2026), or may fall under FDA’s current enforcement discretion policies,
as described in FDA’s General Wellness guidance (U.S. Food and Drug Administration,
2026). More broadly, and beyond informational functions, a device’s risks and benefits
should always be compared against the standard of care, in addition to its intended use,
recognizing that the standard of care is often highly variable and on a spectrum.
Page 2 of 12
Section V: A Competency-Based Approach for Premarket Evaluation of
GenAI-Enabled Devices (Q7-17)
Q7: Is the competency-based approach described above, i.e., device benchmarking followed
by clinical confirmation, a useful and appropriate framework for evaluating GenAIenabled devices?
A7: We appreciate FDA sharing its flexible thinking on least burdensome approaches for
evaluating GenAI-enabled use cases. We support AdvaMed’s comments that “Premarket
evaluation and evidentiary expectations should be calibrated to the device” and that
“FDA should leverage and, when necessary, adapt premarket evidentiary expectations,
methods, and terminology rather than creating new parallel expectations and vocabularies
specific to GenAI” (AdvaMed, 2026).
To build on AdvaMed’s comments, we offer a specific example of how new terminology,
such as a “competency-based” approach, could cause confusion and distract from the
goal of least burdensome approaches that balance patient safety and innovation. The new
terminology gives the impression that “competency assessment” may be a novel, less
burdensome path to market for certain GenAI use cases. However, the “benchmarking”
and “clinical confirmation” steps described in the paper closely mirror existing principles
for medical devices. Examples include “analytical validation” and “clinical validation”
for in vitro diagnostics, and “standalone testing” and “clinical reader studies” for
radiology computer-aided detection (CADe) devices (U.S. Food and Drug
Administration, 2022; U.S. Food and Drug Administration, 2022). If all of FDA’s existing
CADe recommendations, which appear to fit within what FDA describes as “competency
assessment,” were applied to broader GenAI-enabled use cases such as draft reporting,
the path to market would be too burdensome.
Even a standalone-only study, as described in the first CADe guidance above, would be
impractical if it had to be statistically powered for hundreds of findings. This example
leads to two recommendations:
• Rather than developing general terminology and follow-up guidance across many
GenAI use cases, FDA could achieve more by focusing on specific, least burdensome
methods for selected high-impact use cases, ideally through more flexible and timely
outputs such as research articles and Regulatory Science Tools.
• As FDA considers least burdensome approaches for GenAI-enabled use cases (e.g.,
comprehensive draft radiology reports), it should also consider less burdensome
approaches for related outputs that have existed for many years (e.g., alternatives to
clinical reader studies for a narrow CADe device). GenAI, for example foundation
model technology, can make development of more comprehensive CAD type
products more efficient in addition to enabling more novel use cases. In addition,
existing device types may complement more novel use cases; for example,
increasingly we are learning that users may expect localization (CADe type output)
along with a draft report to increase AI explainability. Unless CDRH also considers
less burdensome pathways for CADe, manufacturers may not be able to deliver such
explainable comprehensive products to customers, which could have a negative
Page 3 of 12
impact on patient care. Testing expectations should be driven by the intended use,
benefits, risks, and standard of care, not by the underlying technology (GenAI,
convolutional neural networks, or rule-based algorithms).
Q8-Q15 (performance and benchmarking) summary response: These are the right questions
for FDA to ask about performance and benchmarking, but they can only be answered
meaningfully for specific intended uses.
GenAI enables broad use cases that challenge existing evaluation methods but can also deliver
significant benefits. For example, a comprehensive radiology draft report may cover hundreds of
radiological abnormalities and their attributes. That breadth is what drives radiologist reporting
efficiency, an important benefit given physician shortages. Such a device can only reasonably be
tested using case-based methods, because it is impractical to test every possible output or
finding, even through unpowered subgroup analyses. Asking sponsors to remove findings based
on issues seen in unpowered subgroup analyses can also severely limit a device’s clinical utility
and may confuse users (and potentially introduce risk), because many findings are related.
Therefore, unpowered subgroup analyses should not dictate which outputs remain in the device
when overall case-based criteria are met and demonstrate a favorable benefit-risk profile.
Instead, these results can be shared with users in the labelling where relevant.
Our Harrison.Rad 1.5 Technical Report (see Appendix) describes detailed considerations for
evaluating GenAI-enabled technologies in the context of a radiology foundation model
https://harrison.ai/harrison-rad-1-5-technical-report/ (Mall, et al., 2026). We are not suggesting
that these methods should become the preferred regulatory approaches. Rather, we offer the
report as a contribution to the scientific discussion that FDA’s paper has started.
Some of the methods described in our Harrison.Rad 1.5 technical report are likely closer to the
competency-based approach CDRH is envisioning. Some of these methods are better suited to
evaluating a general-purpose radiology foundation model that can support many different device
and non-device use cases. We recognize, however, that the first use cases to receive marketing
authorization will likely be more narrowly defined. One example is draft reporting, where the
device is not interactive and only produces a draft report from an image and supporting
information such as clinical history. Both general-purpose foundation models and single use
cases such as draft reporting face a similar challenge of output scale, but at different levels:
• A general-purpose model that responds to interactive prompts can serve many use cases,
which may be difficult to define and test comprehensively before the device reaches the
market.
• A single use case such as draft reporting can still produce unconstrained outputs, because
the report may describe any of hundreds of radiological abnormalities. This makes it
nearly impossible to test every output, even through unpowered subgroup analyses.
Unconstrained use cases, and unconstrained outputs within a single use case, raise patient safety
questions. They can also deliver public health benefits that narrower technologies cannot. We
therefore encourage CDRH to consider how device authorizations over the next few years may
open or close paths for future innovation. Just as it would be nearly impossible to evaluate a draft
Page 4 of 12
report at the level of detail typically expected for traditional CAD outputs, it may not be possible
to evaluate every output of a general-purpose foundation model at the level of detail expected for
a single use case such as draft reporting. For assistive technologies where a physician always
reviews the output and where the standard of care baseline itself is highly variable, one way to
address this challenge is to scale the level of evaluation to the intended use, labelling, and user
training described by the manufacturer. For example, if FDA were to ever clear a generalpurpose radiology foundation model with unconstrained but assistive only use cases, the
manufacturer could clearly label what the model was and was not tested for, potentially through
safety by design labelling built into the software interface.
Section VI: Postmarket Monitoring for GenAI-Enabled Devices (Q18-24)
Q18: CDRH is considering whether it may be appropriate to accept greater premarket
uncertainty regarding a GenAI-enabled device’s benefit-risk profile through greater
reliance on postmarket monitoring. Under what conditions might such an approach be
appropriate, and what characteristics of a monitoring program would need to be in place to
justify reduced premarket evidence? Are there device types or risk profiles for which this
approach would not be appropriate?
Q19: Please comment on the potential approaches to postmarket performance evaluation,
including periodic re-benchmarking, sample-based clinician review, and performance
degradation monitoring. What additional approaches should CDRH consider, and how
should the cadence and triggering events for reassessment be determined?
A18 and A19: We support AdvaMed’s comments that “Postmarket surveillance is shared
responsibility with distinct responsibilities for manufacturers based on device-specific
considerations” and that “FDA’s framework should remain flexible, allowing
manufacturers to employ a combination of monitoring approaches proportionate to the
specific risks presented by the device while ensuring continued safety and effectiveness
throughout its lifecycle” (AdvaMed, 2026). Revising the draft AI Lifecycle guidance
(U.S. Food and Drug Administration, 2025) in line with least burdensome principles
would be a practical way to clarify general considerations without creating new GenAIspecific guidance. Specific monitoring approaches depend less on the GenAI technology
itself and more on the specific device design and intended use.
AdvaMed’s comments on the draft AI Lifecycle guidance offer an illustrative example of
where greater reliance on postmarket monitoring may be appropriate (AdvaMed, 2025): a
GenAI-enabled device produces a high-quality radiology report from medical images,
which a radiologist then reviews and revises. It may not be feasible to validate hundreds
of potential findings, many of them rare, with statistical power. In this case, it may be
appropriate to balance premarket and postmarket data requirements so that such devices
can reach the market in situations where premarket data alone would be considered
insufficient.
Page 5 of 12
In such cases, FDA should consider premarket and postmarket approaches that
complement each other and do not overburden device manufacturers or user facilities.
For example, traditional “2+1” reader-based ground truth (two readers, with a third to
resolve disagreements) can be expensive and difficult to obtain in a postmarket setting,
even for a small percentage of exams. Other types of data are readily obtainable but
harder to interpret. For this reason, communication between the manufacturer and end
user should always be the first step when a potential issue arises.
Harrison.ai has published on methods to evaluate such reports, most fully in Section 6.1
of our technical report (Mall, et al., 2026). A key learning is that evaluation metrics
should be multidimensional. A single overall score can give credit to a report that misses
a subtle finding, can accept a vague but incorrect description as a match, and can apply
similar penalties to errors with very different clinical consequences. For example, a report
that identifies the correct diagnosis but omits a minor finding may score about the same
as one that “hallucinates” a dangerous condition that is not present. A single score,
however well calibrated, also does not show why a report is adequate or inadequate:
whether it identified the clinically important findings, whether it would be safe to act on,
and whether a radiologist would accept it. https://harrison.ai/harrison-rad-1-5-technicalreport/
Q20: Please comment on whether the proposed approaches to postmarket monitoring can
be facilitated by machine-based supervisory agents. What considerations, including the
evaluation and reliability of the supervisory agent itself, should CDRH take into account
for such an approach?
A20: The American College of Radiology (ACR) has recently published its experience
using large language models (LLMs) to measure how often AI outputs agree or disagree
with the radiologist’s report for narrower, single-finding AI devices (Schmidt, Brink, &
Bhatia, 2026). This approach could be extended to more complex, multi-finding GenAIenabled use cases such as draft reporting.
Based on our experience developing our C6 framework for postmarket evaluation (Mall,
et al., 2026) and using LLMs for premarket evaluation, we highlight several key
considerations:
• Simple word matching is not sufficient, because physicians and models use many
different phrases to describe the same thing. Several published methods go
beyond word matching, for example by extracting clinical entities or by testing
whether one statement logically supports (“entails”) another. These include
RadGraph (Jain, et al., 2021), GREEN (Ostmeier, et al., 2024), RadFact (Bannur,
et al., 2024), and CheXbert (Smit, et al., 2020).
• Even with more sophisticated methods, the more granular the comparison
between the AI output and the final physician report, the higher the measured
disagreement rate will be.
• Physicians do not necessarily report every abnormality they see, and reporting
practices vary between facilities and between individual physicians.
• Not all differences are clinically meaningful in terms of patient impact and risk.
Page 6 of 12
• For unconstrained outputs, it is impossible to define a fixed list (ontology) of
every possible software output. However, automated methods can build a detailed
ontology that characterizes the AI training inputs.
One way to address some of these issues is to group related radiological abnormalities for
evaluation purposes. These groupings should cover findings that most radiologists
typically report and that are actionable, so that a difference between AI and radiologist
could be clinically meaningful.
These considerations and limitations also apply when LLMs are used as part of least
burdensome premarket approaches. The key takeaway is that it is nearly impossible to
evaluate every output of a draft radiology report, in either the premarket or postmarket
setting, even when semi-automated LLM approaches are used to reduce the burden of
human review and ground truthing.
Q21: What roles might clinicians, healthcare institutions, professional societies, standardssetting bodies, and other stakeholders appropriately play in postmarket monitoring, and
how can these roles be structured without diffusing manufacturer accountability?
A21: A key consideration is the feedback loop to the manufacturer, and our comments
here are no different for GenAI than for any other type of device. If an end user or other
party identifies a potential issue, the manufacturer should be notified first so that it can
determine, through existing quality system processes, whether additional analysis and
investigation are needed.
Q22: CDRH envisions that a manufacturer’s premarket competency-based assessment
might serve as a baseline against which post-deployment modifications could be reevaluated. How might the extent of re-benchmarking or other evidence be scaled to the
nature and expected impact of a given modification? For example, are there categories of
change that might not significantly affect safety or effectiveness of a GenAI-enabled device,
or might be appropriately managed within a sponsor’s quality management system versus
requiring FDA premarket review and authorization before implementation? Are there
categories of changes that might be appropriate for inclusion in a PCCP?
A22: FDA has rightly noted that the burden of evaluating future modifications should
also be a consideration in premarket testing. A specific example illustrates the issues
involved.
For computer-aided detection and diagnosis (CAD) devices, such as lung nodule
detection, FDA’s currently recommended testing has two parts. First, a standalone study
evaluates the software output alone against a reference standard, typically using a “2+1”
clinical reader panel as a proxy for ground truth. Second, a clinical reader study compares
physicians using the AI with physicians alone, typically using a subset of the standalone
cases. While this approach would be overly burdensome to apply to the hundreds of
abnormalities in a GenAI-enabled draft report, typical CAD studies have one advantage
for modifications: the standalone dataset can usually be reused at least a limited number
Page 7 of 12
of times or with small test set augmentation, without new ground truthing or a new reader
study.
By analogy, for certain device modifications, if an LLM was used in premarket testing to
measure agreement and disagreement rates between the AI and the radiologist report,
FDA could consider accepting a repeat of that evaluation on the same dataset, without
repeating any ground truthing performed in the original premarket study. This approach
has limitations, however, because a change in agreement rates after a modification may
not be clinically meaningful. Further evaluation, while possible, would likely require
more human oversight and burden.
To reduce submission burden and focus FDA review resources on the most impactful
changes, FDA should also consider updating its guidance Deciding When to Submit a
510(k) for a Software Change to an Existing Device (U.S. Food and Drug Administration,
2017) to provide more clarity and risk-based flexibility on which changes do not require a
new 510(k), both for GenAI-enabled devices and for other software. For example, to help
industry and FDA keep pace with AI innovation, a change that can be evaluated using the
methods, protocols, and acceptance criteria from the most recently cleared device could
be documented through the quality system rather than in a new 510(k) submission.
Q23: How might PCCP concepts or other change-control approaches be adapted for
GenAI-enabled devices when the nature or scope of future modifications cannot be fully
prespecified?
A23: We support AdvaMed’s recommendation (AdvaMed, 2026) that FDA consider the
approach to predetermined change control plans (PCCPs) outlined in the UK’s recently
published report, National Commission into the Regulation of AI in Healthcare:
Recommendations for a future regulatory framework (National Commission into the
Regulation of AI in Healthcare, 2026). The UK Government has since accepted the
Commission’s recommendations (Medicines and Healthcare products Regulatory Agency,
2026). The report describes an approach “… to PCCPs that allows for manufacturers to
establish boundaries to define the scope of allowable change (rather than only allowing
specific modifications to be prespecified), including clear approaches for adaptive AIenabled medical devices, while maintaining a device’s intended use.” One possible way
to implement such an approach would be to define PCCPs by performance thresholds
rather than by specific changes.
FDA could also consider allowing broader changes to indications for use under a PCCP,
such as adding radiological abnormalities to an assistive AI device. However, if a
manufacturer left these indications out originally because FDA testing expectations were
too burdensome, committing to perform the same studies later would not remove this
barrier to innovation.
Page 8 of 12
Recommendations for FDA next steps
We understand that FDA intended the discussion paper to encourage open stakeholder discussion
rather than to propose draft recommendations. With that in mind, we offer the following
suggestions for FDA’s next steps.
Because factors such as intended use are likely more important than the use of GenAI technology
when evaluating a device’s benefits and risks, GenAI-specific policies may not be needed.
Instead, focusing on the following broadly applicable actions could help balance innovation and
patient safety for GenAI and other technologies:
1. Finalize the AI Lifecycle guidance using least burdensome principles.
2. Update the Deciding When to Submit a 510(k) for a Software Change guidance to allow
more flexibility for device modifications that can be appropriately managed within a
manufacturer’s quality system.
3. Update PCCP policies for to allow manufacturers to define the scope of allowable
changes as opposed to prespecifying specific modifications.
4. Consider review process changes to address product narrowing in response to review
feedback. When sponsors remove features or reduce indications in response to FDA
feedback, this is not systematically tracked, so the issue is difficult to see or measure.
Sponsors also lack a structured way to raise study-design concerns or inconsistent
feedback early, or to request that review teams seek external expert input on novel use
cases where clinical professional guidelines may be lagging. We recommend that CDRH:
(i) track design changes and scope reductions made in response to deficiencies, either as a
CDRH performance metric or through a one-time audit (for example, of AI-enabled
devices);
(ii) audit the consistency of deficiencies within specific device areas;
(iii) extend an equivalent of the “least burdensome” flag to Pre-Submissions; and
(iv) create a manufacturer-initiated pathway to involve CDRH’s Network of Experts,
with appropriate transparency.
We acknowledge that GenAI enables complex use cases that may warrant different evaluation
considerations, but these are likely best addressed in the context of specific intended uses. FDA
has a unique, bird’s-eye view of developments across the industry. Less formal, more flexible
Agency publications describing its thinking, clearly noted as not representing “requirements” or
“recommendations,” could contribute greatly to further stakeholder discussion. Historically, a
first-of-a-kind device is authorized through the De Novo or PMA pathway, and FDA then
develops device-specific guidance, sometimes many years later. This approach may be too slow
to keep pace with the rapidly evolving AI landscape. We encourage FDA to prioritize research
articles and Regulatory Science Tools for high-impact use cases such as radiology draft
reporting. These formats do not carry the same weight as guidance or standards, but they still
share FDA’s current thinking transparently and can be more readily adapted as stakeholder
discussions continue.
As FDA explores least burdensome evaluation methods for GenAI-enabled use cases, we also
encourage the Agency to revisit its testing recommendations for existing devices with similar
risk profiles. As discussed above, applying FDA’s current clinical reader study expectations for
Page 9 of 12
CAD devices to draft reporting would be overly burdensome. We commend FDA for exploring
more fit-for-purpose methods to evaluate GenAI-enabled use cases such as draft reporting. We
also encourage the Agency to revise its recommendations for CAD devices, for example by
relying on standalone testing and case-based evaluation metrics, in place of clinical reader
studies where appropriate, for comprehensive multi-finding CAD devices.
We thank FDA for facilitating public discourse in this important area and for considering these
comments. Please direct any questions to Dr. Mark Phillips, Chief Clinical Officer at Harrison.ai
(Mark.Phillips@harrison.ai), or to the email address associated with this electronic submission.
Level P, 24 Campbell Street Sydney NSW 2000
Website: https://harrison.ai/
APPENDIX: Harrison.Rad 1.5 technical report: A radiology foundation model that can draft
reports from images, priors and clinical context
Page 10 of 12
References
AdvaMed. (2025). Comment letter on the FDA draft guidance Artificial intelligence-enabled
device software functions: Lifecycle management and marketing submission
recommendations . Retrieved from https://www.regulations.gov/comment/FDA-2024-D4488-0031
AdvaMed. (2026). [Comment letter on the FDA discussion paper Considerations for the
regulation of generative AI-enabled medical devices] (Docket No. FDA-2026-N-7874).
Bannur, S., Bouzid, K., Castro, D. C., Schwaighofer, A., Thieme, A., Bond-Taylor, S., . . .
Hyland, S. L. (2024). MAIRA-2: Grounded radiology report generation
(arXiv:2406.04449). arXiv. doi:10.48550/arXiv.2406.04449
Hong, E. K., Song, O., & O’Riordan, K. (2026). Generative artificial intelligence models for
chest radiograph interpretation: Comparison of a general-purpose model, domain-specific
model, and two agentic models. American Journal of Roentgenology.
doi:10.2214/AJR.25.34267
International Medical Device Regulators Forum. (2025). Characterization considerations for
medical device software and software-specific risk (IMDRF/SaMD WG/N81
FINAL:2025). Retrieved from https://www.imdrf.org/sites/default/files/202501/IMDRF_SaMD%20WG_Software-Specific%20Risk_N81%20Final_0.pdf
International Organization for Standardization. (2019). Medical devices: Application of risk
management to medical devices (ISO Standard No. 14971:2019). Retrieved from
https://www.iso.org/standard/72704.html
Jain, S., Agrawal, A., Saporta, A., Truong, S. Q., Duong, D. N., Bui, T., . . . Rajpurkar, P. (2021).
RadGraph: Extracting clinical entities and relations from radiology reports
(arXiv:2106.14463). arXiv. doi:10.48550/arXiv.2106.14463
Mall, S., Nekrasov, V., Kumar, A., Karunasena, S., Nibali, A., Bird, A., . . . Seah, J. (2026).
Harrison.Rad 1.5 technical report: A radiology foundation model that can draft reports
from images, priors and clinical context (arXiv:2607.05880). Harrison.ai. Retrieved from
https://harrison.ai/harrison-rad-1-5-technical-report/
Medicines and Healthcare products Regulatory Agency. (2026). What the National Commission’s
recommendations mean for you. Retrieved from GOV.UK:
https://www.gov.uk/government/publications/what-the-national-commissionsrecommendations-mean-for-you
National Commission into the Regulation of AI in Healthcare. (2026, September 10).
Recommendations for a future regulatory framework. Retrieved from GOV.UK:
https://www.gov.uk/government/publications/national-commission-into-the-regulationof-ai-in-healthcare-recommendations-for-a-future-regulatory-framework
Ostmeier, S., Xu, J., Chen, Z., Varma, M., Blankemeier, L., Bluethgen, C., . . . Delbrouck, J.-B.
(2024). GREEN: Generative radiology report evaluation and error notation. Findings of
the Association for Computational Linguistics: EMNLP 2024. Association for
Computational Linguistics. doi:10.18653/v1/2024.findings-emnlp.21
Schmidt, K., Brink, L., & Bhatia, N. (2026). AI monitoring AI with LLMs: The American
College of Radiology’s Imaging AI Registry. Journal of the American College of
Radiology. Retrieved from https://www.jacr.org/article/S1546-1440(26)00437-0/abstract
Smit, A., Jain, S., Rajpurkar, P., Pareek, A., Ng, A. Y., & Lungren, M. P. (2020). CheXbert:
Combining automatic labelers and expert annotations for accurate radiology report
labeling using BERT. Proceedings of the 2020 Conference on Empirical Methods in
Page 11 of 12
Natural Language Processing (EMNLP) (pp. 1500-1519). Association for Computational
Linguistics. doi:10.18653/v1/2020.emnlp-main.117
U.S. Food and Drug Administration. (2017, October 25). Deciding when to submit a 510(k) for a
software change to an existing device: Guidance for industry and Food and Drug
Administration staff. Retrieved from https://www.fda.gov/media/99785/download
U.S. Food and Drug Administration. (2022, September 28). Clinical performance assessment:
Considerations for computer-assisted detection devices applied to radiology images and
radiology device data in premarket notification (510(k)) submissions. Retrieved from
https://www.fda.gov/media/77642/download
U.S. Food and Drug Administration. (2022, September 28). Computer-assisted detection devices
applied to radiology images and radiology device data: Premarket notification [510(k)]
submissions. Retrieved from https://www.fda.gov/media/77635/download
U.S. Food and Drug Administration. (2025, January 7). Artificial intelligence-enabled device
software functions: Lifecycle management and marketing submission recommendations
[Draft guidance] (Docket No. FDA-2024-D-4488). Retrieved from
https://www.fda.gov/regulatory-information/search-fda-guidance-documents/artificialintelligence-enabled-device-software-functions-lifecycle-management-and-marketing
U.S. Food and Drug Administration. (2026, January 6). Clinical decision support software:
Guidance for industry and Food and Drug Administration staff. Retrieved from
https://www.fda.gov/media/109618/download
U.S. Food and Drug Administration. (2026, August 18). Considerations for the regulation of
generative AI-enabled medical devices: Discussion paper and request for feedback
(Docket No. FDA-2026-N-7874). Retrieved from https://www.fda.gov/medicaldevices/digital-health-center-excellence/considerations-regulation-generative-ai-enabledmedical-devices-discussion-paper-and-request
U.S. Food and Drug Administration. (2026, January 6). General wellness: Policy for low risk
devices: Guidance for industry and Food and Drug Administration staff. Retrieved from
https://www.fda.gov/media/90652/download
Page 12 of 12
Harrison.Rad 1.5 Technical Report
A radiology foundation model that can draft reports from images, priors and clinical context
Suneeta Mall Vladimir Nekrasov Ashnil Kumar Sajith Karunasena
Aiden Nibali Alix Bird Mateo Diaz Shine Jarrel Seah
Harrison.ai
Abstract
Imaging demand is growing faster than the radiology workforce can expand, and the resulting
reporting backlogs cannot be resolved through training and recruitment alone. The most direct
arXiv:2607.05880v1 [cs.CV] 7 Jul 2026
opportunity lies in reducing the time and effort radiologists spend producing reports, a task that
requires interpreting images, integrating clinical history and prior studies, and drafting structured
findings. We present Harrison.Rad 1.5 (HR1.5), a radiology-specific multimodal large language
model that accepts interleaved text and visual inputs and generates structured and unstructured
text across the full breadth of plain-film radiology, spanning computed radiography, diagnostic
x-rays for chest, musculoskeletal, abdominal, spine, and pelvic regions, and mammography.
HR1.5 is trained through a three-stage pipeline: domain adaptation of a base language model on
radiology reports, contrastive vision-encoder training with curriculum-based hard negatives on
approximately six million image-report instances, and visual-question-answering fine-tuning on
multi-turn clinical conversations. We evaluate it with a Findings-Diagnosis scoring framework
that extends RadGraph-XL entity extraction with ontology-based synonym matching and
polarity-contradiction detection, benchmarked on RadBench, a simulated FRCR 2B Short Case
examination scored against Angoff-method thresholds, ReXGradient, and internal multi-modality
datasets. HR1.5 is the only system evaluated to meet the simulated FRCR passing standard and
achieves the highest accuracy on closed-format clinical questions, across anatomical regions, on
internal multi-body-part and mammography report generation, and on the primary clinicallyaligned score for public chest report generation. We further examine explainability and model
behaviour, including question-sensitive Grad-CAM heatmaps, attention analysis, and confidence
estimation, to support responsible future evaluation toward clinical use, and a proposed framework
for a clinically grounded basis for assessing radiology report quality.
This technical report is released for transparency and capability preview; selected sections are intended for
submission to a peer-reviewed venue. This report is released for research purposes only. Harrison.rad.1.5 has not been
cleared for clinical use and is not a medical device. Use must be in accordance with the Terms and Conditions.
Contents
Introduction 3
1 Methodology 5
1.1 Training Methodology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.1.1 Scaling Training to B200-Class Hardware . . . . . . . . . . . . . . . . . . . . 6
1.2 Data Engineering for Training . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3 Serving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Quantitative Results 9
2.1 Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.2.1 FRCR Evaluations — Mock examination . . . . . . . . . . . . . . . . . . . . 13
2.2.2 Closed Question Datasets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.2.3 Open Question Datasets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3 Qualitative Results 21
3.1 Accurate Findings Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2 Draft Report Generation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.3 Longitudinal Reasoning with Priors . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.4 Precise Anatomical Localisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.5 Complex Multi-Finding Reasoning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.6 Style and Template Adaptation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
4 Explainability 33
4.1 Concept Understanding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.2 Concept Localisation: Attention as a Proxy . . . . . . . . . . . . . . . . . . . . . . . 38
4.3 Confidence and Certainty about the Response . . . . . . . . . . . . . . . . . . . . . . 41
4.4 Behaviour on Out-of-Distribution Inputs . . . . . . . . . . . . . . . . . . . . . . . . . 44
5 Discussion 46
6 Future Work 46
6.1 Towards Better Evaluation: The C6 Framework . . . . . . . . . . . . . . . . . . . . . 46
6.2 3D and Volumetric Imaging Support . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Acknowledgements 48
References 49
Introduction
Radiology has become the diagnostic backbone of modern medicine, and that success is now its
central operational problem: imaging demand is rising faster than the workforce that reads it can
grow. In the United States, imaging utilisation is projected to increase by up to 25% over the coming
decades, while radiologist supply is projected to remain constrained relative to demand (Neiman
Health Policy Institute, 2025). In the United Kingdom, the consequences of this imbalance are
already evident: the Royal College of Radiologists reports a 32% shortfall in consultant radiologists
relative to need, a deficit of nearly 2,300 radiologists (Royal College of Radiologists, 2025). In 2024,
imaging demand grew by 8%, while the radiologist workforce grew by only 4.7% (Royal College of
Radiologists, 2025). If these trends continue, the workforce shortfall is projected to reach 40% by
2030 (Royal College of Radiologists, 2025).
Rising imaging volumes relative to workforce growth manifests as reporting backlogs, longer
turnaround times, and reader fatigue that degrades diagnostic accuracy (Krupinski et al., 2010).
The radiology industry is facing a scaling problem that training and recruitment cannot solve.
Modern radiology AI solutions have already demonstrated substantial clinical value through
task specific models that detect important findings, prioritise urgent studies, and assist with
time-consuming measurements. These systems have no doubt improved patient care: accurate
prioritisation of truly urgent cases serves a valuable function in an environment of backlogged
reporting lists, however, it does not solve the underlying problem of the radiologist workforce not
having capacity to read all of the scans being performed.
The greatest opportunity to address radiology’s scaling challenge lies in reducing the effort and
time required for radiologists to report studies, and this is a complex, multifaceted task that involves
not only attention to detail in image interpretation, but also the synthesis of imaging findings with
other clinical information, observation of the patient over time by comparing to prior examinations,
and ultimately the integration of all of this context to consider appropriate differential diagnoses to
answer a clinical question and guide the patient’s treatment.
Harrison.ai’s experience reflects this evolution; our current generation of AI models are trained to
identify a predefined set of findings on images, and whilst they can do this with very high accuracy,
they only assist the radiologist in one of the aforementioned cognitive processes required to produce
a radiology report. Augmenting the radiologist on all aspects of their workflow requires an AI model
that is able to integrate information the way a radiologist does by design.
Harrison.ai’s next phase involves building an AI model to help solve the global radiology capacity
problem from the ground up, a specialised radiology foundation model that can perform accurate
detection, reason through clinical history and prior imaging, and draft a well-structured radiology
report. To support the responsible development and eventual clinical translation of such systems,
it is desirable that such a system also provides calibrated confidence estimates and interpretable
evidence for its conclusions. We develop and evaluate each of these capabilities in the sections that
follow.
These requirements have become attainable through a rapid evolution in multimodal deep
learning. The Transformer (Vaswani et al., 2017) and its visual counterpart, the Vision Transformer
(Dosovitskiy et al., 2021), gave images and text a common modelling substrate; large-scale contrastive
image-text pre-training (CLIP, Radford et al., 2021; SigLIP 2, Tschannen et al., 2025) showed
that aligning the two yields transferable, open-vocabulary recognition. A subsequent line of work
connected pre-trained vision encoders to large language models such as Flamingo (Alayrac et al.,
2022), BLIP-2 (Li et al., 2023b), and the visual instruction tuning of LLaVA (Liu et al., 2023),
turning static classifiers into instruction-following systems that reason over interleaved images and
text, a capability brought to general audiences by GPT-4 with vision (OpenAI, 2023). A vision
Figure 1.1. The radiology foundation-model landscape, 2020–2026. Models are positioned by year of release,
with the bars showing the total released per year; marker colour denotes access level (open source, open weights,
research/closed, API/preview, FDA-cleared/commercial) and marker shape denotes capability tier, from
precursor segmentation and backbone encoders (1), through contrastive or single-modality vision-language
models (2–3), to strong multimodal report-generation VLMs (4) and generalist, context-aware reporting
systems (5). The field is heavily concentrated on the chest radiograph and on single-modality models; the
Harrison.Rad 1 and Harrison.Rad 1.5 models occupy the generalist, context-aware reporting tier.
encoder aligned to a generative language model, and instruction-tuned on multimodal dialogue is
what makes a generalist, natively integrable radiology assistant feasible.
The same framework has been adapted to radiology. Contrastive alignment alone reached expertlevel zero-shot detection from unannotated reports (CheXzero; Tiu et al., 2022), and biomedical-scale
pre-training improved transfer (BiomedCLIP; Zhang et al., 2023). Instruction-tuned biomedical
assistants followed (LLaVA-Med; Li et al., 2023a), alongside increasingly generalist radiology systems:
RadFM extended generative pre-training to web-scale medical data (Wu et al., 2023); ELIXR grafted
an X-ray encoder onto a fixed LLM for broad chest-X-ray tasks (Xu et al., 2023); CheXagent built
an instruction-tuned foundation model and benchmark for chest-X-ray interpretation (Chen et al.,
2024); and MAIRA-1/MAIRA-2 advanced grounded chest-X-ray report generation (Hyland et al.,
2023; Bannur et al., 2024). General-purpose medical multimodal models such as Med-Gemini
(Saab et al., 2024; Yang et al., 2024) and the openly released MedGemma (Sellergren et al., 2025)
demonstrated strong cross-domain capability, and fully open foundation models for chest radiography
have since appeared (Wang, 2025). Figure 1.1 situates these systems by year of release, modality,
and access level. Yet most of these efforts remain confined to a single modality (overwhelmingly the
chest radiograph), do not span the full range of plain-film studies encountered in practice, and are
rarely evaluated against the standardised examinations used to certify human radiologists.
This evaluation gap exposes a deeper limitation: open and general-purpose models tend to fail
on genuinely complex cases because they lack the dense domain knowledge radiological reasoning
requires. The point is concrete: on a simulation of the Fellowship of the Royal College of Radiologists
(FRCR) 2B Rapids examination, which only 40–59% of human candidates pass on first attempt,
Harrison.Rad 1 scored 85.67% (51.4/60), while every other model evaluated scored below 50%
(Harrison.ai, 2024). This advantage has since been borne out independently: in a peer-reviewed
multi-reader study in the American Journal of Roentgenology involving radiologists at Stanford,
Mass General Brigham, and Seoul National University Hospital, Harrison.Rad 1 was the preferred
model for every reader, selected between 39.2% and 66.5% of the time, far ahead of the least-favoured
model, Gemini 2.5 (5.7–9.4%). It also recorded the lowest hallucination rate of any model evaluated,
at 5.7%, compared with 10.8% for Gemini 2.5, 19.3% for MAIRA-2, and 53.8% for MedRAX (Hong,
2026). In the 2025 Healthcare AI Challenge (Mass General Brigham AI Arena and American College
of Radiology, 2025), an independent evaluation conducted by the Mass General Brigham AI Arena
and the American College of Radiology comprising 117 cases, 113 participants, and 2,840 evaluations,
radiology reports drafted by Harrison.Rad 1 were rated clinically acceptable in 65.4% of evaluations,
compared with 79.6% for radiologist-authored reports, indicating that the model approaches but
does not yet match expert reporting quality. Capability of this kind is earned from large-scale,
clinically curated imaging and reporting, not general web data. At the same time, we believe the
competitive landscape should remain open enough for the radiology foundation-model ecosystem to
evolve collectively, through shared benchmarks, comparable protocols, and accessible models, within
the bounds of patient privacy and intellectual property. It is in this spirit that Harrison.ai makes
Harrison.Rad 1 and Harrison.Rad 1.5 available for research and capability evaluation, allowing the
community to probe, compare, and build on them. Currently, however, neither model is cleared or
approved for clinical or diagnostic use in any jurisdiction, and must not be used to inform or drive
patient care.
We present Harrison.Rad 1.5 (HR1.5), an incremental update to our radiology-specific multimodal
large language model Harrison.Rad 1 (HR1) that refines its training and evaluation while retaining
the same overall design. Like its predecessor, HR1.5 accepts interleaved text and visual inputs and
generates structured and unstructured text outputs across the full breadth of computed radiography,
diagnostic x-rays for chest, musculoskeletal, abdominal, spine, and pelvic regions, and mammography.
HR1.5 follows a three-stage training pipeline: domain adaptation of a base LLM on radiology reports,
contrastive vision-encoder training with curriculum-based hard negatives on approximately 6 million
image-report instances, and visual-question-answering fine-tuning on crafted multi-turn clinical
conversations. We evaluate it with a Findings-Diagnosis scoring framework that extends RadGraphXL entity extraction (Jain et al., 2021; Delbrouck et al., 2024) with ontology-based synonym
matching and polarity-contradiction detection, benchmarked on RadBench, a simulated FRCR
2B Short Case examination scored against Angoff-method thresholds, ReXGradient (Zhang et al.,
2025), and internal multi-modality datasets. We further examine model behaviour and explainability
to support responsible future evaluation toward clinical use, by adapting established techniques:
question-sensitive Grad-CAM heatmaps, attention analysis, and confidence estimation.
1 Methodology
This section describes how HR1.5 is built and deployed: the training methodology that produces
the model (§1.1), the data engineering that supplies its learning signal (§1.2), and the serving stack
that delivers it at inference (§1.3). The emphasis throughout is on the principles behind each choice
rather than on implementation specifics.
1.1 Training Methodology
HR1.5 is trained through a multi-stage pipeline, an approach that has become standard for large
vision–language foundation models. The now conventional strategy first establishes competence
within each modality and then aligns them; a language model is adapted to the target domain, a
vision encoder is aligned to text through large-scale contrastive learning in the manner of CLIP
(Radford et al., 2021) and SigLIP (Tschannen et al., 2025), and the two are finally connected and
tuned to follow instructions over interleaved image and text inputs. This approach, pioneered by
Flamingo, introduced gated cross-attention to fuse a frozen vision encoder with a language model
over interleaved sequences (Alayrac et al., 2022); was made more efficient by BLIP-2’s lightweight
querying bridge between frozen image encoders and LLMs (Li et al., 2023b), and popularised as
visual instruction tuning by LLaVA (Liu et al., 2023). This method has been scaled by recent open
multimodal systems such as Qwen2-VL (Wang et al., 2024) and PaliGemma (Beyer et al., 2024).
Staging the training in this way optimises one objective at a time before the components are trained
together, e.g. the language model is first adapted to the target domain, the vision encoder is then
aligned to text through contrastive learning, and only once these are in place are the two wired
for harder joint objective and tuning to follow instructions over interleaved image and text inputs.
Decoupling the stages in this way improves both training stability and sample efficiency.
This paradigm transfers naturally to radiology, and several radiology foundation models adopt a
comparable multi-stage formulation; CheXagent (Chen et al., 2024) and CheXOne (Zhang et al.,
2026), for example, both report a multi-stage training pipeline for chest-X-ray interpretation. HR1.5
follows the same broad philosophy; several stages of pre-training with and without contrastive
objectives, precede one or more fine-tuning stages that confer instruction-style, image-grounded
question-answering ability. Throughout, and in keeping with the overarching goal of a clinically
capable radiology assistant, the learning objectives are chosen to be predominantly radiology-relevant,
so that model capacity is spent on the distinctions that matter for image interpretation and reporting
rather than on general-purpose breadth.
Figure 1.2 summarises this pipeline at a conceptual level: the progression from a domain-adapted
language model, through contrastive vision–language alignment, to instruction fine-tuning for visual
question answering, with the dominant input modality indicated at each stage.
One design choice worth highlighting relates to image captioning, a well-established auxiliary
objective in vision–language model training that has demonstrated clear value across general-domain
systems. In our setting, however, ablations found a dedicated captioning task to be ineffective for
radiology foundation-model training, neither improving nor degrading downstream performance.
HR1.5 therefore omits a separate captioning objective and proceeds directly from contrastive
alignment to instruction fine-tuning.
1.1.1 Scaling Training to B200-Class Hardware
Whereas HR1 was trained on a large fleet of NVIDIA A100 accelerators, HR1.5 was trained at scale
on NVIDIA B200-class hardware. The newer accelerators offer a large step-up in compute and
memory bandwidth, which moved the bottleneck off compute and onto the surrounding data and
memory machinery; realising their throughput meant re-architecting that machinery rather than
simply rescaling it.
Two changes mattered most. We re-architected data loading so that each training rank reads
only the shard it needs, rather than every worker materialising the full dataset, removing the
redundant I/O that otherwise starved the system. And we restructured on-disk storage with lazy,
per-rank caching so that host memory ceased to be the binding constraint. Together, these kept the
Figure 1.2. Conceptual overview of the HR1.5 training pipeline. From radiology data (images, reports, and
clinical context), training proceeds through multi-stage pre-training (with and without contrastive objectives)
followed by instruction fine-tuning for image-grounded visual question answering, yielding Harrison.Rad 1.5.
Objectives are kept predominantly radiology-relevant; internal architecture, stage count, and scale are
intentionally omitted.
B200s consistently fed.
1.2 Data Engineering for Training
The capability of HR1.5 rests as much on how its training data is engineered as on the model
architecture itself. Rather than treating data as a fixed resource to be scraped and lightly augmented,
we approach it as a design problem: the learning signal is constructed from the structure of the
clinical task, so that each example teaches a distinction a radiologist would consider meaningful. This
first-principles, mechanism-design view of crafted data (designing examples around the real-world
mechanism that generates them, rather than around incidental surface statistics) guides both data
workstreams that feed training.
The first workstream concerns contrastive learning, where the difficulty and informativeness of
negative examples largely determine what the vision encoder learns. Building on the established
observation that hard negatives sharpen fine-grained representations (e.g. NegCLIP; Yuksekgonul
et al., 2023), we extend hard-negative construction in both breadth and depth. Breadth refers to
the range of clinically meaningful axes along which a negative can differ from a true description,
so that the encoder must attend to the clinically decisive feature rather than to surface wording.
Depth refers to graded difficulty: negatives follow a curriculum that begins with easily separable
contrasts and progressively introduces harder, more confusable ones as the encoder improves. The
aim is a representation that distinguishes descriptions that read alike but differ clinically, precisely
where lexical shortcuts fail.
Figure 1.3. Data engineering for HR1.5 training, shown conceptually. From a de-identified clinical source,
two engineered learning signals are produced: (top) hard negatives that vary along clinically meaningful axes
(breadth) under a difficulty curriculum (depth), for contrastive alignment; (bottom) context-grounded crafted
dialogues derived from real cases, with deliberate oversampling of under-represented findings, for instruction
tuning. Specific pipelines, prompts, generator models, and dataset scales are intentionally not shown.
The second workstream concerns instruction data, which we craft anchored in real clinical context.
Genuine case material serves as the source from which diagnostic questions and multi-turn exchanges
are derived, spanning the range of tasks a radiologist performs: overall assessment, description of
specific findings, verification of a suspected finding, differential reasoning, anatomical localisation,
and comparison against prior studies. Coverage is shaped deliberately rather than left to the
empirical frequency of conditions (under-represented pathologies and challenging presentations are
oversampled, and example difficulty is controlled) so that the model is not biased toward common
findings at the expense of the rare-but-important ones. The guiding principle throughout is fidelity
to the mechanism that produces real radiological reasoning, so that scale never comes at the cost of
clinical realism.
Figure 1.3 illustrates this philosophy schematically: a single de-identified clinical source is
transformed, along one path, into breadth-and-depth hard negatives for contrastive alignment and,
along the other, into coverage-controlled crafted dialogues for instruction tuning.
All data used for training and evaluation are de-identified following strict institutional and
regulatory guidelines, so that no personal health information is present at any stage of the pipeline.
In aggregate, HR1.5 was trained on approximately 6.5 million studies, yielding in the order of
18 million conversations spanning a broad range of clinical tasks.
1.3 Serving
Because HR1.5 and its predecessor HR1 use a custom architecture, we built a bespoke serving
integration on top of vLLM (Kwon et al., 2023), the high-throughput, low-latency inference engine
for large language models.
At inference, we apply methods that curtail hallucination and keep answers grounded in the
clinically meaningful content of the input. Our approach is inspired by contrastive decoding (Li et al.,
2023d; Leng et al., 2024), which reduces bias and hallucination by contrasting output distributions,
but is developed internally and adapted to our objective of clinically faithful generation. Together
with the served model, this constitutes the HR1.5 Core API.
We extend the Core API through an agentic framework for higher-level tasks such as report
generation and chat-completion/question-answer style interaction, using an architecture analogous
to a system of experts. Within this workflow the constituent models are consulted according to the
clinical task at hand, other harrison.ai models may contribute where appropriate, however HR1.5
is always consulted. This agentic configuration represents the HR1.5+ system reported in the
evaluations that follow.
2 Quantitative Results
We evaluate HR1.5 across a comprehensive suite of benchmarks spanning open-ended and closedformat visual question answering, structured report generation, and clinical examination performance.
Evaluations are reported across all models compared (HR1.5, HR1, CheXagent (Chen et al., 2024),
CheXOne (Zhang et al., 2026), MedGemma (Sellergren et al., 2025), GPT-5.4 (OpenAI, 2026),
Gemini-3-Flash-Preview (Google, 2025), Claude Opus 4.7 (Anthropic, 2026)) wherever applicable,
with results broken down by body part and question type. We additionally evaluated several other
open-source models, but on manual review deemed them unsuitable for the clinical tasks assessed
here; these models were therefore removed from the comprehensive review.
The open-weight baselines were served behind OpenAI-compatible endpoints (vLLM, except
CheXagent which runs locally via HuggingFace Transformers) and queried greedily (temperature 0):
HR1 and HR1.5 used a 300-token generation cap with top_k=1, a frequency penalty of 1.1, and
JSON-schema-constrained (guided) decoding for closed questions; MedGemma (4B/27B) used a
500-token cap; CheXOne used a 300-token cap in instruct mode and 1024 tokens in reasoning mode,
both with a repetition penalty of 1.1; and CheXagent generated up to 300 new tokens in bfloat16
using its shipped generation_config. For the proprietary API models, decoding temperature
was set to 0 wherever the model accepts it (GPT-5.4 and Gemini-3-Flash-Preview) and left at the
provider default otherwise (Claude Opus 4.7, which does not respect a temperature parameter).
Generation was capped at 512 output tokens for GPT-5.4 and Gemini-3-Flash-Preview and at
100 tokens for Claude Opus 4.7; GPT-5.4 was queried through the OpenAI Responses API with
reasoning effort “none”, and Gemini-3-Flash-Preview through its OpenAI-compatible endpoint with
reasoning effort “medium”. All remaining sampling parameters were left at their provider or server
defaults.
2.1 Evaluation
Our evaluation is designed to mirror both how radiologists are certified and how they work day-to-day.
At its core, is a mock simulation of the Royal College of Radiologists (RCR) FRCR examination;
the assessment used to certify practising radiologists. This simulation reconstructs the exam in
both its legacy Rapid Reporting and current Short Case forms (Section 2.2.1). The legacy FRCR
Rapids component (retired in 2025) presented 30 plain radiographs per exam sheet, roughly half
of which were normal, to be reported within 35 minutes. For each film the candidate recorded
only “normal” or the key diagnosis, rewarding fast, accurate triage and penalising over-calling.
With a pass mark of 54 of 60, a score of 2 per question was rewarded if diagnosis was identified
correctly, otherwise 0. The current FRCR 2B Short Cases format (the new standard since 2025)
replaced Rapid Reporting, here we present 25 abnormal cases drawn mainly from chest (50–60%),
musculoskeletal (40–50%) and abdominal (up to 4%) radiographs, each case scored out of up to five
marks (total score 125). Around this we measure open- and closed-format question answering on
plain-film X-ray interpretation tasks, so that both free-text reporting and discrete clinical decisions
are captured.
The benchmarks span public and internal (held out set) sources across a range of body parts
and question types: clinician-authored cases (RadBench), the FRCR examination simulations,
the large-scale public chest-X-ray dataset RexGradient (Zhang et al., 2025), an internal held-out
clinical question set built for broad body-part coverage over a mixed set of medium-to-hard clinical
tasks (RadCoverage-VQA), and mammography (CBIS-DDSM). Together these probe HR1.5 from
standardised examination performance through to everyday reporting across the breadth of plain-film
radiology. The scoring methodology common to these evaluations is described next.
Evaluation Scoring: Findings-Diagnosis Methodology
Our primary evaluation scoring extends RadGraph-XL entity extraction (Jain et al., 2021; Delbrouck
et al., 2024) into a Findings-Diagnosis framework used across RadCoverage-VQA, RexGradient,
and mammography benchmarks. RadGraph-XL extracts clinical entities (anatomy, observations
with polarity labels) and their relations from both predicted and ground-truth text. The FindingsDiagnosis score then applies:
• Normal/abnormal classification: Pattern-based text analysis with negation-aware phrase
matching determines whether the prediction indicates a normal or abnormal study; when
text matching is ambiguous, RadGraph entity analysis checks for the presence of pathological
observations as a fallback.
• Diagnosis overlap scoring: RadGraph-XL extracts clinical entities from both texts; observation
tokens are filtered to remove generic descriptors (severity, laterality modifiers) and meta-terms
(e.g., “finding”, “lesion”), then lemmatised and matched using fuzzy token matching with
ontology-based synonym expansion from a radiology knowledge graph.
• Polarity contradiction detection: Cases where the prediction asserts the opposite polarity
to the ground truth (e.g., definitely denying a finding that is definitely present, or vice versa)
receive zero credit; this is checked bidirectionally across both prediction and reference entity sets.
This scoring is applied consistently across open-ended and closed-format evaluations. For closedformat (multiple-choice) questions, an LLM-as-a-judge approach (Gu et al., 2024) validates answer
correctness when exact matching is insufficient.
Comparing every metric on the same studies, scored across models, makes the deficiencies of
conventional metrics concrete. These metrics track how a report is worded far more than whether it
is clinically correct. Lexical-overlap scores (BLEU, ROUGE-L) reward n-gram agreement with the
reference phrasing, so they collapse toward zero for a factually excellent report worded differently
and can rank a fluent, correct paraphrase below a clumsy near-copy of the reference. Embeddingand entity-based scores (BERTScore, CheXbert-cos, RadGraph) reward semantic or entity similarity
but are largely insensitive to clinical polarity and salience, so they remain high even when a report
hallucinates a clinically significant finding or misses one. Composite error metrics such as RadCliQ
inherit the same blind spots from their components. In short, none of these metrics reliably separates
a clinically correct report from a clinically wrong one, which is why we do not treat them as indicative
of clinical quality.
The Findings-Diagnosis score is introduced to close this gap: by comparing extracted findings
and penalising polarity contradictions, it separates clinically correct from clinically incorrect reports
far more faithfully than any surface metric. It is not, however, infallible; its normal/abnormal
decision and finding overlap are computed automatically, it can still credit a report that misses
a subtle finding or accept a vague but incorrect descriptor as a match. We therefore adopt the
following reporting policy for the open-ended evaluations that follow: the Findings-Diagnosis score is
the primary measure, and where its automated finding-matching is insufficient to adjudicate clinical
correctness, most notably on the free-text report generation of RexGradient, we additionally report
a Judge Findings-Diagnosis score, an LLM-as-a-judge (Gu et al., 2024) adjudication of the same
findings-versus-diagnosis comparison. Conventional report-generation metrics are still tabulated for
consistency with the prior literature, but should be read only as context alongside these clinically
aligned scores.
To make this concrete, the cases in Table 1 place the conventional metrics beside a clinician
reference verdict (GT acceptable — whether a radiologist would accept the report) and the
Findings-Diagnosis (F-D) verdict with its judge-based adjudication (Judge F-D), so the automated
verdicts can be read directly against the clinical reference. The true (ground-truth) answer for each
is given in the first column and the predicted answer verbatim in the second; ✓/✗ indicate whether
the prediction is clinically acceptable.
Table 1: Illustrative per-study comparison of conventional metrics against the Findings-Diagnosis (F-D) verdict, the judge-based Judge F-D, and a
clinician acceptability reference (GT acceptable).
True answer Predicted answer RadGraph-F1 CheXbert-cos BERTScore ROUGE-L BLEU GT acc. F-D Judge F-D
These findings suggest de- MedGemma-27B: “Cervical spondylosis.” 0.000 0.609 0.819 0.000 0.000 ✓ ✓ ✓
generative changes. Anterolisthesis indicates C4 is
slightly displaced forward over
C5. Narrowed discs and facet
joint arthropathy suggest wear.
Foraminal stenosis could lead
to nerve compression. The
left convexity is a mild postural or structural curve. These
changes are common in elderly
patients.
The Hill-Sachs deformity is MedGemma-27B: “Cannot be determined from 0.000 0.307 0.819 0.000 0.000 ✗ ✗ ✗
characterized by a compression this image.”
fracture in the humeral head,
which is consistent with recurrent anteroinferior dislocation.
Increasing ill-defined airspace HR1.5+: “The chest X-ray demonstrates patchy 0.184 0.624 0.855 0.199 0.054 ✓ ✓ ✓
consolidation in the right lung airspace consolidation in the right lower lung
base partially obscuring the zone, consistent with pneumonia in the clinical
right hemidiaphragm. Diffuse context of shortness of breath. Small bilateral
peribronchial cuffing. Small pleural effusions are present with blunting of the
right pleural effusion. Ill- costophrenic angles bilaterally. Bilateral lower
defined opacity in the medial zone interstitial thickening is noted. Aortic arch
aspect of the left lung base calcification is evident, representing age-related
as well. No evidence of pul- vascular changes. Chronic healed or healing rib
monary edema. Heart size is fractures are present on the right side, involvnormal. Calcifications of the ing multiple ribs in the mid to lower thoracic
mitral annulus. Atherosclerotic region, appearing stable without acute displace12 calcifications in the thoracic
aorta. Upper mediastinal conment. The cardiac silhouette is normal in size
without cardiomegaly. No pulmonary edema or
tours are within normal limits. pneumothorax is identified. The mediastinum
Old healed fractures of the pos- appears unremarkable. The lung bases are interolateral aspects of the right completely imaged, which may limit full assesseighth and ninth ribs. ment of pathology in those regions. No calcified
pleural plaques are identified.”
(same ground truth as above) MedGemma-27B: “The cardiomediastinal silhou- 0.017 0.461 0.843 0.187 0.048 ✗ ✗ ✗
ette is within normal limits. The hila are unremarkable. There are diffuse bilateral interstitial opacities, predominantly in the lower lung
zones, with a reticular pattern and some areas
of ground-glass opacity. There is also evidence
of bronchial wall thickening, particularly in the
lower lobes. There are no pleural effusions or
pneumothorax identified. The visualized bony
structures appear unremarkable.”
The bony pelvic ring is intact, HR1.5+: “Mixed lytic and sclerotic lesion in the 0.275 0.843 0.899 0.345 0.102 ✓ ✓ ✓
but there is severe narrowing of left acetabulum with cortical irregularity and
the left hip joint space with sub- protrusio, moderate bilateral hip joint space nararticular sclerosis, subchondral rowing with subarticular sclerosis and marginal
cyst formation, and marginal osteophytes, and degenerative changes in the
osteophytes. Additionally, a sacroiliac joints and lower lumbar spine.”
mixed lytic and sclerotic lesion
is present in the acetabulum
and ilium, with expansile features, cortical breach, and a
soft tissue component.
(same ground truth as above) MedGemma-4B: “Pelvic asymmetry” 0.000 0.653 0.812 0.038 0.000 ✗ ✗ ✗
Mild retrolisthesis at C4/5 and HR1.5+: “Mild retrolisthesis at C4/5 and an- 0.259 0.661 0.871 0.284 0.117 ✓ ✗ ✓
C5/6. Moderate to severe disc terolisthesis at C7/T1. Severe degenerative disc
degenerative changes at C4/5 disease from C4/5 to C6/7, with moderate degenand C5/6 with moderate disc eration at C3/4. Bilateral facet joint arthropathy,
degenerative changes at C6/7. present on the left from C3/4 to C6/7 and on
Right C3/4 facet joint arthropa- the right from C2/3 to C5/6. Severe bony foramthy. Left neural foramina are inal stenosis bilaterally at C2/3. Moderate left
patent. Mild C5/6 foraminal foraminal stenosis at C3/4 and moderate right
narrowing due to uncovertebral foraminal stenosis at C3/4. Mild left foraminal
osteophytes. stenosis at C4/5. Large osteophytes present at
multiple levels. No bone lesions or fractures identified.”
Several patterns illustrate why the surface metrics are not indicative of clinical quality, and why
we rely on the Findings-Diagnosis family instead; throughout, the GT-acceptable column is the
clinician reference against which the automated verdicts are read. First, lexical-overlap metrics
(BLEU, ROUGE-L) and even RadGraph-F1 collapse to zero for a terse but clinically correct answer
(“Cervical spondylosis”), whereas F-D and Judge F-D both credit the correct diagnosis. Second,
those same metrics give near-identical scores to genuinely uninformative outputs: a non-answer
(“Cannot be determined from this image.”) and a vague non-specific descriptor (“Pelvic asymmetry”)
hold BERTScore around 0.81–0.82 (comparable to fully correct reports) yet both F-D and Judge
F-D correctly mark them incorrect. Third, a clinically correct report and a clinically wrong one
on the same study can sit within 0.01–0.09 BERTScore of each other (the chest and pelvis pairs
above); the embedding- and entity-based scores cannot separate them, but F-D marks the report
that mischaracterises the consolidation and denies the effusion, or offers only “pelvic asymmetry”,
as incorrect. Finally, where automated finding-matching is confounded by differing level labels and
marks a thorough, clinically sound report incorrect (automated F-D ✗), the judge adjudication
recovers the clinical agreement (Judge F-D ✓), the regime that motivates reporting Judge F-D on
free-text report generation such as RexGradient.
2.2 Results
2.2.1 FRCR Evaluations — Mock examination
The FRCR examinations are not publicly available. We created a simulation of the FRCR examination, using external data that was not used for model training. In building the examination, we
made our best effort to follow the published examination guidelines.
All results in this section are reported as sheets passed (for FRCR-style exams) or perbenchmark aggregate metrics. We report sheets passed to provide a clinically meaningful measure:
a model either meets the standard for a given exam sheet or it does not. However, we also report
median and interquartile range (IQR; 25th–75th percentiles) of the scores to provide a balance
review across series of mock examination sheets.
Legacy FRCR Rapids (Retired since 2025). The legacy FRCR 2B Rapid Reporting format
consists of a mix of normal and abnormal plain-film X-ray cases. The RCR retired this format in
2025, having replaced it with the Short Case examination (Section 2.2.1); we retain it here as an
additional point of comparison. We evaluate on 70 held-out exam sheets. Each case is assessed in
two sequential parts: first, whether the study is normal or abnormal; and second, where the study is
abnormal, the findings that are present. A case is scored correct only when both parts are answered
correctly: the normal/abnormal call must be right and, if the case is abnormal, the findings must
also be correctly identified. A fully correct case scores 2 and any other response scores 0, giving a
maximum of 60 per sheet. Following the FRCR Rapids guideline, a sheet is counted as passed at a
score of 54 or above. Results are reported as the number of exam sheets passed.
Table 2: Legacy FRCR 2B Rapid Reporting (retired): per-sheet pass rate.
Model Pass Rate
HR1.5+ 15.7%
HR1.5 24.3%
HR1 10.0%
GPT-5.4 0.0%
Gemini-3-Flash-Preview 0.0%
Claude Opus 4.7 0.0%
MedGemma-27B 0.0%
MedGemma-4B 0.0%
As shown in Table 2, HR1.5 leads with 24.3%, followed by HR1.5+ at 15.7%, and HR1 at 10.0%.
None of the general-purpose models nor the medical-domain MedGemma (Sellergren et al., 2025)
variants pass a single sheet, underscoring the difficulty of the legacy FRCR Rapids format and the
advantage of radiology-specific training. We attribute the slight regression of HR1.5+ on FRCR 2B
Rapids to its greater tendency to produce detailed, characterising responses, which are penalised on
normal cases. The same effect is visible in Table 5, where HR1.5+ also regresses slightly on the
closed normal/abnormal question type. This is particularly consequential for FRCR 2B Rapids,
where roughly half of the cases in each sheet are normal.
FRCR 2B Short Cases (New Standard, 2025). The FRCR 2B Short Case examination
represents the new RCR standard since 2025, consisting exclusively of abnormal cases requiring
structured report writing. We evaluate on 16 held-out exam sheets, each containing 25 abnormal
X-ray cases with a body-part distribution mirroring the official RCR specification (approximately
56% chest, with the remainder across MSK upper/lower limb, spine, pelvis, and abdomen).
The model is asked to describe the findings for each case. Responses are scored on a 0–5 rubric
by an LLM-as-a-judge (Gu et al., 2024), with diagnosis as the primary scoring axis aligned to official
FRCR 2B marking criteria. Each sheet yields a cumulative score out of 125.
Following RCR guidance, we estimated the Angoff cut-off (Angoff, 1971) for each exam sheet.
These cut-offs were spot-checked and verified by internal expert radiologists. The human expert
cut-off came in significantly lower (∼50%) than the estimated Angoff cut-off, indicating that we are
holding our models to a deliberately high standard. Results are reported as sheets passed against
the Angoff cut-off.
Table 3: FRCR 2B Short Case examination: mean and median sheet scores (out of 125) along with its 95%
bootstrap confidence interval.
Model Median (/125) Mean (95% CI)
HR1.5+ 86.5 74.8 (61.4–87.2)
HR1.5 73.5 63.5 (52.6–73.5)
HR1 71.0 60.7 (48.4–72.4)
GPT-5.4 44.0 44.6 (39.2–50.0)
Gemini-3-Flash-Preview 38.0 36.9 (32.7–40.8)
Claude Opus 4.7 23.0 23.6 (17.8–29.8)
MedGemma-27B 18.0 21.7 (15.6–28.4)
MedGemma-4B 11.0 17.7 (11.4–24.6)
LLaMa-4-Scout 12.0 14.5 (9.7–19.6)
Table 3 presents the FRCR 2B Short Case results. The Angoff cut-off ranges from 69.3 to 77.3
across sheets (mean 73.2). HR1.5+ and HR1.5 exceed the mean Angoff cut-off (median of 86.5 and
73.5, respectively), with HR1 (median 71.0) falling near the cut-off boundary. All external models
scored well below the passing threshold.
Scoring each sheet against its own set cut-off mark and counting a sheet as passed only when its
cumulative score meets that cut-off, the per-sheet pass rate is as follows:
Table 4: FRCR 2B Short Case examination: per-sheet pass rate against the Angoff cut-off.
Model Pass Rate
HR1.5+ 62.5%
HR1.5 50%
HR1 50%
GPT-5.4 0.0%
Gemini-3-Flash-Preview 0.0%
Claude Opus 4.7 0.0%
MedGemma-27B 0.0%
MedGemma-4B 0.0%
As shown in Table 4, HR1.5+ leads with 62% rate, and HR1 at 50.0%. None of the generalpurpose models nor the medical-domain MedGemma (Sellergren et al., 2025) variants pass a
single sheet, underscoring the difficulty of the FRCR 2B Short Cases format and the advantage of
radiology-specific training.
2.2.2 Closed Question Datasets
To complement publicly available benchmarks, we curated closed evaluation datasets targeting
specific pathology categories and clinical scenarios. These datasets were generated using a structured
pipeline: an LLM produces clinically plausible multiple-choice questions and distractors from
ground-truth case data, and generated questions are subsequently reviewed and validated by expert
radiologists.
Categories covered include abdominal pathologies, airspace opacification, chest lines and tubes,
masses and bone lesions, C-spine levels, and body-part positioning, among others.
We report accuracy broken down by question type across all closed datasets (RadBench-Closed,
ReXGradient-Closed, FRCR-Closed, Pathology-Closed), pooling items of the same type regardless
of source dataset. This avoids aggregation artefacts from dataset-size imbalance. We restrict to
question types with sufficient sample size (n > 500) and note the random baseline for each, which
varies with the number of answer options.
Table 5: Closed-format accuracy by question type, along with the 95% Wilson score confidence intervals.
Diagnosis Finding Normal/Abn Laterality
Model
(n=13390) (n=1443) (n=1053) (n=534)
Random baseline 25.0% 50.0% 50.0% 33.3%
82.0% 81.2% 82.8% 85.4%
HR1.5
(81.3–82.6) (79.1–83.1) (80.4–85.0) (82.2–88.1)
78.0% 77.4% 78.3% 85.5%
HR1.5+
(77.3–78.7) (75.2–79.5) (75.7–80.7) (82.3–88.2)
75.3% 78.0% 81.9% 68.7%
HR1
(74.6–76.0) (75.8–80.1) (79.5–84.1) (64.6–72.5)
77.6% 73.0% 70.6% 88.2%
GPT-5.4
(76.9–78.3) (70.7–75.2) (67.8–73.3) (85.2–90.7)
67.9% 77.1% 66.2% 80.0%
Gemini-3-Flash-Preview
(67.1–68.7) (74.9–79.2) (63.3–69.0) (76.4–83.2)
59.9% 58.1% 56.6% 77.3%
Claude Opus 4.7
(59.1–60.7) (55.5–60.6) (53.6–59.6) (73.6–80.7)
79.3% 51.3% 63.8% 50.2%
MedGemma-27B
(78.6–80.0) (48.7–53.9) (60.9–66.6) (46.0–54.4)
75.3% 76.5% 53.7% 58.2%
MedGemma-4B
(74.6–76.0) (74.2–78.6) (50.7–56.7) (54.0–62.3)
46.0% 55.3% 50.8% 66.3%
LLaMa-4-Scout
(45.2–46.8) (52.7–57.8) (47.8–53.8) (62.2–70.2)
Table 5 shows per-question-type accuracy. HR1.5 leads in Diagnosis (82.0%), Finding (81.2%),
and Normal/Abnormal (82.8%). GPT-5.4 achieves the highest Laterality accuracy (88.2%). All
models substantially exceed random baselines, with the largest margins on Diagnosis where chance
is 25%.
Table 6 presents overall closed-question accuracy grouped by body-part category. Body parts are
mapped to four groups: CXR (chest/thorax), MSK (musculoskeletal, including spine, extremities,
pelvis, head, and dental), Abdomen, and Other. ReXGradient items are assigned to CXR as the
dataset contains only chest X-rays.
Table 6: Closed-format accuracy by body-part group.
CXR MSK Abdomen Other
Model
(n=12103) (n=3768) (n=143) (n=826)
79.0% 88.8% 80.4% 80.3%
HR1.5
(78.3–79.7) (87.8–89.8) (73.1–86.1) (77.4–82.9)
76.2% 82.2% 74.8% 69.6%
HR1.5+
(75.4–77.0) (80.9–83.4) (67.1–81.2) (66.4–72.6)
73.1% 80.0% 74.1% 73.5%
HR1
(72.3–73.9) (78.7–81.2) (66.4–80.6) (70.4–76.4)
77.5% 76.6% 84.6% 57.3%
GPT-5.4
(76.7–78.2) (75.2–77.9) (77.8–89.6) (53.9–60.6)
67.9% 68.3% 79.0% 71.2%
Gemini-3-Flash-Preview
(67.1–68.7) (66.8–69.8) (71.6–84.9) (68.0–74.2)
59.5% 58.7% 60.8% 65.0%
Claude Opus 4.7
(58.6–60.4) (57.1–60.3) (52.6–68.4) (61.7–68.2)
65.6% 52.8% 44.4% 61.9%
MedGemma-27B
(64.7–66.4) (51.2–54.4) (36.5–52.6) (58.5–65.1)
64.6% 68.1% 60.5% 50.8%
MedGemma-4B
(63.7–65.4) (66.6–69.6) (52.3–68.1) (47.4–54.2)
77.6%
CheXOne-Reasoning – – –
(76.8–78.3)
72.7%
CheXOne-Instruct – – –
(71.9–73.5)
34.7%
CheXagent – – –
(33.9–35.6)
44.5% 55.0% 58.7% 56.3%
LLaMa-4-Scout
(43.6–45.4) (53.4–56.6) (50.5–66.4) (52.9–59.6)
HR1.5 leads in CXR (79.0%), MSK (88.8%), and Other (80.3%). GPT-5.4 achieves the highest
Abdomen accuracy (84.6%), though the Abdomen subset is small (n=143). CheXOne-Reasoning
reaches 77.6% on CXR, competitive with GPT-5.4 but below HR1.5. CXR-only models (CheXOne,
CheXagent) are not evaluated on non-chest body parts.
2.2.3 Open Question Datasets
The closed-format results above isolate discrete clinical decisions, but they constrain the model to a
fixed set of options and so understate the harder problem a radiologist actually faces: producing
free text that is both clinically correct and well formed. We therefore turn to open-ended evaluation,
where each model generates unconstrained descriptions or full reports that are scored primarily
with the Findings-Diagnosis methodology (Section 2.1), alongside conventional report-generation
metrics stated in the literature. As discussed earlier, surface-level text-overlap scores can diverge
sharply from clinical correctness, so the Findings-Diagnosis score is treated as the primary signal
and the remaining metrics as context.
The following subsections report open-ended performance across four complementary settings:
clinician-authored cases spanning multiple body parts and question types (RadBench, Section 2.2.3);
large-scale public chest-X-ray report generation (RexGradient, Section 2.2.3); the internal, multibody-part held-out set (RadCoverage-VQA, Section 2.2.3); and mammography (Section 2.2.3).
Together they trace HR1.5’s open-ended behaviour from standardised, clinician-curated questions
through to everyday reporting across the breadth of plain-film radiology.
RadBench Evaluation
RadBench is a clinician-authored evaluation framework developed at harrison.ai for benchmarking
multimodal radiology foundation models. It consists of 105 unique X-ray cases with 497 questions
(378 closed, 119 open) spanning question categories including finding presence, anatomical location,
laterality, diagnosis, clinical reasoning, normal/abnormal classification, and modality identification.
Closed-format evaluation uses binary accuracy, precision, recall, F1, and Matthews correlation
for yes/no questions, and accuracy with option-count-scaled weighting for multi-choice questions.
Open-ended evaluation uses ROUGE (Lin, 2004), BLEU (Papineni et al., 2002), CheXBERT
(Smit et al., 2020), RadGraph F1, the Findings-Diagnosis score described above, and an LLM-as-ajudge (Gu et al., 2024) adjudication (Judge F-D).
A note on data availability. A portion of the RadBench cases is sourced from the NIH
MedPix teaching atlas. Since RadBench was released, MedPix has been permanently retired by
the U.S. National Library of Medicine (NLM), following a realignment of agency priorities and
resources; this is a permanent withdrawal rather than a temporary outage. As a consequence,
many of the source images can no longer be retrieved, and the open-ended split has shrunk to only
69 usable cases. This remaining sample is too small to support an unbiased comparison across
models. We therefore report the RadBench open-ended results below for completeness and posterity
only, and caution against reading them as a reliable ranking; the larger-scale evaluations in the
surrounding sections (ReXGradient, RadCoverage-VQA, and mammography), together with the
FRCR simulations, provide the load-bearing evidence.
Table 7: RadBench open-ended evaluation on the reduced (n=69) set. Reported for completeness only; see
the note on data availability above.
Model RadGraph F1 ↑ RadCliQ-v1 ↓ F-D Score ↑ CheXBERT ↑ BLEU ↑ ROUGE-L ↑ Judge F-D ↑
HR1.5+ 0.008 0.274 58.0% 0.793 0.017 0.201 31.0%
HR1.5 0.020 0.133 73.9% 0.858 0.049 0.470 37.1%
HR1 0.003 0.287 59.4% 0.788 0.000 0.173 28.4%
GPT-5.4 0.006 0.112 78.3% 0.885 0.053 0.497 33.3%
Gemini-3-Flash-Preview 0.007 0.018 71.0% 0.921 0.085 0.608 35.1%
Claude Opus 4.7 0.006 0.460 31.9% 0.654 0.002 0.065 24.9%
MedGemma-27B 0.000 0.149 59.4% 0.872 0.036 0.506 12.8%
MedGemma-4B 0.000 0.122 58.0% 0.886 0.000 0.489 16.5%
Table 7 shows open-ended evaluation results using RadBench. Gemini-3-Flash-Preview achieves
the best scores on most surface metrics, including RadCliQ-v1 (0.018), CheXBERT (0.921), BLEU
(0.085), and ROUGE-L (0.608), while GPT-5.4 leads on the F-D Score (78.3%). HR1.5 attains
the highest Judge F-D score (37.1%) and the highest RadGraph F1 (0.020). Claude Opus 4.7
records the lowest F-D Score (31.9%), possibly because its responses diverge from the expected
radiology format. Given the reduced sample (n=69; see the note above), these results are reported
for completeness only.
A recurring cause of the divergence between surface metrics and clinical quality is easy to see
at the level of individual items: a longer, more specific response can reduce exact-match overlap
even as it improves clinical fidelity. To illustrate, consider the question “What complication of
the oesophageal stent has occurred?”, for which the ground truth is “the oesophageal stent has
migrated.” HR1 answers: “Complication of the oesophageal stent is evident. The oesophageal stent
appears migrated.” HR1.5 answers: “The complication of the oesophageal stent is that it has slipped
out of its original position and now lies horizontally in the fundus of the stomach.”
Under F-D scoring, HR1 is marked correct and HR1.5 incorrect. Yet HR1.5 is the clinically richer
response: it not only identifies the migration but localizes the displaced stent as lying horizontally
in the gastric fundus. The penalty arises purely from surface form. HR1.5 paraphrases “migrated”
as “slipped out of its original position,” failing to reproduce the ground-truth keyword, whereas
HR1’s shorter answer echoes it verbatim. The longer, more specific output is thus scored as wrong
despite its higher clinical fidelity.
RexGradient
RexGradient (Zhang et al., 2025) is a large-scale chest X-ray evaluation benchmark comprising
approximately 160,000 studies. We evaluate on the Zero-shot report generation task i.e. given
patient demographics and clinical indication, generate a detailed findings statement.
Similar to the RadBench evaluation, performance is measured using RadGraph F1, CheXBERT
(Smit et al., 2020), RadCliQ (Yu et al., 2023) composite scores, and our Findings-Diagnosis
methodology. Additionally, we evaluate over 11,000 closed-format questions derived from the
dataset, spanning diagnosis (83%), finding presence, normal/abnormal, clinical diagnosis, report
interpretation, and laterality categories.
Table 8: RexGradient zero-shot report generation.
Model F-D Score ↑ Judge F-D ↑ RadGraph F1 ↑ RadCliQ-v1 ↓ CheXBERT ↑ BLEU ↑ ROUGE-L ↑
49.7% 68.9%
HR1.5 0.124 0.407 0.450 0.042 0.204
(48.8–50.7) (68.0–69.8)
43.1% 73.0%
HR1.5+ 0.142 0.378 0.471 0.039 0.179
(42.2–44.1) (71.7–73.5)
46.7% 63.5%
HR1 0.149 0.384 0.443 0.044 0.199
(45.7–47.7) (62.6–64.5)
47.1% 65.0%
CheXOne-Reasoning 0.238 0.158 0.529 0.110 0.281
(46.1–48.0) (64.1–66.0)
47.7% 66.6%
CheXOne-Instruct 0.232 0.167 0.530 0.108 0.277
(46.7–48.7) (65.7–67.5)
46.7% 65.3%
MAIRA-2 0.152 0.329 0.491 0.073 0.224
(45.7–47.7) (64.4–66.2)
47.0% 60.6%
MedGemma-27B 0.166 0.342 0.460 0.068 0.243
(46.0–48.0) (59.7–61.6)
50.5% 61.0%
MedGemma-4B 0.208 0.266 0.476 0.070 0.243
(49.6–51.5) (60.0–61.9)
25.7% 42.0%
CheXagent 0.055 0.650 0.336 0.020 0.126
(24.9–26.6) (41.1–43.0)
Table 8 shows zero-shot report generation results. RadCliQ-v1 is a composite error metric (lower
is better). CheXOne-Reasoning achieves the highest RadGraph F1 (0.238), RadCliQ-v1 (0.158),
BLEU, and ROUGE-L, while CheXOne-Instruct edges ahead on CheXBERT (0.530). MedGemma4B leads on the automated Findings-Diagnosis score (50.5%), while on the Judge Findings-Diagnosis
score — the more reliable measure on this benchmark — HR1.5+ leads (73.0%). The gap between
text-overlap metrics and F-D score highlights the distinction between surface-level textual similarity
and clinical accuracy.
A note on this benchmark: a substantial portion of RexGradient consists of structured reports
with additional sections and finer-grained nodes than a free-text findings statement. On such
references the automated Findings-Diagnosis score performs suboptimally due to limitations of its
underlying RadGraph entity extraction, whose finding-matching becomes brittle when the ground
truth is heavily structured (for the reasons set out in Section 2.1), and the F-D column above
reflects this. The Judge Findings-Diagnosis score is more representative of clinical correctness
here, with HR1.5+ attaining the highest value (73.0%); even so, a residual gap remains under this
stronger evaluator, underscoring that no current automatic evaluator fully captures report quality
on structured data.
RadCoverage-VQA
RadCoverage-VQA is an internal, held-out evaluation set of clinical questions built for broad bodypart coverage (chest X-ray, MSK, spine, abdomen, pelvis, and mammography) across a mixed set of
medium-to-hard clinical tasks. Cases are drawn from validation-tagged production data, providing
a diverse and clinically representative test set that extends well beyond chest-only evaluation.
Performance is measured using the Findings-Diagnosis methodology, with results broken down
by body-part category. Body parts are grouped using the same mapping as the closed datasets
(Section 2.2.2): CXR, MSK, Abdomen, and Breast, with remaining body parts (clavicle, coccyx,
fibula, finger, sacrum) in Other.
Table 9: RadCoverage-VQA Findings-Diagnosis scores by body-part group.
CXR MSK Abdomen Breast Other Overall
Model
(n=3224) (n=10245) (n=1805) (n=944) (n=340) (n=16558)
70.1% 64.8% 70.0% 67.7% 65.3% 66.6%
HR1.5
(68.5–71.7) (63.9–65.7) (67.8–72.1) (64.7–70.6) (60.1–70.2) (65.9–67.3)
50.2% 47.6% 36.5% 43.5% 47.9% 46.6%
HR1.5+
(48.5–51.9) (46.6–48.6) (34.3–38.7) (40.4–46.7) (42.6–53.2) (45.8–47.4)
65.9% 63.1% 66.9% 67.5% 60.0% 64.3%
HR1
(64.2–67.5) (62.2–64.0) (64.7–69.0) (64.4–70.4) (54.7–65.1) (63.6–65.0)
17.4% 15.7% 10.7% 13.1% 14.1% 15.3%
MedGemma-27B
(16.1–18.7) (15.0–16.4) (9.4–12.2) (11.1–15.4) (10.8–18.2) (14.8–15.9)
25.5% 22.9% 16.0% 16.5% 26.5% 22.4%
MedGemma-4B
(24.0–27.0) (22.1–23.7) (14.4–17.8) (14.3–19.0) (22.1–31.4) (21.8–23.0)
Table 9 shows Findings-Diagnosis scores on RadCoverage-VQA. HR1.5 leads across all body-part
categories, with the largest gains over HR1 in Abdomen (+3.1pp) and CXR (+4.2pp). MedGemma
models score substantially lower, likely because RadCoverage-VQA uses a multi-turn dialogue format
that differs from their training distribution.
Mammography
Mammography evaluation extends the Findings-Diagnosis framework to breast imaging using the
CBIS-DDSM dataset (n=1625).
Table 10: Mammography (CBIS-DDSM) Findings-Diagnosis scores.
Model F-D Score ↑ 95% CI
HR1.5+ 77.8% 75.6–79.7%
HR1.5 64.9% 62.5–67.1%
HR1 64.9% 62.6–67.2%
MedGemma-27B 67.6% 65.3–69.9%
MedGemma-4B 61.4% 59.0–63.7%
Table 10 shows mammography Findings-Diagnosis scores. HR1.5+ leads at 77.8%, a substantial
improvement over HR1.5 and HR1 (both 64.9%). MedGemma-27B achieves 67.6%, outperforming
HR1.5 and HR1 on this modality.
3 Qualitative Results
Beyond aggregate benchmark scores, we also characterised qualitative aspects of HR1.5’s output
through structured radiologist review. This section organises that assessment around six capabilities
that together characterise how HR1.5 performs as compared to publicly available frontier models.
We compare (dated May, 2026) HR1.5+ (Harrison.Rad 1.5, agent-driven VQA, temperature
0) against three publicly available frontier multimodal models: GPT-5.4 (OpenAI), Gemini-3Flash-Preview (Google) and Claude Opus 4.7 (bedrock-anthropic-claude-opus-4-7, AWS Bedrock).
Decoding temperature was set to 0 wherever the model accepts it (HR1.5+, GPT-5.4 and Gemini-3Flash-Preview) and left at the provider default otherwise: Claude Opus 4.7 (which does not respect
a temperature parameter). All remaining sampling parameters were left at provider defaults. Every
model received the same prompt and image(s) per case.
3.1 Accurate Findings Detection
Accurate image interpretation is core to the function of a radiology foundation model: correctly
identifying the finding that is actually present, especially when it is subtle and easily mistaken for a
more dramatic abnormality.
HR1.5+ names the scaphoid waist fracture directly. All three frontier models miss it, each with a
different wrong read — GPT-5.4 a soft-tissue mass with “bones intact,” Gemini a giant-cell tumour
of the tendon sheath, and Claude Opus 4.7 a distal radius (Colles’) fracture (figure 3.1).
3.2 Draft Report Generation
The primary requirement is that the draft be accurate and faithfully to the findings, and correctness
is never traded off against style. Given an accurate draft, the secondary aim is to present it in a
form close to what the reporting radiologist would have written, so the content reads naturally
and reduces cognitive load rather than expecting them to change practice. Crucially, matching the
radiologist’s prose is intended to ease review, not to substitute for it. The draft is a starting point
that the radiologist verifies and edits, and familiarity of style should make discrepancies easier to
spot rather than encouraging uncritical acceptance.
Reference case 3.2, a portable chest radiograph demonstrating innumerable bilateral pulmonary
metastases with an indwelling PICC line. The verbatim prompt and each model’s verbatim answer
are shown below, first to the image alone (figure 3.2.1), then with the clinical history added (drafting
a report is exactly where history should be used), with the key finding in each answer highlighted in
bold (figure 3.2.2).
From the image alone, HR1.5+ names the metastases directly, while all three frontier models
read the identical film as pneumonia / COVID-19.
With the history supplied, HR1.5+, Gemini-3-Flash-Preview and Claude Opus 4.7 all identify
the bilateral pulmonary metastases, and HR1.5+ additionally reports the appropriately-sited PICC
line. GPT-5.4 still reads the nodules as multifocal pneumonia and never mentions metastatic disease,
a persistent miss even when handed the diagnosis-relevant history.
Figure 3.1. Scaphoid fracture. Case courtesy of Subash Thapa, Radiopaedia.org (case 40623), used under
licence.
Figure 3.2.1. Metastatic lung disease, portable chest radiograph. Case courtesy of Daniel Hyeong Seok
Kim, Radiopaedia.org (case 153711), used under licence. Image only (no clinical history).
Figure 3.2.2. Metastatic lung disease, portable chest radiograph. Case courtesy of Daniel Hyeong Seok
Kim, Radiopaedia.org (case 153711), used under licence. Image with clinical history.
3.3 Longitudinal Reasoning with Priors
When prior imaging is available, HR1.5 automatically compares the current study against it and
surfaces interval change, described in plain clinical language rather than as a flagged checklist. The
comparison is performed at the moment the case is opened, so the report is contextualised against
the patient’s history from the outset.
Reference case 3.3, current and prior chest radiographs (supplied labelled), with the clinical
history “Day 2 of antibiotics, follow-up imaging.” The reference optimal report describes interval
worsening of right-sided airspace opacification with a large right pleural effusion and new left
mid-zone opacity (figure 3.3).
HR1.5+ reports the interval progression correctly, with increased right-sided airspace opacification, a new left mid-zone opacity and the enlarging right effusion, matching the reference, as does
GPT-5.4. Gemini-3-Flash-Preview adds a spurious cardiomegaly, and Claude Opus 4.7 misses the
large effusion and fabricates bilateral central venous catheters that are not present. Overt interval
progression is thus within reach for HR1.5+; subtler change, for example a fracture occult on the
prior and newly visible, is harder for a single pass, which the production comparison route addresses
by reading each study independently and diffing the two.
3.4 Precise Anatomical Localisation
Findings are placed with greater anatomical specificity, including quantitative measurements such
as the cardiothoracic ratio. The improvement is most pronounced on the spine, abdomen, and
post-procedural imaging, among the areas of highest diagnostic complexity.
Reference case 3.4, a chest radiograph demonstrating a left apical mass in the left upper lobe,
partially obscured by the clavicle (figure 3.4). HR1.5+ localises the mass correctly to the left upper
lobe, as does Gemini-3-Flash-Preview; GPT-5.4 gets the side but mischaracterises it as a calcified
mediastinal/chest-wall lesion. Claude Opus 4.7 lateralises it to the wrong (right) side, a laterality
error with direct clinical consequences.
3.5 Complex Multi-Finding Reasoning
HR1.5 shows improved performance on high-cognitive-load studies that contain multiple findings,
precisely the cases in which a finding is most likely to be missed. It also offers stronger coverage of
rare-but-real findings that fall outside the classical reporting ontology.
Multiple coexisting findings (a chest study). Reference case 3.5.1, PA and lateral chest radiographs
demonstrating congestive cardiac failure, a study whose several coexisting findings (cardiomegaly,
pulmonary venous congestion, interstitial oedema) must be tied to a single diagnosis (figure 3.5.1).
HR1.5+ ties the cardiomegaly, pulmonary venous congestion and interstitial oedema together as
congestive cardiac failure, as does GPT-5.4. Gemini-3-Flash-Preview instead diagnoses sarcoidosis,
hallucinating bilateral hilar and paratracheal lymphadenopathy; Claude Opus 4.7 diagnoses combined
pulmonary fibrosis and emphysema, calling the cardiomediastinal silhouette normal and so missing
the cardiomegaly, two confident non-cardiac misreadings of a classic heart-failure film.
A complex single-region injury. Reference case 3.5.2, a knee radiograph series demonstrating
three coexisting findings: a lateral patellar dislocation, a medial patellar avulsion fracture, and a
joint effusion (figure 3.5.2).
HR1.5+ enumerates all three coexisting findings, as does Claude Opus 4.7 (describing the medial
avulsion as an osteochondral fragment). GPT-5.4 mis-reads the displaced patella as a normal
bipartite-patella variant, concluding “no acute fracture or dislocation,” while Gemini under-grades
Figure 3.3. Pneumococcal pneumonia, current (left) and prior (right) chest radiographs. Case courtesy of
Jack Ren, Radiopaedia.org (case 29090), used under licence.
Figure 3.4. Left upper lobe mass. Case courtesy of Dai Roberts, Radiopaedia.org (case 80445), used under
licence.
the dislocation as subluxation, the high-cognitive-load setting where a missed finding is most
consequential.
Figure 3.5.1. Congestive cardiac failure, PA and lateral chest radiographs. Case courtesy of Frank Gaillard,
Radiopaedia.org (case 29092), used under licence.
Figure 3.5.2. Patellar dislocation with medial avulsion fracture and joint effusion. Case courtesy of Rodney
Strahan, Radiopaedia.org (case 51097), used under licence.
3.6 Style and Template Adaptation
Slotting into an existing reporting workflow means following a prescribed report template exactly:
populating the sections that apply and, just as importantly, suppressing those that do not, rather
than padding the report with empty boilerplate. The test is therefore not whether a model
can produce section headings (all of them can) but whether it treats the template’s conditional
instructions as rules to obey.
Reference case 3.6, as shown in figure 3.6, a chest radiograph demonstrating interstitial thickening
(a fibrosis pattern), with the clinical history “Shortness of breath” supplied. The study is reported
against a fixed sectioned template (mirroring the structured variant the rad-agent production
pipeline emits) in which Lines & Devices, Abdominal and Soft Tissue are explicitly marked optional,
to be included only when relevant, while the remaining regions take a normal statement where no
positive finding is present.
All four models detect the bilateral interstitial/fibrotic process, follow the capitalised heading
convention, and produce the sectioned layout, so neither detection nor formatting separates them.
The differentiator is the conditional instruction to include the optional sub-regions only when relevant.
HR1.5+ and GPT-5.4 follow it exactly here, emitting only the populated regions and dropping all
three optional sub-regions (Lines & Devices, Abdominal, Soft Tissue), none of which apply. Claude
Opus 4.7 appends a redundant Soft Tissue: Unremarkable, and Gemini-3-Flash-Preview appends
both Abdominal and Soft Tissue.
Figure 3.6. Interstitial thickening. Case courtesy of Bruno Di Muzio, Radiopaedia.org (case 45314), used
under licence.
4 Explainability
Instruction-tuned vision-language models tend to present their answers as fluent text without
surfacing the basis for them, which makes them harder to interpret. We see value in complementing
HR1.5’s outputs with explanations that show the reasoning behind an answer and connect it back
to verifiable features in the image. Such grounding makes the model easier to adopt and to work
with (Reyes et al., 2020; Tonekaboni et al., 2019): it lets a reader trace why an answer was given,
form a view of how reliable it is (Hendricks et al., 2016; Selvaraju et al., 2017), and develop a clearer
sense of where the model’s strengths and limits lie (Ghassemi et al., 2021). The methods described
below are exploratory, early steps in this direction for HR1.5 rather than a finished or validated
capability, and they treat the model’s own internal signals as additional evidence to be surfaced
rather than discarded.
To this end, with HR1.5 we have built a post-hoc explainability engine that derives interpretable
outputs alongside each prediction, using only quantities the model already computes during a normal
forward pass (activations, gradients, and attention weights) with no retraining or architectural
change. The engine is organised around three questions a clinician asks of any model output:
• Concept understanding (Section 4.1): Does the model truly know this finding? We test
whether findings are genuinely encoded and used internally, not just named.
• Concept localisation (Section 4.2): Where on the image is this answer coming from? We map
each answer back to the image regions that drove it.
• Confidence and certainty (Section 4.3): How much should I trust this answer? We attach a
reliability score to each answer and correct it for known sources of false certainty.
4.1 Concept Understanding
A central question for any clinical model is whether it genuinely represents the findings it reports. A
linear probe that separates positive from negative examples shows that a concept is decodable from
the model’s activations, but decoding alone does not establish that the model uses that concept when
forming an answer. The distinction matters clinically: only a finding the model actively uses, rather
than one we can merely read off its activations, is a reliable handle for monitoring or intervention.
To examine this we apply an offline analysis we term concept tracing, which combines four lines
of evidence: extraction of a concept direction, validation of its separability, characterisation of its
geometry and depth, and a causal steering test. The results below cover the two largest free-text
radiography domains, chest radiography (CXR) and musculoskeletal imaging (MSK).
For each domain we curate a vocabulary of clinically salient findings and extract, for each,
a unit-norm linear direction from labelled activations. The primary estimator is a difference of
means, the unit-normalised difference between the mean positive and mean negative activations,
which has been shown to be more causally implicated in model outputs than more elaborate probes
(Marks and Tegmark, 2023); L2-regularised logistic regression (Alain and Bengio, 2017) serves as an
alternative. Two extraction sites are probed independently: the decoder residual stream at the last
prompt token, where the model has integrated the full image and question, and the image-side latent
representation pooled across the bridger’s slots, which indicates whether a concept is already linear
before text conditioning. Probes are trained within each domain so that a direction discriminates
the finding itself rather than the imaging modality.
On a held-out split, each concept is scored by its positive-to-negative projection gap, ROC-AUC,
and Cohen’s d, and assigned to the decoder layer at which its probe is strongest. Across the full
vocabularies, both domains separate held-out data well: 17 CXR concepts reach a mean ROC-AUC
of 0.873 (median 0.853, mean Cohen’s d 1.82) and 24 MSK concepts reach 0.895 (median 0.907,
mean Cohen’s d 1.91). Tables 11 and 12 report every concept tested in each domain, with its
held-out ROC-AUC and Cohen’s d.
Table 11: Held-out concept separability for CXR Table 12: Held-out concept separability for MSK
(ROC-AUC and Cohen’s d). (ROC-AUC and Cohen’s d).
Concept ROC-AUC Cohen’s d Concept ROC-AUC Cohen’s d
scoliosis 0.999 3.26 loose body 0.999 3.22
hyperinflation 0.995 3.43 bone cyst 0.985 3.32
endotracheal tube 0.965 2.30 osteomyelitis 0.969 2.81
pacemaker 0.956 2.10 spiral fracture 0.961 2.63
rib fracture 0.934 3.16
comminuted fracture 0.951 2.15
pleural thickening 0.917 1.87
joint replacement 0.942 2.55
subcutaneous emphysema 0.915 2.52
displaced fracture 0.916 1.74
central venous catheter 0.897 1.54
ankylosis 0.915 1.62
pulmonary oedema 0.853 1.46
pneumonia 0.843 1.55 joint space narrowing 0.914 1.80
cardiomegaly 0.832 1.27 scoliosis 0.913 1.96
lung collapse 0.799 1.06 avulsion fracture 0.912 1.98
pneumothorax 0.797 1.20 spondylosis 0.907 1.66
ground glass opacity 0.791 1.04 soft tissue swelling 0.907 1.90
atelectasis 0.785 1.08 spondylolisthesis 0.903 1.66
consolidation 0.781 1.05 osteopenia 0.883 1.64
pleural effusion 0.776 1.08 lytic lesion 0.881 2.51
kyphosis 0.858 1.47
degenerative change 0.846 1.22
sclerotic lesion 0.845 1.18
osteoarthritis 0.842 1.49
subluxation 0.825 1.33
dislocation 0.813 1.27
joint effusion 0.803 1.53
fracture 0.800 1.22
These concepts span a clinically meaningful range, and the model encodes each as a linear
direction that separates held-out data. Discrete, well-circumscribed findings are recovered with
particular clarity: structural and device findings such as scoliosis, hyperinflation, the endotracheal
tube, and the loose body and bone cyst in MSK exceed 0.96 with large effect sizes, and specific
fracture morphologies (comminuted, spiral, displaced) and focal lesions (lytic lesion, osteomyelitis)
are similarly well separated. Distributed, interpretation-heavy findings are encoded cleanly as
well: pulmonary oedema, pneumonia, and cardiomegaly in CXR, and degenerative changes such
as osteoarthritis, spondylosis, and joint space narrowing in MSK, all decode with ROC-AUC in
the high 0.8s to low 0.9s despite being defined by diffuse and frequently co-occurring appearances,
indicating that these directions capture the finding itself rather than a single localised cue.
Before asking whether the model uses these directions, it is worth examining how the directions
relate to one another. We measure the pairwise cosine similarity between every pair of concept
directions in each domain. The directions are distinct but structured: the mean absolute cosine
similarity is 0.335 for CXR and 0.305 for MSK, and the first three principal components capture
59% and 56% of the variance respectively. The findings occupy separate axes rather than collapsing
onto one, while still sharing a compact, low-dimensional geometry (Figures 4.1 and 4.2).
Figure 4.1: CXR concept cosine similarity.
Where directions do align, the alignment tracks clinical relationships rather than arbitrary
correlation. In CXR, the overlapping airspace and parenchymal processes group together: atelectasis
and consolidation are the most aligned pair (0.77), with lung collapse (0.69) and ground glass
opacity (0.67) nearby, while the support devices form their own cluster, the central venous catheter
and endotracheal tube aligning at 0.75. In MSK, the degenerative-spine findings cluster tightly
(osteoarthritis with joint space narrowing at 0.82, spondylosis with spondylolisthesis at 0.81, kyphosis
with spondylosis at 0.75), and the fracture family clusters separately (the generic fracture aligning
with displaced fracture at 0.74 and comminuted fracture at 0.67). The two clusters sit on opposite
sides of the space: acute fracture directions are most anti-aligned with chronic degenerative ones
(fracture versus spondylosis at −0.76), mirroring the clinical distinction between traumatic and
degenerative disease. The representation therefore organises findings the way a radiologist would
group them.
Beyond confirming that each finding is decodable, we test whether the model actually uses
these directions rather than merely carrying them. For samples negative for a concept, the probe
direction d is added to the residual stream during a forward pass as ∆ = strength · s · d at every
token position, with s the mean residual-stream norm so that strength is comparable across layers,
swept over {−1.0, 0.0, +1.0, +2.0}. This follows additive activation steering (Turner et al., 2023;
Rimsky et al., 2024) along an inference-time intervention direction (Li et al., 2023c), and the effect
is read as the shift in the concept’s first-subword token logit relative to the unsteered baseline. We
classify each concept vector by the control it affords, using suppression (∆logit < 0 at strength
−1.0) and induction (∆logit > 0 at both +1.0 and +2.0) as the two tests:
• Bidirectional: subtracting suppresses the finding and adding induces it; the model both reads
the direction and writes to it, so it is a usable handle for monitoring and for controlled intervention
in either direction.
Figure 4.2: MSK concept cosine similarity.
• Subtractive: subtracting reliably suppresses the finding, but adding does not sustain its
induction; dependable for detection and suppression, not for induction.
• Additive: adding induces the finding, but subtracting does not suppress it; usable for eliciting
or amplifying the finding, not for detection or suppression.
• None: neither direction produces the expected effect.
Tables 13 and 14 report the mean ∆logit at each strength for every concept, with its classification.
Table 13: CXR concept steering: mean ∆logit at each intervention strength.
Concept ∆logit @ −1.0 ∆logit @ +1.0 ∆logit @ +2.0 Behaviour
pacemaker −2.18 +5.86 +7.61 bidirectional
hyperinflation −3.49 +4.21 +4.71 bidirectional
pulmonary oedema −2.72 +3.12 +3.49 bidirectional
cardiomegaly −4.32 +1.36 +1.39 bidirectional
scoliosis −6.10 +2.66 +1.31 bidirectional
pleural thickening −2.35 +1.64 +1.13 bidirectional
rib fracture −3.59 +0.71 +0.95 bidirectional
endotracheal tube −1.02 +0.91 +0.50 bidirectional
pneumonia −3.82 +1.09 −0.57 subtractive
consolidation −2.22 −1.88 −1.48 subtractive
ground glass opacity −1.39 −1.24 −1.77 subtractive
subcutaneous emphysema −3.55 −0.53 −1.87 subtractive
central venous catheter −0.60 −2.13 −2.63 subtractive
pneumothorax −3.65 −0.60 −2.64 subtractive
lung collapse −3.22 −2.56 −4.54 subtractive
pleural effusion −2.01 −3.20 −4.95 subtractive
atelectasis −4.44 −7.44 −7.63 subtractive
Table 14: MSK concept steering: mean ∆logit at each intervention strength.
Concept ∆logit @ −1.0 ∆logit @ +1.0 ∆logit @ +2.0 Behaviour
soft tissue swelling −10.07 +11.41 +11.42 bidirectional
kyphosis −5.42 +6.36 +6.28 bidirectional
degenerative change −4.88 +5.14 +4.88 bidirectional
comminuted fracture −6.22 +4.47 +4.55 bidirectional
osteoarthritis −5.67 +2.68 +2.33 bidirectional
joint space narrowing −4.35 +2.33 +2.32 bidirectional
avulsion fracture −9.08 +2.26 +2.32 bidirectional
dislocation −8.28 +2.00 +2.03 bidirectional
fracture −8.86 +1.54 +1.52 bidirectional
scoliosis −9.55 +0.87 +0.93 bidirectional
subluxation −9.83 +1.08 +0.75 bidirectional
spondylolisthesis −9.56 −0.73 −1.10 subtractive
spiral fracture −5.97 −1.56 −1.57 subtractive
loose body −3.89 −3.75 −3.70 subtractive
joint replacement −7.10 −4.13 −4.16 subtractive
spondylosis −9.06 −4.85 −5.25 subtractive
displaced fracture −7.46 −5.10 −5.39 subtractive
sclerotic lesion −7.80 −5.37 −5.43 subtractive
joint effusion −5.53 −5.47 −5.49 subtractive
Concept ∆logit @ −1.0 ∆logit @ +1.0 ∆logit @ +2.0 Behaviour
lytic lesion −2.80 −5.55 −5.63 subtractive
osteopenia −8.04 −6.57 −6.93 subtractive
osteomyelitis −7.60 −7.12 −7.41 subtractive
bone cyst −4.40 −7.39 −7.44 subtractive
ankylosis −9.91 −9.74 −10.21 subtractive
In both domains subtraction lowers the corresponding logit for every concept, so each direction
is at least subtractive; no additive-only or unresponsive cases were observed. Eight of 17 CXR
concepts and 11 of 24 MSK concepts are fully bidirectional, including many clinically central findings
(pulmonary oedema, cardiomegaly, and rib fracture in CXR, and comminuted fracture, dislocation,
osteoarthritis, and a generic fracture in MSK) for which the model both reads the direction and
writes to it. The remaining concepts are dependable for detection and suppression; their use for
induction should be confirmed against generated content rather than assumed. On naturalistic
transcripts the directions also fire preferentially on the tokens where their finding is discussed (mean
peak projection ratio of present to absent cases: CXR 1.13×, MSK 1.81×), confirming that they
carry genuine signal under deployment-like conditions.
4.2 Concept Localisation: Attention as a Proxy
Localisation answers “where in the image did this answer come from?” by attributing each
generated answer, and each individual answer word, back to spatial regions of the input. The engine
offers several attribution methods that all produce the same artefact, a heatmap overlaid on the
image, but derive it from different internal signals. Offering several is deliberate: gradient-based
and attention-based attributions have different failure modes, and agreement between them is more
trustworthy than any single map.
• GradCAM (answer-level): A gradient-based saliency method in the spirit of GradCAM
(Selvaraju et al., 2017), adapted for HR1.5, attributes the model’s answer to the visual features
that most influenced it and renders them as a heatmap over the image. The heatmap is questionsensitive: asking “Is there a pneumothorax?” vs “Are there any rib fractures?” on the same
chest X-ray produces different maps focused on the relevant anatomical regions.
• Per-token gradient attribution: Because the answer is generated word by word, the same
attribution can be computed for a single selected word rather than the whole answer, isolating
the region responsible for “effusion” separately from the region responsible for “left”.
• Attention rollout (Abnar and Zuidema, 2020): Composes the model’s attention across layers
to approximate the cumulative flow of information that single-layer attention misses.
• Layer-wise Relevance Propagation (LRP) (Chefer et al., 2021): Propagates relevance
backward from the answer to the input image regions, giving a more principled attribution than
raw attention weights.
• Cross-modal attention: Reads out, directly from the model’s attention, which parts of the
image each part of the answer drew on, requiring no gradients.
Mapping back to the image: Each method produces a value per image region, which is
arranged into a spatial map and overlaid on the image. To keep the overlay faithful, regions that
do not correspond to image content are excluded before the map is formed. The map can also be
distilled into a set of discrete regions of interest, the contiguous areas of strongest attribution.
Attention as a proxy: It is well established that attention weights are not, by themselves,
faithful explanations (Jain and Wallace, 2019), though counter arguments show they are not
meaningless either (Wiegreffe and Pinter, 2019). These maps should not be read as a direct
delineation of pathology: they indicate the regions the model attended to in forming its answer, not
a segmentation of the finding itself. This distinction shapes how their quality should be measured.
Overlap metrics such as IoU or DICE, which reward pixel-accurate agreement with a reference mask,
are the wrong yardstick for an attention map: the model can attend to the correct anatomical region
without its attribution tracing the lesion’s exact boundary. The appropriate measure is a hit rate:
whether the attended region falls on the clinically relevant area the model should have considered
when forming its view. We therefore adopt hit rate as a fidelity score for attended regions, and
report it on two localisation datasets below.
We evaluate localisation on two public datasets that provide reference regions: FracAtlas
(Abedeen et al., 2023), a musculoskeletal radiograph dataset with fracture bounding boxes, for MSK;
and MS-CXR (Boecking et al., 2022), a chest X-ray phrase-grounding dataset with radiologist-drawn
bounding boxes, for CXR. A case counts as a hit when the model’s attended region falls within the
reference region for the finding. We report the localisation hit rate alongside answer accuracy on
the same cases (Table 15).
Table 15: Localisation hit rate and answer accuracy on FracAtlas (MSK) and MS-CXR (CXR).
Dataset Modality Answer accuracy Localisation hit rate
FracAtlas (Abedeen et al., 2023) MSK 90.2% 78.7%
MS-CXR (Boecking et al., 2022) CXR 61.2% 90.2%
The model attends to the clinically relevant region in the large majority of cases in both
modalities (78.7% MSK, 90.2% CXR). Notably, on MS-CXR the attended region is correct 90.2%
of the time even though answer accuracy on the same cases is lower (61.2%), indicating that the
model looks in the right place even when it does not arrive at the expected answer, evidence that
the localisation reflects genuine visual grounding rather than post-hoc rationalisation of the answer
(Figure 4.3).
These two gaps map onto the classical taxonomy of radiological interpretation error established
through eye-tracking studies of visual search (Kundel et al., 1978), which distinguishes search errors
(the finding is never fixated), recognition/perception errors (the region is fixated but the abnormality
is not perceived, a category tied in human readers to insufficient dwell time), and cognition/decision
errors (the abnormality is perceived but misinterpreted). The MSK gap, where the attended region
more often falls outside the reference, is the analogue of a search error — the model’s attention
does not reach the relevant area. The CXR gap, where the region is attended but the answer is still
wrong, is the analogue of a cognition error — the model looks in the right place but draws the
wrong conclusion. The intermediate, dwell-dependent recognition error has no natural counterpart
here: inference completes in a single rapid forward pass, so dwell time is not a meaningful quantity
for these systems.
Figure 4.3: FracAtlas localisation: ground-truth fracture regions (left of each pair) versus the model’s
attended regions (right) across a sample of cases. MS-CXR examples cannot be shown here owing to dataset
licensing restrictions, but the localisation behaviour is analogous and is deducible from the hit-rate results
above.
4.3 Confidence and Certainty about the Response
Generative models, including large language models and vision language models, do not natively
expose any signal of how confident they are in a given output. In high-stakes settings, where
each output should be reviewed for correctness and clinical impact, this absence is a real barrier
to adoption: a reader has no internal cue for which answers deserve closer scrutiny. Emerging
approaches aim to derive a meaningful confidence estimate from the model’s own computation,
but they remain early in their development and depend on substantial stabilisation and post-hoc
steps such as tuning/calibration before a model’s reported confidence can be aligned with its actual
reliability. With HR1.5 we explore this direction, with the aim of giving our users a more meaningful
signal alongside each answer. The method described below is an early implementation: it already
surfaces useful, discriminative signals, though it is not yet tuned/calibrated.
The confidence feature answers “how much should this answer be trusted?” by converting
three internal signals (Table 16), all read at the final generated token, into a single uncertainty
value, with confidence defined as its complement (Galil et al., 2023); in the spirit of Kadavath et al.
(2022), showing a model’s activations carry information about its own reliability:
Table 16: Internal signals combined into the confidence (uncertainty) feature.
Signal Description
Attention entropy Shannon entropy of the last token’s attention distribution,
normalised by log(sequence_length) to [0, 1]; diffuse attention indicates the model has not settled on what to attend
to.
Inter-head disagreement Variance across attention heads of their attention distributions for that token; when heads disagree the aggregate is
less reliable.
Representation-trajectory in- Coefficient of variation of the L2 norm of the token’s hidden
stability state across layers; spikes or collapses indicate an unsettled
representation.
These are combined as a weighted sum U = we Hattn +wd Dhead +wcv CV, inverted to confidence =
1 − U (clipped to [0, 1]), and bucketed for display as HIGH / MEDIUM / LOW. This is a
relative trust indicator, not a calibrated probability of correctness.
Why the raw score is biased by attention sinks: Transformer models exhibit attention
sinks (Xiao et al., 2024; Sun et al., 2026), a small number of tokens (often positional or low-content)
that attract a disproportionate share of attention mass regardless of semantics, tied to massive
activation spikes in a few hidden channels. A sink concentrates attention mass on one or two
tokens, making the distribution look sharp (low entropy), which the score above reads as high
confidence, but the sharpness is a structural artefact of the architecture, not evidence the model
is genuinely certain about this image and this question. Left uncorrected, sinks therefore deflate
entropy and inflate confidence, and do so most where the model is least engaged with the
content.
Sink detection and correction: The engine corrects for this in three steps:
1. Detect: for each candidate key position the received attention mass is averaged over batch,
heads, and query positions; any position exceeding a fixed threshold is flagged as a sink (the
same scan reports per-head sink ratios).
2. Mask and renormalise: the sink columns are removed from the last token’s attention row and
the remainder renormalised to a proper distribution (p˜ = p[non-sink] / Σ p[non-sink]);
entropy and head disagreement are recomputed on this reduced support. The representationtrajectory term is intrinsic to the hidden state and left unchanged.
3. Recompute and report inflation: the composite uncertainty is recalculated, and the confidence inflation inflation = max(0, (1 − Uraw ) − (1 − Ucorrected )) is reported explicitly. By
construction it is non-negative: the correction can only reveal hidden uncertainty, never manufacture confidence.
Rather than silently replacing the number, the engine surfaces both values and the gap between
them, for example, “Raw: HIGH → Corrected: MEDIUM (sinks inflate by 0.23)”, so a reader can
see when the model’s apparent certainty was an artefact. This is a post-hoc, signal-level correction
complementary to, not a substitute for, formal probability calibration such as temperature scaling
on held-out data (Guo et al., 2017).
Worked example, see figure 4.4: For the chest X-ray below, asked “Describe the findings in
this chest x-ray,” HR1.5 responded with “The chest x-ray shows a 3.5cm mass in the left upper lobe,
which is suspicious for malignancy. There are no other pulmonary nodules or masses identified. No
pleural effusion is present. The cardiac size and mediastinal contours appear normal.”
Figure 4.4: Left apical lung mass case. Case courtesy of Dai Roberts, Radiopaedia.org (case 80445), used
under licence.
The engine reports a raw confidence of 0.81. Detecting and correcting for attention sinks lowers
this to 0.79, an inflation of 0.03. A small portion of the apparent certainty was a structural artefact
rather than genuine engagement with the image and question. The modest gap leaves the overall
trust assessment unchanged, while making explicit that the corrected figure is the more faithful one.
On the follow-up question “Are you sure about the size of the mass?”, the model answered “Yes,
the mass in the left upper lobe measures 3.5cm,” with a raw confidence of 0.82 corrected to 0.80.
The confidence holds steady under direct challenge, and the sink correction again accounts for only
a small (0.03) inflation, indicating the model’s certainty about the measurement is genuine rather
than an attention artefact.
Activation highlighting for the lung mass case indicates the image regions the model draws on
when forming its answers, concentrated over the left upper lobe mass described in the response.
Worked example, shown in figure 4.5, a multi-turn conversation: The exchange below
is on a Weber B ankle fracture case. Each turn shows the question, the model’s answer, and the
raw and sink-corrected confidence.
Figure 4.5: Ankle Weber B fracture. Case courtesy of Frank Gaillard, Radiopaedia.org (case 7480), used
under licence.
Across the conversation the corrected confidence stays in a narrow band (0.80–0.82) with a
consistent, small sink inflation (∼0.02), showing that the model’s certainty is stable as the dialogue
moves from description to classification to a full report, and does not spuriously spike. The activation
highlighting shows where in the image the model draws this information from, and how that focus
shifts as the question changes between the opening description and the final challenge. Activation for
the opening question (“Describe the findings”) is spread across the distal fibula and both malleoli,
where the fractures and soft-tissue swelling are described; activation for the fourth question (the
soft-tissue swelling query) is narrowed toward the malleoli where the swelling is described, illustrating
that the model re-focuses on the area in question as the dialogue shifts.
Worked example, shown in figure 4.6, an out-of-distribution input: To probe how the
confidence signal behaves on atypical inputs, we presented the first medical radiograph ever taken
(Röntgen’s 1895 X-ray of a hand) with a deliberately off-domain question.
The answer is correct, but the confidence falls to 0.76, noticeably below the 0.80+ reported on
the in-distribution clinical cases above. The lower figure tracks the unusual, low-contrast, historically
Figure 4.6: Querying with the historical first Röntgen’s 1895 X-ray of a hand. Underlying radiograph:
Wikimedia Commons (Wilhelm Röntgen, X-ray of Albert von Kölliker’s hand, 1896).
degraded image and the off-domain question: the confidence signal registers that this input sits
outside the model’s familiar distribution, which is exactly the behaviour a trustworthy uncertainty
estimate should show.
While these results make this a promising direction for ascertaining confidence, the approach
remains at an early stage and will require further development before it can serve as a consistent,
interpretable indicator of confidence. We are optimistic that, with further stabilisation and calibration, these signals can be made substantially more meaningful and reliable, and refining them is an
area we intend to focus on in future work.
4.4 Behaviour on Out-of-Distribution Inputs
A model’s response to inputs it could never have seen in training is as revealing as its performance
on genuine studies. To probe this, we presented a synthetic chest X-ray generated with the GPT-5.5
image API from the prompt “generate a realistic looking x-ray of an apple in a human chest”, an
image that looks radiographic but contains an object with no counterpart in any real study (figure
4.7).
The model accepts the synthetic image as a genuine radiograph and, when reporting findings,
maps the unfamiliar object onto the nearest concept it has learnt (a left breast prosthesis) holding
that interpretation with high confidence even when directly challenged. This exposes a real limitation:
on concepts entirely outside its training distribution, the model is incorrect. Importantly, though, the
error is a reasonable one: the synthetically generated apple has overlapping imaging characteristics
with breast prosthesis or implant. The output is plausible given the model training data, rather
than being outright nonsensical.
This points to a limitation that no confidence signal can fully remove: when an out-of-distribution
input maps plausibly onto a learnt concept, as the apple does onto a prosthesis, the model can
be confidently wrong, so a high value alone does not guarantee correctness. As the Röntgen case
above shows, the signal can sometimes register an unfamiliar input as less certain, though not yet
reliably. In our own system we already gate on this, rejecting inputs identified as out-of-distribution
so that unreliable cases are held back rather than reported. Calibrating these confidence signals
and measuring their reliability quantitatively remains future work; what we present here is an early,
qualitative result.
Figure 4.7: Query with a synthetically generated x-ray image. The activation overlay is concentrated on
the apple-shaped opacity that the model labels a prosthesis, confirming that the confident misinterpretation
is driven by the anomalous object itself, and that the model substitutes its closest learnt concept for one it
was never trained to recognise.
5 Discussion
Considered in aggregate, the evaluations support a consistent conclusion: HR1.5 attains its largest
advantages on the tasks of greatest clinical consequence. On our internal, held-out mock simulations
of the FRCR examinations, which reconstruct the assessments through which the Royal College of
Radiologists certifies practising radiologists and serve as the closest available proxy in this study to
authentic diagnostic reporting, HR1.5 and its agentic configuration HR1.5+ are the only systems to
meet the passing standard, satisfying the Angoff-referenced threshold of the current 2B Short Case
format. Every other general-purpose and medical-domain model evaluated falls substantially below
it. Because this benchmark most directly approximates the clinical reporting task, the magnitude
of this margin is of particular practical significance. We note that these examinations are internally
constructed simulations rather than the official RCR sittings, which are not publicly available
(Section 2.2.1).
This advantage generalises beyond the examination setting. On closed-format clinical questions
and on the multi-body-part internal evaluation, HR1.5 attains the highest accuracy across diagnosis,
finding presence, and normal/abnormal classification, and across chest, musculoskeletal, and other
anatomical regions, indicating that the improvement is not specific to a single modality or question
type.
We are equally explicit about the settings in which HR1.5 does not lead. On a few openended chest-radiograph report-generation benchmarks, frontier general-purpose models and chestspecialised systems achieve higher values on individual metrics, and the agentic HR1.5+ configuration
does not uniformly improve upon HR1.5, most notably under multi-turn dialogue evaluation. We
attribute this in part to genuine headroom that remains to be addressed, and in part to the metric
limitations characterised in Section 2.1, in which surface- and entity-level scores reward lexical form
over clinical correctness. Neither consideration qualifies the principal result: on the task that most
closely resembles, and is most likely to inform, routine radiological practice (examination-grade
reporting) HR1.5 outperforms all evaluated systems by a substantial margin.
6 Future Work
6.1 Towards Better Evaluation: The C6 Framework
Section 2.1 established why automated metrics diverge from clinical utility: conventional reportgeneration metrics track wording far more than clinical correctness, and even the Findings-Diagnosis
score we adopt as the primary measure, with the Judge Findings-Diagnosis variant where automated
matching is insufficient, remains a single aggregate. As noted there, it inherits the central limitation
of all scalar scores: it can credit a report that misses a subtle finding, can accept a vague but
incorrect descriptor as a match, and assigns comparable penalties to errors of very different clinical
consequence. A report that identifies the correct diagnosis but omits a minor finding is scored
similarly to one that hallucinates a dangerous pathology. A scalar score, however well-aligned, does
not indicate why a report is adequate or inadequate: whether it identified the clinically important
findings, whether it would be safe to act upon, and whether a radiologist would accept it.
Recent work has reached the same conclusion, that single-score metrics are insufficient for
clinical text, and has begun to decompose evaluation into interpretable axes. DocLens (Xie et al.,
2024) evaluates medical text along completeness, conciseness, and attribution; GREEN (Ostmeier
et al., 2024) introduces error-aware evaluation with interpretable explanations; RadCliQ (Yu et al.,
2023) combines several signals into a clinically grounded composite; and CT-FineBench (2026)
demonstrates that conventional metrics fail to capture granular diagnostic accuracy in CT report
generation. These efforts point toward a consensus: clinically meaningful evaluation requires
fine-grained, multi-dimensional assessment rather than a single opaque number.
To this end, we are developing the C6 evaluation framework, which scores a draft report along
six interpretable clinical pillars (Table 17) and combines them into a single composite (Figure 6.1).
Figure 6.1: The C6 framework. A draft report is scored along six clinical pillars — Coverage, Correctness,
Comprehensiveness, Care, Coherence, and Conformity — each yieldingP a pillar score pi . A weighted sum
with deployment-tunable weights wi produces a single Credibility score, i wi pi .
Table 17: The six clinical pillars of the C6 framework.
Pillar Evaluation question
Coverage Are all findings that should be reported identified? (recall)
Correctness Are the reported findings true, with no hallucinated pathology? (precision)
Comprehensiveness Are correctly identified findings described in sufficient
detail to act on?
Care What is the patient-safety impact of any errors?
Coherence Is the report clear, internally consistent, and wellcommunicated?
Conformity Does the reporting style match institutional expectations?
The six pillars can be evaluated independently or combined into an overall Credibility score,
P
defined as the weighted sum Credibility = i wi pi of the per-pillar scores. The decomposition is
what makes the framework diagnostic rather than merely summative: it separates failure modes
that a scalar metric conflates, distinguishing a model with high Coverage but low Correctness
(over-calling) from one with high Correctness but low Comprehensiveness (under-describing). The
Care pillar introduces explicit clinical severity weighting, absent from existing metrics, so that safetycritical errors such as a missed pneumothorax are penalised more heavily than benign omissions or
stylistic divergence.
The weights wi are tunable per deployment, reflecting that different clinical settings require
different emphases: emergency triage prioritises Care and Coverage, where detecting an acute,
actionable finding is paramount, whereas oncology surveillance emphasises Comprehensiveness and
Correctness, where precise characterisation and interval comparison are most important.
Underlying all six pillars is a radiology knowledge base that anchors each extracted finding to a
stable clinical concept, extending the ontology-based synonym matching and polarity logic already
employed by the Findings-Diagnosis score (Section 2). This grounding enables C6 to recognise
that “no airspace opacification” contradicts “consolidation” despite the lexical dissimilarity, and
conversely that “opacity” and “consolidation” are clinically related rather than contradictory, so
that related descriptors are not penalised as opposing assertions.
We intend to adopt C6 as the primary evaluation framework for future model iterations and
to release it as an open evaluation tool for the broader community, providing a shared, clinically
grounded basis for assessing radiology report quality.
6.2 3D and Volumetric Imaging Support
HR1.5 is trained exclusively on 2D X-ray modalities. A natural extension is to support 3D
volumetric imaging, particularly CT scans, which represent a substantial and growing share of
radiology workload. Recent work has demonstrated the feasibility of foundation models for 3D
radiology: CT2Rep (Hamamci et al., 2024) introduced automated report generation for 3D medical
images, Argus (2024) established the first comprehensive benchmark for 3D radiograph report
generation, and several 2026 efforts have tackled challenges specific to volumetric data including
entity hallucination mitigation, region-grounded report generation, and hierarchical vision-language
modelling for efficient volume processing.
Extending HR1.5 to 3D introduces significant architectural and data challenges. Volumetric
inputs are orders of magnitude larger than single X-ray images, requiring efficient spatial encoding
strategies, whether through hierarchical 2.5D approaches, sparse attention over volume patches,
or multi-view aggregation. The training data pipeline must be adapted for DICOM series with
slice-level annotations and volumetric spatial relationships.
We plan to address these challenges in future model iterations, with the goal of delivering a
unified radiology foundation model capable of interpreting both 2D and 3D imaging modalities
within a single architecture.
Acknowledgements
We thank the radiologists and clinical collaborators whose expertise informed the design, scoring,
and evaluation of HR1.5, and our colleagues across Harrison.ai for their engineering and research
contributions. We gratefully acknowledge the open-source community, in particular PyTorch and
Hugging Face and their contributors, whose frameworks supported the training, development, and
evaluation of HR1.5, and the vLLM project and its contributors, whose tools underpin our serving
stack, and Radiopaedia.org and its contributors for the teaching cases used under licence in the
qualitative examples.
References
Iftekharul Abedeen et al. FracAtlas: A dataset for fracture classification, localization and segmentation of musculoskeletal radiographs. Scientific Data, 2023. URL https://pmc.ncbi.nlm.nih.
gov/articles/PMC10404222/.
Samira Abnar and Willem Zuidema. Quantifying attention flow in transformers. In Annual Meeting
of the Association for Computational Linguistics (ACL), 2020.
Guillaume Alain and Yoshua Bengio. Understanding intermediate layers using linear classifier
probes. In ICLR Workshop, 2017. arXiv:1610.01644.
Jean-Baptiste Alayrac et al. Flamingo: a visual language model for few-shot learning. In Advances
in Neural Information Processing Systems (NeurIPS), 2022.
William H. Angoff. Scales, norms, and equivalent scores. In Educational Measurement. 1971.
Anthropic. Claude Opus 4.7. https://www.anthropic.com/claude, 2026.
Argus Authors. Argus: A comprehensive benchmark for 3d radiograph report generation. arXiv
preprint arXiv:2406.07146, 2024. arXiv:2406.07146.
Shruthi Bannur et al. MAIRA-2: Grounded radiology report generation. arXiv preprint
arXiv:2406.04449, 2024.
Lucas Beyer et al. PaliGemma: A versatile 3b vlm for transfer. arXiv preprint arXiv:2407.07726,
2024.
Benedikt Boecking et al. Making the most of text semantics to improve biomedical vision-language
processing (MS-CXR). In European Conference on Computer Vision (ECCV), 2022. URL
https://physionet.org/content/ms-cxr/1.1.0/.
Hila Chefer, Shir Gur, and Lior Wolf. Transformer interpretability beyond attention visualization.
In Conference on Computer Vision and Pattern Recognition (CVPR), 2021.
Zhihong Chen et al. A vision-language foundation model to enhance efficiency of chest x-ray
interpretation (CheXagent). 2024. arXiv:2401.12208.
CT-FineBench Authors. CT-FineBench: Fine-grained evaluation of ct report generation. arXiv
preprint arXiv:2604.24001, 2026. arXiv:2604.24001.
Jean-Benoit Delbrouck et al. RadGraph-XL: A large-scale expert-annotated dataset for entity and
relation extraction from radiology reports. In Findings of the Association for Computational
Linguistics (ACL), 2024.
Alexey Dosovitskiy et al. An image is worth 16x16 words: Transformers for image recognition at scale
(ViT). In International Conference on Learning Representations (ICLR), 2021. arXiv:2010.11929.
Ido Galil et al. A framework for benchmarking class-out-of-distribution detection and its application
to ImageNet. In International Conference on Learning Representations (ICLR), 2023.
Marzyeh Ghassemi, Luke Oakden-Rayner, and Andrew L. Beam. The false hope of current approaches
to explainable artificial intelligence in health care. The Lancet Digital Health, 3(11):e745–e750,
2021. doi: 10.1016/S2589-7500(21)00208-9.
Google. Gemini 3 Flash (preview). https://deepmind.google/technologies/gemini/, 2025.
Jiawei Gu et al. A survey on LLM-as-a-Judge. arXiv preprint arXiv:2411.15594, 2024.
Chuan Guo et al. On calibration of modern neural networks. In International Conference on
Machine Learning (ICML), 2017. arXiv:1706.04599.
Ibrahim Ethem Hamamci et al. CT2Rep: Automated radiology report generation for 3d medical
imaging. In Medical Image Computing and Computer Assisted Intervention (MICCAI), 2024.
Harrison.ai. Reimagining medical ai: Harrison.rad.1, a large multimodal foundation model designed
to excel in radiology, 2024. URL https://harrison.ai/radiology-foundation-model/.
Lisa Anne Hendricks et al. Generating visual explanations. In European Conference on Computer
Vision (ECCV), 2016. arXiv:1603.08507.
others Hong. Generative artificial intelligence models for chest radiograph interpretation: Comparison
of a general-purpose model, domain-specific model, and two agentic models. American Journal of
Roentgenology, 2026. URL https://doi.org/10.2214/AJR.25.34267.
Stephanie L. Hyland et al. MAIRA-1: A specialised large multimodal model for radiology report
generation. arXiv preprint arXiv:2311.13668, 2023.
Saahil Jain et al. RadGraph: Extracting clinical entities and relations from radiology reports. In
NeurIPS Datasets and Benchmarks, 2021.
Sarthak Jain and Byron C. Wallace. Attention is not explanation. In NAACL-HLT, 2019.
arXiv:1902.10186.
Saurav Kadavath et al. Language models (mostly) know what they know. arXiv preprint
arXiv:2207.05221, 2022.
Elizabeth A. Krupinski et al. Long radiology workdays reduce detection and accommodation
accuracy. Journal of the American College of Radiology, 7(9):698–704, 2010.
Harold L. Kundel, Calvin F. Nodine, and Dennis Carmody. Visual scanning, pattern recognition
and decision-making in pulmonary nodule detection. Investigative Radiology, 1978.
Woosuk Kwon et al. Efficient memory management for large language model serving with PagedAttention (vLLM). In ACM Symposium on Operating Systems Principles (SOSP), 2023. URL
https://github.com/vllm-project/vllm. arXiv:2309.06180.
Sicong Leng et al. Mitigating object hallucinations in large vision-language models through visual
contrastive decoding (VCD). In Conference on Computer Vision and Pattern Recognition (CVPR),
2024. arXiv:2309.09117.
Chunyuan Li et al. LLaVA-Med: Training a large language-and-vision assistant for biomedicine in
one day. arXiv preprint arXiv:2306.00890, 2023a.
Junnan Li et al. BLIP-2: Bootstrapping language-image pre-training with frozen image encoders
and large language models. In International Conference on Machine Learning (ICML), 2023b.
arXiv:2301.12597.
Kenneth Li et al. Inference-time intervention: Eliciting truthful answers from a language model. In
Advances in Neural Information Processing Systems (NeurIPS), 2023c. arXiv:2306.03341.
Xiang Lisa Li et al. Contrastive decoding: Open-ended text generation as optimization. In Annual
Meeting of the Association for Computational Linguistics (ACL), 2023d. arXiv:2210.15097.
Chin-Yew Lin. ROUGE: A package for automatic evaluation of summaries. In Text Summarization
Branches Out, ACL Workshop, 2004.
Haotian Liu et al. Visual instruction tuning (LLaVA). In Advances in Neural Information Processing
Systems (NeurIPS), 2023. arXiv:2304.08485.
Samuel Marks and Max Tegmark. The geometry of truth: Emergent linear structure in large
language model representations of true/false datasets. arXiv preprint arXiv:2310.06824, 2023.
Mass General Brigham AI Arena and American College of Radiology. Healthcare ai challenge —
results, 2025. URL https://healthcareaichallenge.org/ai-results/.
Meta AI. The Llama 4 herd. https://ai.meta.com/blog/llama-4/, 2025.
Neiman Health Policy Institute. Projected US imaging utilization, 2025 to 2055. Journal of the
American College of Radiology, 2025. URL https://www.neimanhpi.org/.
OpenAI. GPT-4 technical report. arXiv preprint arXiv:2303.08774, 2023.
OpenAI. GPT-5.4. https://openai.com, 2026.
Sophie Ostmeier et al. GREEN: Generative radiology report evaluation and error notation. In
Findings of the Association for Computational Linguistics (EMNLP), 2024.
Kishore Papineni, Salim Roukos, Todd Ward, and Wei-Jing Zhu. BLEU: a method for automatic
evaluation of machine translation. In Annual Meeting of the Association for Computational
Linguistics (ACL), 2002.
Alec Radford et al. Learning transferable visual models from natural language supervision (CLIP).
In International Conference on Machine Learning (ICML), 2021.
Mauricio Reyes et al. On the interpretability of artificial intelligence in radiology: Challenges and
opportunities. Radiology: Artificial Intelligence, 2(3):e190043, 2020. doi: 10.1148/ryai.2020190043.
Nina Rimsky et al. Steering Llama 2 via contrastive activation addition. In Annual Meeting of the
Association for Computational Linguistics (ACL), 2024. arXiv:2312.06681.
Royal College of Radiologists. Clinical radiology workforce census 2025. Technical report,
Royal College of Radiologists, London, 2025. URL https://www.rcr.ac.uk/media/n1fjvrv4/
rcr-2025-clinical-radiology-workforce-census-report.pdf.
Khaled Saab et al. Capabilities of Gemini models in medicine (Med-Gemini). arXiv preprint
arXiv:2404.18416, 2024.
Andrew Sellergren et al. MedGemma technical report. arXiv preprint arXiv:2507.05201, 2025.
Ramprasaath R. Selvaraju et al. Grad-CAM: Visual explanations from deep networks via
gradient-based localization. In International Conference on Computer Vision (ICCV), 2017.
arXiv:1610.02391.
Akshay Smit et al. CheXbert: Combining automatic labelers and expert annotations for accurate
radiology report labeling using BERT. In Conference on Empirical Methods in Natural Language
Processing (EMNLP), 2020.
Mingjie Sun et al. The spike, the sparse and the sink: Anatomy of massive activations and attention
sinks. arXiv preprint arXiv:2603.05498, 2026.
Ekin Tiu et al. Expert-level detection of pathologies from unannotated chest x-ray images via
self-supervised learning (CheXzero). Nature Biomedical Engineering, 6:1399–1406, 2022. URL
https://www.nature.com/articles/s41551-022-00936-9.
Sana Tonekaboni et al. What clinicians want: Contextualizing explainable machine learning for
clinical end use. In Machine Learning for Healthcare (MLHC), PMLR 106, 2019. arXiv:1905.05134.
Michael Tschannen et al. SigLIP 2: Multilingual vision-language encoders with improved semantic
understanding, localization, and dense features. arXiv preprint arXiv:2502.14786, 2025.
Alexander Matt Turner et al. Activation addition: Steering language models without optimization.
arXiv preprint arXiv:2308.10248, 2023.
Ashish Vaswani et al. Attention is all you need. In Advances in Neural Information Processing
Systems (NeurIPS), 2017. arXiv:1706.03762.
others Wang. A fully open ai foundation model applied to chest radiography. Nature, 2025. URL
https://www.nature.com/articles/s41586-025-09079-8.
Peng Wang et al. Qwen2-VL: Enhancing vision-language model’s perception of the world at any
resolution. arXiv preprint arXiv:2409.12191, 2024.
Sarah Wiegreffe and Yuval Pinter. Attention is not not explanation. In Conference on Empirical
Methods in Natural Language Processing (EMNLP), 2019. arXiv:1908.04626.
Chaoyi Wu et al. Towards generalist foundation model for radiology by leveraging web-scale 2d &
3d medical data (RadFM). arXiv preprint arXiv:2308.02463, 2023.
Guangxuan Xiao et al. Efficient streaming language models with attention sinks. In International
Conference on Learning Representations (ICLR), 2024. arXiv:2309.17453.
Yiqing Xie et al. DocLens: Multi-aspect fine-grained evaluation for medical text generation. In
Annual Meeting of the Association for Computational Linguistics (ACL), 2024.
Shawn Xu et al. ELIXR: Towards a general purpose x-ray artificial intelligence system through
alignment of large language models and radiology vision encoders. arXiv preprint arXiv:2308.01317,
2023.
Lin Yang et al. Advancing multimodal medical capabilities of Gemini. arXiv preprint
arXiv:2405.03162, 2024.
Feiyang Yu et al. Evaluating progress in automatic chest x-ray radiology report generation (RadCliQ).
Patterns, 2023.
Mert Yuksekgonul et al. When and why vision-language models behave like bags-of-words (NegCLIP).
In International Conference on Learning Representations (ICLR), 2023.
Sheng Zhang et al. BiomedCLIP: A multimodal biomedical foundation model pretrained from fifteen
million scientific image-text pairs. arXiv preprint arXiv:2303.00915, 2023.
Xiaoman Zhang et al. ReXGradient-160K: A large-scale publicly available dataset of chest radiographs
with free-text reports. arXiv preprint arXiv:2505.00228, 2025.
Xiaoman Zhang et al. A reasoning-enabled vision-language foundation model for chest x-ray
interpretation (CheXOne). arXiv preprint arXiv:2604.00493, 2026.