Themes it raises

5 of the 21 themes in the docket, each with the passage we counted, verbatim.
What makes a function high riskFDA Q1, Q2, Q5
“The framework should name changed-treatment-decision as a consequence category, not just device failure modes.”
Whether benchmark results prove anythingFDA Q9, Q10, Q16
“Standard accuracy benchmarks will not surface these. Evaluation should include structured adversarial cases drawn from real psychiatric presentations.”
Harm from an output that was not wrongFDA Q1, Q2
“(a) sycophantic outputs that affirm delusional or grandiose thinking, which feel "helpful" while being dangerous;”
Whether the user can judge the outputFDA Q3, Q4
“(b) confident de-escalation advice that is subtly wrong in ways only a trained psychiatric clinician would catch.”
Watching the device after it shipsFDA Q19, Q20
“Performance degradation across the product life cycle will not be visible in lab conditions. Require sponsors to monitor device outputs in deployed settings”
Machine-assisted draft, pending human review. The source text and highlighted passages appear below. Read the filing on regulations.gov ↗

The comment as filed

Comment submitted on regulations.gov. Passages we counted are highlighted.

I am a board-certified psychiatric-mental health nurse practitioner and family nurse practitioner operating a telehealth psychiatry practice across six states. I use AI tools daily in clinical workflows. I write to urge FDA to account for mental-health realities as it develops its framework for generative AI-enabled devices.

**1. The risk framework’s consequence axis underweights psychiatric harm.**
The discussion paper’s two-axis framework — degree of device activity versus consequences of relying on an incorrect output — is sound, but "consequences" reads as if harm means a device malfunction. In psychiatry, the harm pathway is a changed clinical decision. A confabulated but fluent medication summary, risk assessment, or treatment suggestion does not look broken; it looks authoritative. There is no lab value, image, or vital sign downstream to catch it — only a clinician’s judgment, often exercised under time pressure, over video, with a patient in crisis. The framework should name changed-treatment-decision as a consequence category, not just device failure modes.

**2. Premarket evaluation must include mental-health-specific adversarial testing.**
Sponsors should be required to test how a device performs with patients expressing suicidality, psychosis, and mania — not just general medical Q&A. Two failure modes deserve explicit attention: (a) sycophantic outputs that affirm delusional or grandiose thinking, which feel "helpful" while being dangerous; and (b) confident de-escalation advice that is subtly wrong in ways only a trained psychiatric clinician would catch. Standard accuracy benchmarks will not surface these. Evaluation should include structured adversarial cases drawn from real psychiatric presentations.

**3. Postmarket monitoring must happen in real telehealth workflows.**
Performance degradation across the product life cycle will not be visible in lab conditions. Require sponsors to monitor device outputs in deployed settings — including telehealth, where the clinician cannot read the room the way an in-person visit allows — across diverse patient populations. Critically, uncertainty must be visible to the clinician using the tool, not buried. A device that reports confidence it does not have is more dangerous than one that admits limits.

**4. Treat the clinician handoff as part of the device’s risk surface.**
For any function above nondirective information, the person answering the patient’s "what does this mean?" is usually a nurse or nurse practitioner. The framework should treat that handoff — what the clinician sees, what caveats travel with the output, how disagreement is surfaced — as part of what gets evaluated, not as someone else’s problem.

**5. Solicit the practitioner voice before guidance.**
The clinicians who will live with these tools are underrepresented in this docket. I urge the Center to actively solicit comment from practicing nurses, nurse practitioners, and frontline prescribers — not only manufacturers and health-system executives — before drafting guidance. We are the ones who will explain these outputs to patients, notice when they are wrong, and decide what to do next.

Respectfully submitted,
Brandon Kruse, PMHNP-BC, FNP-BC