ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical
Devices: Discussion Paper and Request for Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
Executive Summary of ISPE Comments
The International Society for Pharmacoepidemiology (ISPE) respectfully submits the following
comments in response to the discussion paper, "Considerations for the Regulation of Generative
AI-Enabled Medical Devices: Discussion Paper and Request for Feedback," issued by the Center
for Devices and Radiological Health (CDRH) in August 2026 (Docket No. FDA-2026-N-7874).
ISPE is a global scientific society dedicated to advancing the science and practice of
pharmacoepidemiology and the use of real-world evidence (RWE) to inform decisions about
human health and the use of medical products. These comments draw on the membership's
expertise in observational studies, comparator selection, measurement validity, subgroup
identification and assessment, the use of real-world data, and postmarket surveillance, the
methods on which many of the approaches described in this paper ultimately depend. ISPE
commends CDRH for seeking early input before settling on a regulatory approach, and for
recognizing that evaluation methods developed for software with bounded inputs and fixed
outputs may not be adequate for generative AI-enabled devices. We offer the following
recommendations to strengthen the methodological rigor, clarity, and practical applicability of
the approaches described.
Apply risk assessment across the lifecycle, using worked examples and postmarket
evidence to reassess risk and safeguards.
Strengthen clinical evaluation through meaningful performance measures, validated
scoring, appropriate comparators, subgroup assessment, and detailed characterization of
the intended use population under routine clinical care.
Consider rigorous non-U.S. real-world evidence as primary and/or supplementary
evidence of clinical performance and safety when data are relevant and reliable and
findings apply to intended U.S. use.
Ensure actionable, feasible postmarket monitoring, with timely detection and mitigation,
adequate resources, and clear manufacturer accountability.
Use synthetic data for targeted supplementary evaluation, with real-patient validation and
explicit acknowledgment of remaining evidence gaps.
Match clinical evidence to intended use and risk, retaining adequate premarket
confirmation where errors could cause serious or irreversible harm.
The following table presents ISPE's detailed comments on the discussion questions and additional
points of feedback found throughout the discussion paper:
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
Discussion Questions ISPE Response
Does the two-axis risk framework, organized
around device activity and the consequence of The two-axis framework provides a useful starting point for organizing the risks
of GenAI-enabled device software functions. The examples accompanying the
relying on an incorrect output, appropriately
framework are helpful. We recommend extending Figure 1 into a matrix/table
capture the dimensions most relevant to the risk of with worked examples across the activity categories and consequence levels.
a GenAI-enabled software function? If there are We also recommend that CDRH clarify how the framework aligns with the
additional dimensions—such as the reversibility existing risk-based device classification system (Class I, II, and III), including
of a resulting action, the availability of worked examples illustrating how a function’s position within the framework
downstream safeguards, the time pressure of the relates to device classification and the associated evidence requirements and
deployment setting, or the traceability of the regulatory controls.
output (i.e., to primary source materials)—that Additionally, we recommend explicitly describing the two-axis framework as a
should be represented in a risk framework, please lifecycle assessment that is revisited as postmarket evidence accumulates.
describe and provide examples of how they should Initial placement on the axis would reflect the intended use and evidence
be represented. available before deployment. Postmarket monitoring should inform
reassessment. CDRH should clarify when such findings would warrant a change
in the software function’s position within the framework and when they would
instead suggest additional recommendations/safeguards to support keeping its
position.
CDRH is considering that clinical confirmation To select and justify a clinical confirmation approach tailored to a device’s
for a GenAI-enabled device might not require a intended use and proportionate to its risk profile, a sponsor would need to
prospective clinical study in every case, and has consider the following:
- Whether a non-GenAI version of the device exists and has been
described above a range of approaches of
previously cleared or approved for its intended use by CDRH; this
increasing rigor and patient exposure. How might would inform whether existing performance data from routine clinical
a sponsor select and justify a confirmation care might be available to serve as a retrospective comparison group
approach tailored to a device’s intended use and - Whether the GenAI-enabled device integrates into an existing clinical
proportionate to the device’s risk profile? Are workflow or whether it introduces a new step or sufficiently changes an
there device types or risk profiles for which one or existing clinical workflow; this would inform whether an approach such
as a shadow deployment could be used for clinical confirmation
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
more of these approaches would be insufficient or - Whether the indicated use population is well understood in terms of
inappropriate? How might the anticipated clinical and demographic characteristics, current standard of care, and
distribution of real-world inputs be taken into natural history of disease; this would inform whether an approach using
real world patient data could be used and evaluated for the
consideration? Are there other methods of clinical
representativeness of a selected real world data source and the
confirmation that might help inform the corresponding patient cohort. It could also inform assessment of the
evaluation of GenAI-enabled devices? representativeness of an approach such as standardized patient
interactions. Finally, this could also inform an understanding of the
degree of heterogeneity in clinical care for the intended use population
and thus, whether clinical adjudication of real use cases is an appropriate
approach.
- Whether the GenAI-enabled device is functioning to estimate prognosis
or functioning to recommend or operationally trigger clinical action.
- For a prognostic function that reports risk or expected course
without recommending an action, confirmation should follow
the prediction-model practice, that independent external
validation using data that reproduce the information available at
the intended prediction time, with assessment of calibration,
discrimination, prediction error, and clinically important
subgroups. [1]
- A function should enter the treatment-decision pathway when it
recommends treatment, referral, timing, or dosage, or directly
triggers a high-risk action. For such functions, retrospective real
cases and independent specialist adjudication against the
applicable standard of care can establish an output-quality and
premarket safety floor, but cannot establish clinical benefit. If
the device can change treatment or referral, act autonomously,
or cause serious or irreversible harm when wrong, confirmation
should include a prospective comparison under actual-use
conditions and, when ethical and feasible, a randomized trial.
Shadow deployment, standardized patients, and offline expert
review may provide supporting evidence, but should not replace
this evaluation.
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
The approach for clinical confirmation taken by a sponsor should not only be
proportionate to the device’s known risk profile, according to the intended use
of the device or predicate devices which are not GenAI-enabled, but should also
be proportionate to the device class (tied to risk) as well as how disruptive the
GenAI-enabled device is to existing clinical workflows.
Sponsors should prespecify the anticipated distribution of patients, settings,
users, input formats, and rare high-consequence scenarios, and should show that
the confirmation data adequately cover that distribution. The labeled intended
use should be limited to populations and input conditions directly supported by
the evidence. Unevaluated populations or scenarios should be excluded from
the initial indication and considered later through postmarket evidence and
formal indication expansion rather than inferred from average performance.
Where clinician oversight is claimed as a safeguard, standalone device
performance and clinician-AI team performance should be evaluated
separately, and the intended users' ability to detect and correct unsafe outputs
should be demonstrated [2]. Postmarket monitoring may detect drift and support
later expansion, but should not replace adequate premarket confirmation for
high-risk treatment-decision functions.
Additional methods of clinical confirmation may include the use of a real-world
external control arm compared to a single-arm prospective clinical study of the
GenAI-enabled device, the use of a real-world external control arm to serve as
representative contextual information on the intended use population in addition
to clinical adjudication of real use cases, standardized patient interactions,
and/or shadow deployment.
References
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
1. Riley et al. Evaluation of clinical prediction models (part 2): how to undertake
an external validation study. BMJ. 2024;384:e074820.
2.Vasey et al. DECIDE-AI. Nature Medicine. 2022;28:924–933.
How should sponsors achieve statistically Statistically meaningful performance measurement should use prespecified
meaningful performance measurement for measures and acceptance criteria aligned with the device’s intended use and
GenAI-enabled devices? Where synthetically clinical risks, supported by confidence intervals rather than a single aggregate
score or p-value. Diagnostic functions should follow established device and
generated inputs supplement real patient data,
IVD principles: sensitivity and specificity against an appropriate reference
how should sponsors account for differences standard, or positive and negative percent agreement against a non-reference
between the synthetic and real-world distributions comparator. Other functions should use appropriate measures, including
when estimating performance, and under what calibration and discrimination for prognostic outputs [1], and clinically
conditions, if any, is it appropriate to combine meaningful error rates and severity for open-ended outputs. Sponsors should
benchmarking evidence and clinical confirmation prespecify the population, device version, sampling design, unit of analysis,
evidence to support a single performance denominators, and handling of unevaluable outputs and repeated observations
across cases, users, and raters. Sample sizes should support adequate precision
estimate? and assessment against acceptance criteria. Where applicable, standalone
device and clinician–AI team performance should be reported separately.
The primary performance estimate should be based on real patient data sampled
or weighted to a prespecified anticipated distribution of settings, users, case
complexity, missing information, and input types as aligned with the
distribution of those characteristics among a real-world intended use
population. Adequate precision is required for the overall intended-use
population and for any subgroup-specific claim. Other subgroup analyses may
be descriptive, but an anomalous point estimate should prompt investigation,
additional data, or restriction of labeled use rather than dismissal as
underpowered noise. Rare but high-consequence scenarios may be enriched for
stress testing but should be reported separately.
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
Synthetic inputs should be generated and curated independently of the evaluated
device and, where a generative model is used, should come from an independent
third-party model that is neither the tested device nor a closely related model.
They should be used only for separately reported stress tests, should not enter
overall or subgroup performance estimates, and should not support claims
without real-patient anchors.
The evidence needed should reflect the device’s intended use, risk, and
applicable regulatory pathway, including use of justified objective performance
criteria or performance goals where appropriate. When benchmarking and
clinical confirmation evidence are both available, they should generally be
reported separately. A combined performance estimate should be considered
only when the sources assess the same performance measure in sufficiently
comparable populations and conditions, with methods that account for
differences between sources and any overlapping cases.
References
1. Riley et al. Evaluation of clinical prediction models (part 2): how to undertake
an external validation study. BMJ. 2024;384:e074820.
For which clinical domains, device functions, or Synthetic data may support targeted stress testing when clinically important
subpopulations is synthetic data particularly well- scenarios relevant to the device’s intended use are insufficiently represented in
suited, or particularly inadequate, as a supplement real-patient data, such as rare presentations or missing, conflicting, or atypical
inputs. Testing should address identified evidence gaps and be proportionate to
to real-world evidence? What safeguards would
device risk.
mitigate the risk that synthetic data generated by
models of the same class as the device under They may be less suitable for evaluating complex longitudinal trajectories,
evaluation reproduces the very performance gaps treatment-response relationships, rare serious adverse events, or behaviors and
healthcare interactions that are poorly captured in source data.
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
the evaluation is intended to detect, particularly Sponsors should prespecify the purpose and estimand, disclose data provenance
for underrepresented subgroups? and generation methods, and demonstrate relevance and reliability against realworld data. Validation should assess clinically important relationships rather
than only the plausibility of individual cases or similarity of marginal
distributions [1].
When AI generates evaluation data, the generator should, where feasible, be
independent of the evaluated device and sufficiently distinct from its underlying
model to reduce correlated failures. Evaluation should include real-patient
anchor cases, particularly for underrepresented subgroups. Synthetic data may
help identify vulnerabilities but should not substitute for real-patient evidence
supporting subgroup safety or performance claims. Limited representation
should prompt additional real-world data collection and postmarket
surveillance.
References
1. Yan C, Yan Y, Wan Z, Zhang Z, Omberg L, Guinney J, Mooney SD, Malin
BA. A multifaceted benchmarking of synthetic electronic health record
generation models. Nature Communications. 2022;13(1):7609.
doi:10.1038/s41467-022-35295-1.
For open-ended device outputs, how should ISPE recommends that CDRH align performance comparators and acceptance
performance comparators and acceptance criteria criteria with the device’s intended use, clinical risks, and conditions of use.
be selected? When a panel of qualified clinicians Comparators may include a panel of qualified clinicians or an appropriate
reference standard; established objective performance criteria (OPCs) or
serves as the comparator, how should the
justified performance goals (PGs) may provide acceptance targets where
applicable standard (for example, the standard of applicable. For open-ended outputs, sponsors should prespecify how clinical
care versus the performance of a median clinician correctness, completeness, and potentially harmful errors will be assessed.
in practice) be defined and justified? Should
generalist or specialist physicians be used as a
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
performance standard? When should human-AI
team performance, rather than the device
operating alone, serve as the basis for evaluation? When qualified clinicians serve as the comparator, sponsors should define and
justify whether the applicable standard represents the standard of care or the
performance of a median clinician in practice. Generalist or specialist
physicians should be selected according to the clinical task and expertise
required. Human–AI team performance should serve as the basis for evaluation
when the intended use depends on clinician interaction with the device, with
standalone device performance assessed as appropriate to characterize errors
and reliance on clinician oversight.
Fair comparison requires holding case-mix constant (paired designs, or
standardization as in provider profiling), with attention to overlap (positivity)
and a specified device version (consistency), given GenAI stochasticity.
Context-specific rubrics should be developed and validated prior to
deployment. CDRH may suggest mixed-methods approaches incorporating
both quantitative and qualitative techniques. [1]
References:
1. Hong C, Chowdhury A, Sorrentino AD, et al. Application of unified
health large language model evaluation framework to In-Basket
message replies: bridging qualitative and quantitative assessments.
Journal of the American Medical Informatics Association.
2025;32(4):626-637. doi:10.1093/jamia/ocaf023.
Are there ways in which performance might be There are several approaches to assess GenAI-enabled device performance
assessed relative to the care, technology, or course relative to the care, technology, and course of action likely to occur in the
of action likely to occur in the absence of the absence of the device, rather than the comparators described:
- Compare to an exchangeable real-world patient population in terms of
device, rather than to the comparators described in
the impact on the subsequent patient journey (e.g. medications and
this section? What approaches might be used to procedures the patient receives as a result of the GenAI-enabled device)
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
identify and justify a comparator such as unaided and separately the impact on clinical and safety endpoints relevant to
clinical judgment, delayed specialist review, or no the intended use population.
intervention?
When considering a comparator such as unaided clinical judgement, delayed
specialist review, or no intervention, it will be important to concretely specify
the counterfactual for each approach. A well specified counterfactual (e.g. the
patient outcomes if the patient is referred to a specialist 3 months later than they
would having used the GenAI-enabled device) is critical to a valid and specific
design of a non-active comparator of interest, of which each of these approaches
would classify. Understanding the clinical and demographic characteristics, as
well as standard of care treatment patterns, of these potential counterfactual
populations is important when considering the degree of exchangeability
between the population receiving the GenAI-enabled device (as an intervention
or under routine clinical care), particularly when outside the realm of a
randomized controlled clinical trial.
CDRH is considering whether it may be Greater reliance on postmarket monitoring may be appropriate when residual
appropriate to accept greater premarket premarket uncertainty can be reliably detected, measured, and mitigated
uncertainty regarding a GenAI-enabled device’s through a robust, prospective, risk-based surveillance program. Such an
approach builds on a premarket performance baseline appropriate to the
benefit-risk profile through greater reliance on
device’s intended use, risk, and applicable regulatory requirements and devicepostmarket monitoring. Under what conditions specific guidance, against which real-world performance can be assessed over
might such an approach be appropriate, and what time. The monitoring program should evaluate safety, effectiveness, reliability,
characteristics of a monitoring program would input-data quality, and performance across relevant patient and clinical
need to be in place to justify reduced premarket subgroups, while considering the device, sensing or data collection components,
evidence? Are there device types or risk profiles and AI algorithm separately.
for which this approach would not be appropriate?
Greater reliance on postmarket evidence should reflect both device activity and
the consequences of an incorrect output, as well as risk-based device
classification; informational status alone should not justify reduced premarket
evidence. For patient-facing functions, real-time logging should be paired with
timely detection and intervention, since recording an unsafe output does not
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
prevent harm. Findings should trigger reassessment of the device’s risk profile
and safeguards.
Monitoring should combine automated, continuous surveillance with
appropriately targeted human expert review and should include predefined
performance thresholds, triggers, and response procedures. The balance
between human review and automated monitoring of devices post-deployment
should reflect the device’s intended use and risk, with lower-risk systems
relying more on automated monitoring with periodic human audits, and higherrisk devices using a hybrid approach combining continuous automated
surveillance with expert review. Regular benchmarking against established
reference datasets or appropriate validated control populations can help identify
changes in calibration or performance. Monitoring should also account for
changes in clinical practice, patient populations, workflows, data sources, and
device components. [1]
Reducing premarket evidence in reliance on postmarket monitoring may be less
appropriate when device errors could cause serious or irreversible harm before
detection and mitigation, even where postmarket surveillance is warranted or
required. It may also be inappropriate where clinically meaningful outcomes or
reference standards cannot be reliably measured in routine practice. In such
cases, robust premarket evidence should remain an important component of the
overall benefit-risk assessment.
References
1. U.S. Food and Drug Administration. Artificial Intelligence-Enabled Device
Software Functions: Lifecycle Management and Marketing Submission
Recommendations. Draft Guidance for Industry and Food and Drug
Administration Staff. January 7, 2025. Section XI: Device Performance
Monitoring.
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
Please comment on the potential approaches to
postmarket performance evaluation, including ISPE supports all three approaches as complementary components of a riskbased postmarket evaluation strategy but suggests CDRH to frame postmarket
periodic re-benchmarking, sample-based clinician
monitoring as epidemiologic surveillance, not periodic re-examination. This
review, and performance degradation monitoring. perspective is consistent with recent calls for an epidemiology of AI that
What additional approaches should CDRH systematically measures AI exposure and its population-level health effects
consider, and how should the cadence and [1,2]. Applied to GenAI-enabled devices, this supports monitoring real-world
triggering events for reassessment be determined? use and associated clinical outcomes across populations and over time [3]. Each
approach is a measurement instrument whose own validity must be established.
Periodic re-benchmarking can determine whether performance remains
consistent with the established premarket baseline, particularly when changes
occur in sensors, data sources, patient populations, clinical practice, or the
context of use. While periodic re-benchmarking can be useful for detecting
changes in model performance, its value depends heavily on the quality and
relevance of the benchmark. A change in benchmark performance is most
informative when the benchmark has demonstrated construct validity for the
clinical risks of interest, such as using an established reference dataset or
validated control populations. [4] Public benchmarks may also become less
informative over time as models are increasingly exposed to them during
development or as performance approaches a ceiling. Re-benchmarking may
therefore be most useful as a regression check against a prespecified baseline
rather than as evidence, on its own, of continued performance in clinical
practice. Re-benchmarking may also be useful when the GenAI-enabled device
under routine care evolves and is no longer aligned with the initial test
population. Where feasible, evaluation should rely on sequestered benchmarks
with procedures to assess and limit parametric and non-parametric
contamination.
Sample-based clinician review offers a proximal assessment of performance in
realistic clinical scenarios. However, the human reference standard introduces
its own sources of uncertainty. [5] Expert reviewers may disagree when
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
evaluating open-ended clinical outputs, especially for complex or ambiguous
cases. Monitoring programs utilizing clinician review should pre-specify how
cases and reviewers are sampled, use validated evaluation criteria, calibrate
reviewer responses when possible, and report reliability metrics.
Performance-degradation monitoring presents additional challenges for
generative AI because many open-ended outputs do not have readily available
ground-truth labels. Conventional approaches to drift detection may not reliably
indicate changes in clinical performance.[6] Therefore, monitoring plans should
distinguish meaningful drift, including changes to the input population and
measurement between inputs and outputs.
Automated performance monitoring should provide continuous or frequent
surveillance for changes in input distributions, output performance, error rates,
adverse events, and other relevant indicators. Additional monitoring measures
can include characterization of the patient population using the devices.
continued Sample-based clinician review can then provide contextual
assessment of flagged cases, rare events, and clinically meaningful changes that
automated metrics may not capture.
Monitoring should not rely solely on fixed calendar intervals. Rather, it should
combine continuous operational monitoring with more in-depth evaluation
when triggers occur. These triggers might include sponsor-initiated updates,
changes to an underlying model or sensor, shifts in the data environment,
clinical guidelines, or practice, accumulating safety concerns, data drift,
subgroup-specific performance changes, or evidence that the device is being
used outside its evaluated context. Additionally, sponsors must recognize
changes to the users of the tools and their environments when considering
measurement validity behind the monitoring program.
References
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
1. Parikh H, McCormick T, Johnson EK, et al. The epidemiology of artificial
intelligence. Nature Health. Published online September 15, 2026.
doi:10.1038/s44360-026-00188-w.
2. Embi PJ. Algorithmovigilance—advancing methods to analyze and monitor
artificial intelligence–driven health care for effectiveness and equity. JAMA
Network Open. 2021;4(4):e214622.
doi:10.1001/jamanetworkopen.2021.4622.
3. Feng J, Xia F, Singh K, et al. Not all clinical AI monitoring systems are
created equal: review and recommendations. NEJM AI. 2025;2(2).
4. Bean AM, Kearns RO, Romanou A, et al. Measuring what matters:
construct validity in large language model benchmarks. Advances in Neural
Information Processing Systems. NeurIPS. 2026.
https://neurips.cc/virtual/2025/loc/san-diego/poster/121477
5. Tam TYC, Sivarajkumar S, Kapoor S, et al. A framework for human
evaluation of large language models in healthcare derived from literature
review. npj Digital Medicine. 2024;7(1):258. doi:10.1038/s41746-02401258-7.
6. Finlayson SG, Subbaswamy A, Singh K, et al. The clinician and dataset
shift in artificial intelligence. New England Journal of Medicine.
2021;385(3):283-286. doi:10.1056/NEJMc2104626.
What roles might clinicians, healthcare Postmarket monitoring should draw on stakeholder expertise. Clinicians and
institutions, professional societies, standards- healthcare institutions can identify clinically meaningful errors, unexpected use,
setting bodies, and other stakeholders and workflow effects, and link device use to longitudinal real-world data to
assess downstream actions and outcomes. Professional societies such as ISPE
appropriately play in postmarket monitoring, and
can potentially develop common definitions, outcome measures, and evaluation
how can these roles be structured without methods. Standards bodies can support interoperability, device-version
diffusing manufacturer accountability? documentation, and consistent data capture.
Monitoring should integrate into routine workflows and capture which device
and version was used, the context of use, and whether outputs were presented
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
and acted upon. These contributions should complement manufacturer
responsibilities: manufacturers should remain accountable for maintaining
monitoring systems, evaluating and investigating signals, and implementing
corrective or risk-mitigation actions.
Other areas of requested feedback
Other Areas of Requested Feedback ISPE Response
Section V.C.1, Clinical Confirmation Approaches (p. Methodologically rigorous studies using real-world data collected
15) — real-world data and data collected outside the outside the United States should be considered for clinical
United States confirmation when the data are fit for purpose (e.g. both relevant and
reliable) and findings are applicable to the intended U.S. population
CDRH recognizes these approaches may incorporate and clinical setting. Multi-database networks and outcome validation
real-world data (RWD) and/or data collected outside the studies highlight the importance of addressing differences in data
United States (OUS) and seeks input on specific sources, coding practices, populations, and measurement validity
considerations for the use of such data in the clinical across settings [1,2].
confirmation of GenAI-enabled devices.
Transportability (the extent to which findings apply to other
populations and clinical settings) must consider differences in clinical
workflows, professional roles and their designated responsibilities,
treatment guidelines, language, documentation practices, healthcare
access and patterns of interaction with the device, because these may
modify device performance even when the underlying patient
populations are clinically similar. Applicability of OUS evidence
should therefore consider both transportability of the patient
population and transportability of the deployment context.
References
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
1. Haber R, Webster-Clark M, Pratt N, et al. Core Concepts in
Pharmacoepidemiology: Multi-Database Distributed Data Networks.
Pharmacoepidemiol Drug Saf. 2025;34(7):e70177.
doi:10.1002/pds.70177.
2. Weinstein EJ, Ritchey ME, Lo Re V. Core concepts in
pharmacoepidemiology: Validation of health outcomes of interest
within real-world healthcare databases. Pharmacoepidemiol Drug Saf.
2023;32(1):1–8. doi:10.1002/pds.5537.
Section V.C.1, Clinical Confirmation Approaches (p. Synthetic data may supplement evaluation of small or hard-to-sample
16) — synthetic data as a supplement for subgroup subgroups, but particular caution is needed when limited real-world
evaluation data motivate their use. Their representativeness depends on the
information available to generate and validate them. Sponsors should
Where subgroup evaluation would be difficult, demonstrate that synthetic data preserve clinically important
particularly for small or hard-to-sample subgroups where relationships among patient characteristics, treatments, outcomes, and
real-world data is limited, CDRH is considering that healthcare context, rather than relying only on plausible individual
synthetically generated data modeled on real patient data cases or similar marginal distributions [1].
could potentially supplement evaluation, assuming such
synthetic data is representative of the environment in For underrepresented subgroups, synthetic data may support targeted
which the device is intended to be deployed. challenge testing and identify performance vulnerabilities, but should
not substitute for real-patient evidence supporting subgroup-specific
safety claims. Remaining uncertainty should be acknowledged and
may warrant targeted real-world data collection and enhanced
postmarket monitoring. A lack of adequate real-world data remains an
evidence gap that synthetic data do not necessarily resolve.
References
1. Yan C, Yan Y, Wan Z, Zhang Z, Omberg L, Guinney J, Mooney
SD, Malin BA. A multifaceted benchmarking of synthetic electronic
health record generation models. Nature Communications.
2022;13(1):7609. doi:10.1038/s41467-022-35295-1.
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
Section V.C.1, Clinical Confirmation Approaches (p. Endpoints and sample sizes should reflect the device’s intended
16) — endpoints and sample size clinical use, the purpose of clinical confirmation, and the consequences
of error. Alongside technical and clinical performance, evaluation
CDRH acknowledges that the approaches outlined above should consider downstream effects on clinical decisions, time to
may not be powered around traditional effectiveness appropriate care, healthcare utilization, and patient outcomes. Sampleendpoints, and is interested in stakeholder input on how size planning should address precision around clinically consequential
endpoints and sample sizes might be determined. error rates, including uncommon serious errors that aggregate accuracy
may obscure [1].
Rare, high-consequence scenarios may warrant targeted or enriched
evaluation, with results reported separately from routine-use
performance estimates. Clinically important subgroups and the
precision needed to support intended-use claims should be
prespecified. Postmarket real-world evidence may complement
premarket evaluation for outcomes or populations that cannot feasibly
be assessed with sufficient precision before deployment.
References
1. Riley et al. Evaluation of clinical prediction models (part 3):
calculating the sample size required for an external validation study.
BMJ. 2024;384:e074821.
Section VI.A, Postmarket Monitoring Approaches (p.
20) — feasibility and practicality of implementation ISPE recommends that CDRH consider implementation feasibility
using an established framework such as the Nonadoption,
CDRH seeks feedback on these approaches, as well as the Abandonment, Scale-up, Spread, and Sustainability (NASSS)
feasibility and practicality of implementing these or other framework [1]. NASSS helps explain how a technology, its users,
approaches to postmarket monitoring for GenAI-enabled healthcare organizations, and the wider health system interact to
devices. support or hinder successful implementation over time. Applying such
a framework could help identify where monitoring strategies are most
likely to encounter practical barriers over time. The success of
monitoring depends on clinical workflow, incentives, data access, and
ISPE Comments on FDA CDRH Discussion paper: Considerations for the Regulation of Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback
Docket No. FDA-2026-N-7874 Submitted: October 14, 2026
organizational capacity, in addition to technical measurement
strategies.
ISPE therefore recommends that CDRH consider two implementation
principles. First, monitoring plans should address the practical
requirements for implementation, including data access, workflow
integration, staffing, and heterogeneity across deployment settings, in
addition to specifying performance metrics. Second, guidance should
acknowledge the role of health systems, professional societies,
standards organizations, and other ecosystem partners that may
contribute data, expertise, or adjudication while maintaining clear
manufacturer accountability.