Themes it raises

12 of the 21 themes in the docket, each with the passage we counted, verbatim.
Escalating too little and too muchFDA Q6
“Under- and over-escalation should be characterized, measured, and controlled separately because their likelihood, clinical consequences, and operational impacts may differ.”
Whether benchmark results prove anythingFDA Q9, Q10, Q16
“For a benchmark on given data - externally sourced or internally - to be reliable, the underlying data and its limitations, strengths, and weaknesses need to be understood to a sufficient extent.”
Proving the device works in real careFDA Q11, Q12, Q13, Q14, Q15
“ISPE believes that prospective clinical studies should not be universally required for a confirmation approach.”
Trading premarket certainty for postmarket monitoringFDA Q18
“ISPE supports an approach that accepts greater premarket uncertainty regarding a GenAI-enabled device's benefit-risk profile by placing greater reliance on postmarket monitoring.”
Watching the device after it shipsFDA Q19, Q20
“The performance of such agents should be verified themselves, as they form part of the control strategy. As a matter of scalability, a portion of instances may be retained for (additional) human review to allow for a monitoring loop for the supervisory agent itself.”
Controlling a device that keeps changingFDA Q22, Q23, Q24, Q25
“Where these controls do not provide sufficient notification, traceability and assurance for the intended use and risk, the model or supplier should be considered unsuitable.”
Whether human oversight is real oversightFDA Q3, Q4, Q14, Q20, Q21, Q26
“Human oversight may mitigate residual risk, but should receive credit only where its effectiveness has been demonstrated.”
How this fits rules that already existFDA Q8, Q9, Q16, Q25
“Voluntary measures, such as codes of conduct, are implemented under the EU AI Act (see references), while specific requirements are also imposed for certain general-purpose models. ISPE would welcome alignment with such considerations to leverage existing transparency efforts.”
What makes a function high riskFDA Q1, Q2, Q5
“CDRH may consider risk assessment outcomes that describe a range of modes within the informational vs. directional spectrum, while manufacturers would leverage these risk outcomes through targeted control strategies commensurate with the respective risk to the patient, which may change throughout a conversational trajectory.”
Whether the user can judge the outputFDA Q3, Q4
“The entire risk management exercise should consider the clinical action that may result from the output and whether the intended HCP has the competence, authorization, and relevant scope of practice to take that action.”
Judging devices the way clinicians are credentialedFDA Q7, Q8
“ISPE believes that the competency-based framework is an appropriate approach depending on the purpose of the GenAI system whilst recognizing the challenges of defining competencies.”
Who is accountable when something goes wrongFDA Q21
“These complementary roles should be clearly defined and should not transfer or dilute the manufacturer’s accountability for device performance and regulatory compliance.”

FDA questions it names

Questions this filing names by number.

Q1 · The two-axis risk frameworkQ2 · The spectrum of device activityQ3 · When an output becomes directiveQ4 · Generalist and specialist usersQ5 · Multi-turn conversations that migrateQ6 · Care escalation functionsQ7 · The competency-based approachQ8 · Mapping the risk grid to evidenceQ9 · The benchmarking structureQ10 · Benchmark contamination and saturationQ11 · Clinical confirmation without a prospective trialQ12 · Statistically meaningful performanceQ13 · Synthetic dataQ14 · Comparators and acceptance criteriaQ15 · Performance against usual careQ16 · Independent third partiesQ17 · Devices with many functionsQ18 · Trading premarket certainty for postmarket monitoringQ19 · Postmarket performance evaluationQ20 · Machine-based supervisory agentsQ21 · Clinicians, institutions and societiesQ22 · Re-benchmarking after a modificationQ23 · PCCPs for GenAI devicesQ24 · Third-party foundation model changesQ25 · Foundation Model Master FilesQ26 · Agentic devices

Machine-assisted draft, pending human review. The source text and highlighted passages appear below. Read the filing on regulations.gov ↗

The comment as filed

Comment submitted on regulations.gov. Passages we counted are highlighted.

Comments from The International Society for Pharmaceutical Engineering (ISPE)

Attachment

Attachment, text extracted from the filed document. Passages we counted are highlighted.

ISPE Comments on FDA-2026-N-7874, Considerations for the Regulation of
Generative AI-Enabled Medical Devices: Discussion Paper and Request for
Feedback

Submit Date: October 14, 2026

Dockets Management Staff (HFA-305), Food and Drug Administration, 5630 Fishers Lane,
Room 1061, Rockville, MD 20852

Re: Docket No. FDA-2026-N-7874, Considerations for the Regulation of Generative AIEnabled Medical Devices: Discussion Paper and Request for Feedback

To Whom It May Concern:

The International Society for Pharmaceutical Engineering (ISPE) appreciates the opportunity to
comment on the discussion paper titled "Considerations for the Regulation of Generative AIEnabled Medical Devices."

ISPE supports FDA's development of a risk-based framework for GenAI-enabled devices. Our
comments emphasize that evidence expectations should be proportionate to risk and context of
use, allowing sponsors to justify fit-for-purpose approaches rather than applying uniform
requirements.

Because GenAI is evolving rapidly, we also encourage a technology-agnostic, total-productlifecycle approach that connects premarket evaluation, postmarket monitoring, and change
control. Under such an approach, CDRH would specify expected outcomes rather than
particular tools or benchmarks, and manufacturers would remain accountable for device
performance.

ISPE appreciates FDA's consideration of these discussion points.

Sincerely,

Michael S. Martin, President and CEO
International Society for Pharmaceutical Engineering (ISPE)
6110 Executive Blvd., Suite 600, North Bethesda, MD 20852
Tel. +1 301-364-9201 | www.ispe.org

Enclosure: ISPE Discussion Paper Response

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 1 of 14 Of Pharma™
Responses to FDA Request for Comment: FDA-2026-N-7874, Considerations for
the Regulation of Generative AI-Enabled Medical Devices

IV. Risk Assessment

Question 1
FDA question: Does the two-axis risk framework, organized around device activity and the
consequence of relying on an incorrect output, appropriately capture the dimensions most relevant to
the risk of a GenAI-enabled software function? If there are additional dimensions—such as the
reversibility of a resulting action, the availability of downstream safeguards, the time pressure of the
deployment setting, or the traceability of the output (i.e., to primary source materials)—that should
be represented in a risk framework, please describe and provide examples of how they should be
represented.
Position summary
ISPE supports the two-axis framework, similarly to what was presented in FDA's draft guidance
Considerations for the Use of Artificial Intelligence to Support Regulatory Decision-Making for Drug and
Biological Products

Question 2
FDA question: CDRH seeks input on how to account for the spectrum of GenAI-enabled informational
functions that vary in the degree to which they direct a user to a particular action (i.e., between “nondirective” and “action-directing”). What characteristics of an output—such as its wording, specificity,
personalization, or context—could be considered as modifiers of the risk associated with an
informational function, after accounting for the device's overall functionality and intended use? What
additional information would provide manufacturers with sufficient clarity and predictability around
risk assessment for informational functions while recognizing that directiveness may exist along a
continuum rather than as a binary distinction?
Position summary
ISPE supports the position that there may be further considerations beyond signaling words to delineate
between informational and directional objectives.
Rationale/detail
Where wording, specificity, or the nature of the output materially affects directiveness or risk, these
characteristics should be identified during the initial risk assessment and addressed during risk
assessment and risk control steps in risk management exercise. Risk control recommendations should be
taken by a team of experts familiar with the situation and potentially tested in an evaluation exercise.
Recommendation to CDRH
While the delineation between informational and directional functions is helpful, CDRH may consider
emphasizing the de facto impact on the output from the decision-making process in the risk
management exercise, which may be influenced by signal words like "recommend.”

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 2 of 14 Of Pharma™
Question 3
FDA question: CDRH seeks input on whether and when a GenAI-enabled function that results in
delivery of clinical information to patients, as opposed to HCPs, could present different or higher risks,
while also recognizing the potential benefits associated with improved patient empowerment,
engagement, and access to clinical information. What device characteristics, output features, or
safeguards might mitigate risks that could arise when a user lacks the domain knowledge to
independently evaluate an output, without unnecessarily underestimating patient capability?
Position summary
ISPE believes that the level of literacy and the ability of critical thinking of the desired output should be
reflected in the device design, in concordance with the target population (users/patients) and
development, using risk management, i.e., the risk identification step.
Rationale/detail
It is well known, for example, in ICH Q9(R1) (Quality Risk Management), that subjectivity, e.g., bias, can
impact all stages of a risk management exercise.
Recommendation to CDRH
CDRH may consider mentioning that the risk management process used during development should be
designed to minimize bias during the exercise and in the resulting output.

Question 4
FDA question: CDRH seeks input on whether and how the distinction between generalist and
specialist physicians might inform the assessment of risk for HCP-facing GenAI-enabled functions.
Under what circumstances, if any, might risk be affected when an HCP who lacks the relevant clinical
specialist knowledge receives information that falls within a specialist area of practice? What device
characteristics or safeguards might mitigate such risks?
Position summary
ISPE believes that a distinction between estimated competencies of generalist and specialist physicians
should be factored into the risk management exercise.
Rationale/detail
The entire risk management exercise should consider the clinical action that may result from the output
and whether the intended HCP has the competence, authorization, and relevant scope of practice to
take that action.

Recommendation to CDRH
The CDRH may consider including discussion of factoring HCP competence into device development
recommendations.

Question 5
FDA question: For multi-turn conversational GenAI-enabled devices that may migrate from providing
“non-directive” information to “action-directing” information over the course of an exchange, how

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 3 of 14 Of Pharma™
should risk be assessed across realistic conversational trajectories? How could the intended use of
such a device be characterized when its behavior is emergent across a conversation?
Position summary

ISPE acknowledges the possibility that conversations may drift across the informational to directional
spectrum. The intended operational boundaries and context of use should be clearly defined, and the
system should explicitly communicate how its output is intended to be interpreted. The associated risk
management process should support these boundaries through risk-proportionate test coverage and
appropriate control measures.

Rationale/detail
An adaptive approach would accommodate both a) the nature of conversational trajectories and b)
sufficient rigor when more directional portions of the spectrum are proposed for use.
Recommendation to CDRH
CDRH may consider risk assessment outcomes that describe a range of modes within the informational
vs. directional spectrum, while manufacturers would leverage these risk outcomes through targeted
control strategies commensurate with the respective risk to the patient, which may change throughout
a conversational trajectory.

Question 6
FDA question: For GenAI-enabled care escalation functions, CDRH is considering how the evaluation
may account for both under-escalation and over-escalation. How could manufacturers characterize
and weigh these two directions of error, given that they may not be commensurable and that
acceptable trade-offs may vary by clinical context?
Position summary
ISPE believes that under- and over-escalation should be evaluated, controlled, and monitored in the risk
management exercise because their clinical consequences and operational impacts may differ. The
appropriate trade-off and control strategy should reflect the criticality of the condition, the performance
and error modes of the existing care pathway, and the demonstrated effectiveness of human oversight.
Users should be able to escalate or override outputs based on additional evidence and clinical judgment,
subject to controls proportionate to the resulting risk.
Rationale/detail
Under- and over-escalation should be characterized, measured, and controlled separately because their
likelihood, clinical consequences, and operational impacts may differ.
The control strategy should be
informed by the performance and error modes of the existing care pathway and should reflect the
relative severity of each potential error in the intended context. Human oversight may mitigate residual
risk, but should receive credit only where its effectiveness has been demonstrated.
Users should be able
to escalate or override outputs based on additional evidence and clinical judgment, subject to controls
proportionate to the resulting risk.
Recommendation to CDRH
CDRH may consider providing guidance that over- and under-escalation should be treated separately in
risk management exercises.

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 4 of 14 Of Pharma™
V. Premarket Evaluation

Question 7
FDA question: Is the competency-based approach described above, i.e., device benchmarking
followed by clinical confirmation, a useful and appropriate framework for evaluating GenAI-enabled
devices?
Position summary
ISPE believes that the competency-based framework is an appropriate approach depending on the
purpose of the GenAI system whilst recognizing the challenges of defining competencies.

Question 8
FDA question: How might the two-axis risk framework described in Section IV be considered within
the competency-based approach described above to help determine the level of evidence needed for
a premarket submission?
Position summary
ISPE supports using risk management to determine the nature and rigor of evidence required, which
should include a proposed, justified control strategy in a pre-market submission. Higher-risk, actiondirecting or autonomous functions should receive greater scrutiny of robustness, safety-critical
recognition and escalation, and adherence to intended-use boundaries.
Rationale/detail
ISPE considers the two-axis framework a useful starting point for the device development risk
management exercise.
Recommendation to CDRH
CDRH should present the two-axis framework as a useful starting point for the risk management
exercise. The guidance could consider including all important factors received as part of this
consultation exercise.

Question 9
FDA question: Would a benchmarking structure such as the one described above be likely to provide
adequate evidence of clinical knowledge, analytic capabilities, safety behavior, communication, and
generalizability to support a reasonable assurance of safety and effectiveness? Are there elements
that are missing, redundant, or inappropriately categorized? Are there externally developed
standards that could be leveraged?
Position summary
ISPE generally supports describing a benchmarking structure; however, care should be taken to allow for
changing technology.
Rationale/detail
Since technology is rapidly evolving, the competencies applicable to Agentic AI to ensure acceptable
performance may change.

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 5 of 14 Of Pharma™
Recommendation to CDRH
CDRH may consider a technology-agnostic approach in benchmarking to cover a broad range of
changing technologies.

Question 10
FDA question: CDRH recognizes that publicly available benchmarking assets may be subject to data
contamination, saturation, and limited real-world representativeness. How should a sponsor establish
that performance on a given benchmark predicts safe and effective real-world behavior for the
device's intended use? What evidence should support the construct validity of a benchmark used to
gate device evaluation, and what role should sponsor-developed benchmarks play given potential
concerns around independence and optimization to the test?
Position summary
ISPE shares FDA's view on the limitations of publicly available data. For a benchmark on given data -
externally sourced or internally - to be reliable, the underlying data and its limitations, strengths, and
weaknesses need to be understood to a sufficient extent.
In the event of limitations, further sponsordeveloped benchmarks may be needed, leveraging targeted data in the context of use of the AI-enabled
device where evidence is currently uncertain.
Rationale/detail
ISPE has developed a fitness for purpose concept, comprising four properties
• Reliability,
• relevance
• representativeness
• abundance
These could serve as an orientation for assessing and, where appropriate, producing additional data for
benchmarking.
Recommendation to CDRH
CDRH may consider requiring that all benchmarks, whether publicly available or sponsor-developed, are
fully justified in relation to the overall risk characterization, perhaps specifically mentioning the need to
consider independence and optimization.

Question 11
FDA question: CDRH is considering that clinical confirmation for a GenAI-enabled device might not
require a prospective clinical study in every case, and has described above a range of approaches of
increasing rigor and patient exposure. How might a sponsor select and justify a confirmation
approach tailored to a device's intended use and proportionate to the device's risk profile? Are there
device types or risk profiles for which one or more of these approaches would be insufficient or
inappropriate? How might the anticipated distribution of real-world inputs be taken into
consideration? Are there other methods of clinical confirmation that might help inform the evaluation
of GenAI-enabled devices?
Position summary
ISPE believes that prospective clinical studies should not be universally required for a confirmation
approach.
The confirmation approach, clinical or non-clinical, should be proportionate to the device’s

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 6 of 14 Of Pharma™
risk profile, with prospective evidence required to provide sufficient assurance of safety and
effectiveness.
Rationale/detail
Selection should consider, for example, system novelty and autonomy; consequences, detectability and
reversibility of error; quality and representativeness of existing evidence; anticipated real-world inputs;
and the demonstrated effectiveness of proposed risk controls and human oversight.
Recommendation to CDRH
CDRH should permit sponsors to select and justify a fit-for-purpose confirmation approach with
appropriate justification.

Question 12
FDA question: How should sponsors achieve statistically meaningful performance measurement for
GenAI-enabled devices? Where synthetically generated inputs supplement real patient data, how
should sponsors account for differences between the synthetic and real-world distributions when
estimating performance, and under what conditions, if any, is it appropriate to combine
benchmarking evidence and clinical confirmation evidence to support a single performance estimate?
Position summary
ISPE believes that statistically meaningful performance will depend on several factors, and the suitability
of the data used should be justified accordingly. For instance, the amount of data required to derive
statistically meaningful conclusions may depend on the complexity of the model, the scope of the task,
and the heterogeneity of data subgroups, each of which may require evidence. When real-world data is
scarce or may not fully capture the range of expected real-world scenarios, synthetic data may be used
with justification. However, transparency regarding the nature and proportions of the data, as well as
the performance of the synthetic data-generating approach, is necessary. Full reliance on synthetic data
should require extended reasoning on suitability of the chosen approach in relation to risk.
Rationale/detail
Regarding use of data, and data expectations, trade-offs and compromises may be required. Taking
account of risk, various factors such as facilitating data sharing may be supportive measures, as well as
use of synthetic data where its adequacy for the device's context can be demonstrated.
Recommendation to CDRH
CDRH may consider a variety of approaches to establish a robust data foundation for deriving a
performance estimate, while considering a range of factors that should be considered to satisfy data
quality needs; for instance, decision consequence, prevalence of the condition, complexity of the AI
approach, heterogeneity of the device's anticipated performance across subgroups and sub-populations,
etc.

Question 13
FDA question: For which clinical domains, device functions, or subpopulations is synthetic data
particularly well-suited, or particularly inadequate, as a supplement to real-world evidence? What
safeguards would mitigate the risk that synthetic data generated by models of the same class as the

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 7 of 14 Of Pharma™
device under evaluation reproduces the very performance gaps the evaluation is intended to detect,
particularly for underrepresented subgroups?
Position summary
ISPE believes that the use or restriction of synthetic data should be determined by its fitness for the
intended purpose and the level of risk, including the extent and significance of gaps in available realworld data. Its purpose, generation process, and limitations should be specified and assessed, and its
relevance, reliability, and representativeness should be demonstrated against independent real-world
data, particularly for relevant populations, conditions, and failure modes.
Rationale/detail
ISPE believes that the use or restriction of synthetic data should not be determined solely by clinical
domain or device classification. Its suitability should be assessed for the intended purpose and
proportionate to the risk, considering the limitations of available real-world data, the role and relative
contribution of synthetic data within the evidence package, and the extent to which its relevance,
representativeness and reliability can be demonstrated against independent real-world data. Synthetic
data should supplement, rather than obscure, evidence gaps and should not be used where its validity
for the relevant populations, conditions or failure modes cannot be adequately established.
Recommendation to CDRH
CDRH should avoid prescribing suitability for synthetic data by clinical domain or device classification
alone. Sponsors should justify its use based on intended purpose, risk and gaps in real-world data, and
demonstrate its relevance, reliability and representativeness against independent real-world evidence,
including for underrepresented subgroups and relevant failure modes. CDRH may consider the need for
transparency on the purpose of synthetic data generation, its generation process, and a structured
assessment of potential deficiencies or weaknesses, to derive the incremental value of such data
compared to other evidence sources.

Question 14
FDA question: For open-ended device outputs, how should performance comparators and acceptance
criteria be selected? When a panel of qualified clinicians serves as the comparator, how should the
applicable standard (for example, the standard of care versus the performance of a median clinician
in practice) be defined and justified? Should generalist or specialist physicians be used as a
performance standard? When should human-AI team performance, rather than the device operating
alone, serve as the basis for evaluation?
Position summary
ISPE believes that performance comparators and acceptance criteria for open-ended outputs should be
selected and justified according to the intended use, clinical context and level of risk. Evaluation
methods should be fit for purpose and shown to produce reliable, clinically meaningful and reproducible
assessments, while the appropriate comparator (including current standard of care, qualified clinicians,
alternative technologies or human-AI team performance) should reflect the decision being supported
and the reasonably available alternatives.
Rationale/detail
Where clinicians provide the comparator, their relevant specialty, competence, experience and clinical
responsibility should reflect the intended users and decisions supported by the device.

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 8 of 14 Of Pharma™
Recommendation to CDRH
CDRH may consider the need to recommend rationales to justify the chosen performance comparators,
acceptance criteria, and benchmarks, and to consider alternative options, as well as elevated
requirements in cases of higher complexity or risk (e.g., when aiming for a shift from HCP-supervised to
autonomous mode). All these considerations should be related to the context of use.

Question 15
FDA question: Are there ways in which performance might be assessed relative to the care,
technology, or course of action likely to occur in the absence of the device, rather than to the
comparators described in this section? What approaches might be used to identify and justify a
comparator such as unaided clinical judgment, delayed specialist review, or no intervention?
Position summary
ISPE considers that this question should be answered by those with significant clinical expertise and
experience in the use of devices, especially devices which include some form of model.

Question 16
FDA question: Should independent third parties be involved in some or all aspects of a competencybased approach, including device benchmarking and clinical confirmation? If so, in what ways might
qualified, independent third-party participation contribute to this approach, and what qualifications
and independence criteria should apply? Are there aspects of the assessment for which third-party
involvement would be impractical or inadvisable? What safeguards or program design features
would be critical to prevent third-party participation from limiting competition or preventing
innovation?
Position summary
Third-party evaluation should be available as a risk-based assurance mechanism but should not be
universally mandated. Its value depends on the system’s risk, novelty, and complexity, as well as the
independence or specialist expertise required for a credible assessment.
Rationale/detail
A secondary, independent AI system could support evaluation, for example, by reviewing outputs at
scale, detecting inconsistencies or identifying cases requiring expert review. However, it should not
automatically be treated as an independent third party. Its independence would depend on separate
model provenance, training data, architecture, provider, and incentives.
Recommendation to CDRH
CDRH should avoid requiring third-party evaluation in every case. CDRH should permit risk-based
justification of evaluation approaches considering the context of use. CDRH should preserve
manufacturer accountability regardless of who performs the assessment.

Question 17
FDA question: CDRH is exploring whether a competency-based approach could be applied to devices
that incorporate a variety of underlying model architectures, such as multimodal vision-language
models and generative or predictive world models. Are there aspects of the described approach that

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 9 of 14 Of Pharma™
might be ineffective or inapplicable to these devices? What additional or distinct considerations
might need to be addressed?
Position summary
ISPE supports an approach, as largely presented in the document, that is technology-agnostic and should
avoid specifying benchmarks relying on the clinical context and the technology choice.
Rationale/detail
For example, scope maintenance and boundary adherence may be defined for various modalities of
input, demonstrated across a suitable set of data and/or scenarios, while applying to the general logic of
this benchmark category.
Recommendation to CDRH
CDRH may consider a competency-based approach that is largely technology-agnostic, depends on
technology choice and context of use, and is underpinned by an adequate risk-based rationale.

VI. Postmarket Monitoring

Question 18
FDA question: CDRH is considering whether it may be appropriate to accept greater premarket
uncertainty regarding a GenAI-enabled device's benefit-risk profile through greater reliance on
postmarket monitoring. Under what conditions might such an approach be appropriate, and what
characteristics of a monitoring program would need to be in place to justify reduced premarket
evidence? Are there device types or risk profiles for which this approach would not be appropriate?
Position summary
ISPE supports an approach that accepts greater premarket uncertainty regarding a GenAI-enabled
device's benefit-risk profile by placing greater reliance on postmarket monitoring.
The design of the
postmarketing monitoring program should be based on the estimated residual risk from the premarket
development program, particularly when residual risk reduction would be most effective through a
carefully designed postmarketing evaluation program.
Rationale/detail
Allowing more premarket uncertainty may lead to faster development with the benefit of devices being
available to patients earlier. Use of a postmarket monitoring approach could in some cases allow more
real patient data to become available and be used to increase the assurance of an evaluation of the
performance of the GenAI-enabled device.
Recommendation to CDRH
While CDRH may generally link expectations on acceptable premarket uncertainty and appropriate
depth of monitoring activities to the level of risk and the clinical context, certain circumstances, such as
a lack of alternative options and scarcity of data, may admit for higher premarket uncertainty compared
to settings in which more treatment options or data are available. Where greater premarket uncertainty
is accepted, sponsors should clearly characterize and document the remaining uncertainties and
associated risks, communicate them appropriately to users, and explain the postmarket measures and
actions intended to address them.

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 10 of 14 Of Pharma™
Question 19
FDA question: Please comment on the potential approaches to postmarket performance evaluation,
including periodic re-benchmarking, sample-based clinician review, and performance degradation
monitoring. What additional approaches should CDRH consider, and how should the cadence and
triggering events for reassessment be determined?
Position summary
Postmarket performance evaluation should build upon the premarket evaluation framework, using
consistent and comparable measures where appropriate as discussed in the answer to Question 18. A
coherent lifecycle approach supports detection of performance changes over time while improving the
efficiency of evaluation design and implementation.
Rationale/detail
Further coverage of relevant elements (data, adaptive behavior) may allow for greater control and
earlier detection of deficiencies, rather than the lagging measure of device performance. Continuity with
premarket evaluation improves comparability over time and enables an efficient, coherent lifecycle
approach.
Recommendation to CDRH
CDRH may consider additional dimensions that may be relevant for postmarket performance, including
data and adaptive behavior. CDRH should encourage postmarket evaluation frameworks that build on
premarket methods and measures.

Question 20
FDA question: Please comment on whether the proposed approaches to postmarket monitoring can
be facilitated by machine-based supervisory agents. What considerations, including the evaluation
and reliability of the supervisory agent itself, should CDRH take into account for such an approach?
Position summary
ISPE welcomes the possibility to employ machine-based supervisory agents for postmarket monitoring
purposes. The performance of such agents should be verified themselves, as they form part of the
control strategy. As a matter of scalability, a portion of instances may be retained for (additional) human
review to allow for a monitoring loop for the supervisory agent itself.

Rationale/detail
With the supervisory agent becoming part of the control system and simultaneously exhibiting model
characteristics, such functionality would be subject to a control strategy in itself.
Recommendation to CDRH
CDRH may consider the use of machine-based supervisory agents with appropriate demonstration of
their suitability and adequate performance monitoring in place.

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 11 of 14 Of Pharma™
Question 21
FDA question: What roles might clinicians, healthcare institutions, professional societies, standardssetting bodies, and other stakeholders appropriately play in postmarket monitoring, and how can
these roles be structured without diffusing manufacturer accountability?
Position summary
ISPE believes that postmarket monitoring can be supported through user education, incident reporting,
clinical review, performance-data sharing, and the identification of emerging risks. These
complementary roles should be clearly defined and should not transfer or dilute the manufacturer’s
accountability for device performance and regulatory compliance.

Rationale/detail
Training, clear documentation, and accessible user support can help users evaluate outputs, recognize
limitations, and report potential deficiencies. Stakeholder feedback should be collected through defined
channels, assessed systematically, and incorporated into the manufacturer’s risk management,
corrective action, preventive action, and change management systems.
Recommendation to CDRH
CDRH should encourage defined stakeholder roles, reporting pathways, and information-sharing
arrangements while maintaining clear manufacturer ownership of monitoring, investigation, and
corrective action.

Question 22
FDA question: CDRH envisions that a manufacturer's premarket competency-based assessment might
serve as a baseline against which post-deployment modifications could be re-evaluated. How might
the extent of re-benchmarking or other evidence be scaled to the nature and expected impact of a
given modification? For example, are there categories of change that might not significantly affect
safety or effectiveness of a GenAI-enabled device, or might be appropriately managed within a
sponsor's quality management system versus requiring FDA premarket review and authorization
before implementation? Are there categories of changes that might be appropriate for inclusion in a
PCCP?
Position summary
ISPE generally supports the possibility of including modifications to GenAI-enabled medical devices in
PCCP. However, even seemingly routine changes, such as prompt alterations, may cause unintended
performance shifts, and establishing sufficient evidence for such changes may be challenging. Such
changes should therefore be supported by rigorous verification against predefined benchmark
expectations and acceptance thresholds, including, where relevant, measures of performance shifts
between model versions.
Recommendation to CDRH
CDRH may consider inclusion of modifications to AI-enabled medical devices in a PCCP, in due
consideration of effective assurance of performance of such changes, whether carried out manually or
automatically.

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 12 of 14 Of Pharma™
Question 23
FDA question: How might PCCP concepts or other change-control approaches be adapted for GenAIenabled devices when the nature or scope of future modifications cannot be fully prespecified?
Position summary
ISPE believes that sponsors should maintain effective configuration and change control across all device
components, consistent with established practices for externally sourced software.
Rationale/detail
Controls should provide traceability of component versions, dependencies and changes, together with
impact assessment, predefined acceptance criteria, regression testing and rollback capability. Sponsors
should use effective and current configuration-management methods appropriate to the device, its
architecture and level of risk.
Recommendation to CDRH
CDRH should specify expected control outcomes without prescribing configuration-management tools
or technologies.

Question 24
FDA question: For devices built on third-party foundation models, changes to the underlying model
may be initiated by the third-party model developer rather than the device manufacturer. How can a
manufacturer detect, evaluate, and respond to such changes in a timely manner? What
mechanisms—for example, contractual, technical, or through a PCCP—could provide reasonable
assurance that third-party developer-initiated changes do not compromise the safety or effectiveness
of the device?
Position summary
Generally, notifications of changes should be expected, similar to changes performed by cloud
providers. In addition, technical measures such as periodic re-benchmarking against a fixed reference
set may unveil potential shifts in performance in the context of use, while considering potential
probabilistic behavior of models. As described above, a PCCP with performance controls may provide
the basis to efficiently accommodate such changes.
Rationale/detail
The sponsor’s model selection should consider the third-party provider’s configuration and change
management capabilities. Where these controls do not provide sufficient notification, traceability and
assurance for the intended use and risk, the model or supplier should be considered unsuitable.

Recommendation to CDRH
CDRH may specifically consider third-party initiated changes such as those to foundation models as
applicable within the PCCP framework under appropriate control conditions, while considering technical
means such as periodic re-benchmarking. CDRH should expect sponsors to assess third-party providers’
configuration and change-management capabilities as part of supplier qualification and model selection.

6110 Executive Blvd., Suite 600, North Bethesda, MD 20852 USA Shape
T 1 301-364-9201 F 1 240-204-6024 ispe.org The Future
Page 13 of 14 Of Pharma™
VII. Other Topics

Question 25
FDA question: Would voluntary Foundation Model MAFs be practical for foundation model
developers to provide and to keep current, and what content might they include? Given that
participation would be voluntary and that model developers may have limited incentive to disclose
safety-relevant information, what would make such a program sufficiently useful for premarket
review? Are there alternative mechanisms CDRH should consider for obtaining information about
underlying foundation models?
Position summary
Voluntary measures, such as codes of conduct, are implemented under the EU AI Act (see references),
while specific requirements are also imposed for certain general-purpose models. ISPE would welcome
alignment with such considerations to leverage existing transparency efforts.

Rationale/detail
For instance, Annex XII of the EU AI Act demands technical documentation for providers of generalpurpose AI models to downstream providers that integrate the model into their AI system, including
tasks, acceptable use, versions, architecture, modalities, information used on the data for training,
testing and validation, and where applicable the type and provenance of data and curation
methodologies.
Recommendation to CDRH
CDRH may consider comparable frameworks encouraging transparency of foundation model providers,
such as those implemented under the EU AI Act.
Supporting evidence/citations
https://digital-strategy.ec.europa.eu/en/policies/contents-code-gpai https://eurlex.europa.eu/eli/reg/2024/1689/oj/eng

Question 26
FDA question: Are there additional considerations that inform the premarket and postmarket
evaluation of agentic GenAI-enabled devices, beyond those applicable to non-agentic GenAI-enabled
devices? How should the elevated risk associated with autonomous multi-step action, tool use, and
reduced opportunity for human review be reflected in acceptance criteria and oversight?
Position summary
ISPE considers the same principles as for general Generative AI applicable; this includes specification and
targeted selection of contextual data, use of tools and choice of human quality gates, alongside
guardrails (e.g., input controls, controls on multi-agent behavior), while demonstrating their
effectiveness and suitability. These principles should stress the use of risk management when applying
these principles.
Recommendation to CDRH
CDRH may consider using, in guidance, the same principles as for general Generative AI, using the same
or very similar language and terms. It is recognized that providing illustrative examples is a challenge in
guidance.
End of comments