Jagadeesha Pampapathi
“Agentic AI is discussed but I think the risk is greater than portrayed.”
What they argued
Agentic AI needs approval checkpoints and runtime supervision before high-risk autonomous deployment; endorses competency framework; locked model versions plus PCCP-specific foundation-model provisions.
Themes it raises
Across the five cross-cutting questions
High-consequence work: Advises
The comment as filed
I reviewed the FDA discussion paper Discussion Paper Considerations for the Regulation of Generative AI-Enabled Medical Devices. Overall, I think this is one of the most thoughtful FDA discussion papers published to date on GenAI regulation and it aligns well with a Total Product Lifecycle (TPLC) approach. It appropriately recognizes that traditional software validation approaches are insufficient for open-ended GenAI systems and introduces a risk-based + competency-based framework. However, As someone working on cloud-based medical software platforms, there are the areas where I believe additional inputs you can consider before making any final paper/draft guideline/Recommendations. Pl find the enclosed document for Suggestions.
Attachment
Review comments By Jagadeesha:
1. Missing Explicit Multiple Function Device Product (MFDP) Considerations
Although FDA briefly references multiple-function products, the discussion does not
sufficiently address:
• Medical + non-medical GenAI functions in the same platform
• Shared foundation models servicing both regulated and non-regulated functions
• Contamination or influence between regulated and non-regulated outputs
• Segregation requirements
• Architecture Recommendations inclusion Medical + non-medical GenAI functions
This is particularly important for platforms like where some functions could be medicaldevice functions while others remain non-device administrative functions.
Recommendation
FDA should provide:
• Explicit MFDP examples
• Boundary control expectations
• Evidence requirements demonstrating independence between regulated and nonregulated functions
2. Cybersecurity is Underrepresented
The paper references guardrails and prompt injection but cybersecurity requirements are not
sufficiently integrated into the framework.
Particularly absent:
• Model poisoning
• Retrieval database attacks
• Prompt injection impact on clinical outputs
• Adversarial attacks
• Agentic tool abuse
• LLM supply chain security
Recommendation
Add a dedicated cybersecurity competency area or explicitly align with:
• FDA Cybersecurity Guidance
• SBOM expectations
• IEC 81001-5-1
• NIST AI RMF
3. Human Factors / Usability Not Prominent Enough
The paper discusses communication quality and automation bias. However, it does not
explicitly connect them to traditional human factors engineering requirements.
For GenAI systems FDA should evaluate:
• User trust calibration
• Overreliance
• Alert fatigue
• Automation bias
• Explainability effectiveness
Recommendation
Human factors validation should become a standalone evaluation pillar.
4. Lack of Clear Treatment for Retrieval-Augmented Generation (RAG)
Most healthcare GenAI systems will not rely solely on foundation models.
Instead, they will rely upon:
• RAG architectures
• Clinical knowledge repositories
• Guideline retrieval systems
• Enterprise databases
The discussion focuses heavily on foundation models but insufficiently addresses:
• Retrieval quality
• Source governance
• Knowledge base drift
• Citation accuracy
Recommendation
FDA should define validation expectations separately for:
• Foundation model
• Retrieval layer
• Orchestrator layer
• Final clinical output
5. Statistics and Acceptance Criteria Remain Vague
One major weakness is that the document does not provide practical approaches for:
• Sample size determination
• Confidence intervals
• Performance thresholds
• Statistical significance for open-ended outputs
FDA raises the question but does not suggest a path forward.
Recommendation
Consider:
• Non-inferiority approaches
• Bayesian approaches
• Risk-adjusted performance thresholds
• Human-AI team effectiveness measures
6. Agentic AI Requires Stronger Controls
Agentic AI is discussed but I think the risk is greater than portrayed.
Agentic systems can:
• Execute actions
• Trigger workflows
• Modify records
• Interact with connected systems
Single-step validation is insufficient.
Recommendation
Require:
• Action approval checkpoints
• Tool-use verification
• Audit trails
• Rollback capabilities
• Runtime supervision
before high-risk autonomous deployment.
7. Third-Party Foundation Model Change Management
This is probably the biggest practical regulatory challenge.
The paper recognises that foundation model providers may change their models without
sponsors initiating updates.
For medical device manufacturers, this creates a major regulatory problem because:
• Behaviour may change overnight
• Performance may drift
• Safety profiles may change
• FDA-cleared evidence may no longer reflect deployed behaviour
Recommendation
FDA should strongly consider:
• Locked model versions
• Mandatory model version traceability
• Supplier quality agreements
• Triggered revalidation criteria
• PCCP-specific provisions for foundation-model updates